An Apparatus for the Simultaneous Measurements of Thermal Conductivity, Thermal Expansion and Thermal Diffusivity of FRPs Using GM Refrigerator

The applications of fibre reinforced plastic (FRP) materials in cryogenic engineering have stimulated keen interest in the investigation of its properties. The reliable design data generated by a precisely controlled setup at identical environment of its applications are extremely important. This paper describes an apparatus based on a GM refrigerator for the simultaneous measurements of thermal conductivity, thermal expansion and thermal diffusivity using a double‐specimen guarded‐hotplate, 3‐terminal capacitance technique and Angstrom method respectively in the temperature range from 30 K to 300 K. An integrated and perfectly insulated sample holder is designed and fabricated in such a way that the simultaneous measurements of the above properties are conveniently and accurately carried out at different temperatures. A set of stability criteria has been followed during the measurements to ensure the accuracy of the experimental data. The setup is calibrated with stainless steel and copper and the experi...