Use of two real‐time polymerase chain reactions (PCRs) to detect herpes simplex type 1 and 2‐DNA after automated extraction of nucleic acid

Herpes simplex virus infections may be diagnosed by several techniques, including conventional cell culture and the polymerase chain reaction (PCR). This prospective study compares the analytical performances and usefulness of an in‐house real‐time PCR method and the Light Cycler HSV ½ detection kit™ (Roche Diagnostics, Mannheim, Germany). The results of both PCRs were then compared to those obtained by conventional cell culture. A total of 313 samples were tested (70 dermal samples, 81 cerebrospinal fluids (CSF), 47 ocular, 42 anogenital, 34 throat swabs, and 33 oral samples, 3 whole blood, 2 biopsies, and 1 bronchoalveolar lavage). Samples for molecular assays were extracted twice with the MagNa Pure instrument™ (Roche Molecular Biochemicals, Mannheim, Germany) and tested blind in parallel by the two PCR methods. Most (226) samples were also examined by cell culture. Forty three samples were found positive by both PCRs, whereas 267 were negative. The HSV‐1 and ‐2 typing of positive samples was identical. Three of the samples were positive in the in‐house PCR and negative in the Light Cycler HSV ½ detection kit. There was no statistically significant difference between the two tests. Only one sample gave an invalid result due to negative PCR and negative internal control result. Seven samples were positive by both real‐time PCRs and negative in conventional culture. The PCRs were significantly (P < 0.05) more sensitive. The results show good agreement between the two real‐time PCR methods, with the molecular tests being more sensitive than cell culture. J. Med. Virol. 74:459–462, 2004. © 2004 Wiley‐Liss, Inc.

[1]  P. Massip,et al.  Automated Extraction and Quantification of Human Cytomegalovirus DNA in Whole Blood by Real-Time PCR Assay , 2003, Journal of Clinical Microbiology.

[2]  D. Murdoch,et al.  Failure To Genotype Herpes Simplex Virus by Real-Time PCR Assay and Melting Curve Analysis Due to Sequence Variation within Probe Binding Sites , 2003, Journal of Clinical Microbiology.

[3]  Frank T. Hufert,et al.  Rapid Detection of Herpes Simplex Virus and Varicella-Zoster Virus Infections by Real-Time PCR , 2003, Journal of Clinical Microbiology.

[4]  C. Mengelle,et al.  Quantitation of human cytomegalovirus in recipients of solid organ transplants by real‐time quantitative PCR and pp65 antigenemia , 2003, Journal of medical virology.

[5]  A. Osterhaus,et al.  Diagnosing Herpesvirus Infections by Real-Time Amplification and Rapid Culture , 2003, Journal of Clinical Microbiology.

[6]  A. Nitsche,et al.  Detection and subtyping of Herpes simplex virus in clinical samples by LightCycler PCR, enzyme immunoassay and cell culture , 2002, BMC Microbiology.

[7]  Thomas F. Smith,et al.  Detection of Herpes Simplex Virus DNA in Genital and Dermal Specimens by LightCycler PCR after Extraction using the IsoQuick, MagNA Pure, and BioRobot 9604 Methods , 2001, Journal of Clinical Microbiology.

[8]  Thomas F. Smith,et al.  Evaluation of LightCycler PCR for Implementation of Laboratory Diagnosis of Herpes Simplex Virus Infections , 2000, Journal of Clinical Microbiology.

[9]  A. Berger,et al.  Detection of Herpes Simplex Virus DNA by Real-Time PCR , 2000, Journal of Clinical Microbiology.

[10]  Thomas F. Smith,et al.  Diagnosis of Herpes Simplex Virus Infections in the Clinical Laboratory by LightCycler PCR , 2000, Journal of Clinical Microbiology.

[11]  D. Persing,et al.  Laboratory diagnosis of central nervous system infections with herpes simplex virus by PCR performed with cerebrospinal fluid specimens , 1997, Journal of clinical microbiology.

[12]  P. Lebon,et al.  Amplification and characterization of herpesvirus DNA in cerebrospinal fluid from patients with acute encephalitis , 1991, Journal of clinical microbiology.