Valve prosthesis-patient mismatch (VP-PM): a long-term perspective.

The concept/phenomenon of valve prosthesis/patient mismatch (VP-PM), described in 1978, has stood the test of time. From that time to 2011, VP-PM has received a great deal of attention but studies have come to varying conclusions. This is largely because of the determination of prosthetic heart valve area [called effective orifice area index (EOAi)] by projection rather than by actual measurement, variable criteria to assess severity of EOAi and the timing of determination of EOAi. All prosthetic heart valves have some degree of VP-PM which must be placed in a proper clinical perspective. This can be done by determining its effects on function and outcomes. For mortality one needs to focus especially on severe/critical degree of VP-PM and determine the cause of death was due to VP-PM. For the period "beyond 2011" a road map is suggested that will have uniformity of assessment of VP-PM and a focusing on the important goals of VP-PM.

[1]  R GORLIN,et al.  Hydraulic formula for calculation of the area of the stenotic mitral valve, other cardiac valves, and central circulatory shunts. I. , 1951, American heart journal.

[2]  G. Melina,et al.  Moderate chronic kidney disease and left ventricular hypertrophy after aortic valve replacement for aortic valve stenosis. , 2010, The Journal of thoracic and cardiovascular surgery.

[3]  Rüdiger Lange,et al.  Impact of the indexed effective orifice area on mid-term cardiac-related mortality after aortic valve replacement , 2010, Heart.

[4]  I. Wilson,et al.  Patient-prosthesis mismatch does not affect survival following aortic valve replacement. , 2006, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[5]  S. Gelsomino,et al.  Patient‐Prosthesis Mismatch After Small‐Size Stentless Aortic Valve Replacement , 2004, Journal of cardiac surgery.

[6]  P. Pibarot,et al.  Patient-prosthesis mismatch can be predicted at the time of operation. , 2001, The Annals of thoracic surgery.

[7]  M. Enriquez-Sarano,et al.  Impact of Prosthesis-Patient Mismatch on Long-Term Survival in Patients With Small St Jude Medical Mechanical Prostheses in the Aortic Position , 2006, Circulation.

[8]  Bijoy K Khandheria,et al.  Recommendations for evaluation of prosthetic valves with echocardiography and doppler ultrasound: a report From the American Society of Echocardiography's Guidelines and Standards Committee and the Task Force on Prosthetic Valves, developed in conjunction with the American College of Cardiology Card , 2009, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[9]  K. Hashimoto,et al.  Patient–prosthesis mismatch may be irrelevant after aortic valve replacement with the 19-mm Perimount pericardial bioprosthesis in patients aged 65 years or older , 2007, Journal of Artificial Organs.

[10]  Ryozo Nagai,et al.  Gene Expression in Fibroblasts and Fibrosis: Involvement in Cardiac Hypertrophy , 2002, Circulation research.

[11]  A. Della Corte,et al.  Prosthesis-patient mismatch in the elderly: survival, ventricular mass regression, and quality of life. , 2008, The Annals of thoracic surgery.

[12]  Fernando Hornero,et al.  Prosthesis - patient mismatch with latest generation supra-annular prostheses. The beginning of the end? , 2007, Interactive cardiovascular and thoracic surgery.

[13]  S. Rahimtoola,et al.  Perspective on valvular heart disease: an update. , 1989, Journal of the American College of Cardiology.

[14]  G. Troise,et al.  Impact of the improvement of valve area achieved with aortic valve replacement on the regression of left ventricular hypertrophy in patients with pure aortic stenosis. , 2005, The Annals of thoracic surgery.

[15]  T. G.,et al.  Albert Einstein: , 1935, Nature.

[16]  S. Rahimtoola,et al.  Percentage of Left Ventricular Stroke Work Loss: A Simple Hemodynamic Concept for Estimation of Severity in Valvular Aortic Stenosis , 1967, Circulation.

[17]  T. Bové,et al.  Stentless and stented aortic valve replacement in elderly patients: Factors affecting midterm clinical and hemodynamical outcome. , 2006, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[18]  Philippe Pibarot,et al.  Impact of Prosthesis-Patient Mismatch on Cardiac Events and Midterm Mortality After Aortic Valve Replacement in Patients With Pure Aortic Stenosis , 2006, Circulation.

[19]  M. Beer,et al.  Impact of Myocardial Fibrosis in Patients With Symptomatic Severe Aortic Stenosis , 2009, Circulation.

[20]  A. Della Corte,et al.  Aortic valve replacement with 19-mm bileaflet prostheses in the elderly: left ventricular mass regression and quality of life. , 2008, Journal of Heart Valve Disease.

[21]  G. Troise,et al.  Impact of valve prosthesis-patient mismatch on left ventricular mass regression following aortic valve replacement. , 2005, The Annals of thoracic surgery.

[22]  R. Virmani,et al.  Pathological Healing Response of Explanted MitraClip Devices , 2011, Circulation.

[23]  M. De Carlo,et al.  Left ventricular reverse remodeling in percutaneous and surgical aortic bioprostheses: an echocardiographic study. , 2011, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[24]  S. Rahimtoola,et al.  Prosthetic Heart Valve , 2011, Circulation.

[25]  Philippe Pibarot,et al.  Comparison of the hemodynamic performance of percutaneous and surgical bioprostheses for the treatment of severe aortic stenosis. , 2009, Journal of the American College of Cardiology.

[26]  M Jones,et al.  In vitro hemodynamic characteristics of tissue bioprostheses in the aortic position. , 1986, The Journal of thoracic and cardiovascular surgery.

[27]  T. Sundt,et al.  Doppler echocardiography of 240 normal Carpentier-Edwards Duraflex porcine mitral bioprostheses: a comprehensive assessment including time velocity integral ratio and prosthesis performance index. , 2009, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[28]  S. Rahimtoola,et al.  Still searching for the elusive Scarlet Pimpernel of today. , 2012, European heart journal.

[29]  S. Rahimtoola,et al.  Choice of prosthetic heart valve for adult patients. , 2003, Journal of the American College of Cardiology.

[30]  A. Nag,et al.  Study of non-muscle cells of the adult mammalian heart: a fine structural analysis and distribution. , 1980, Cytobios.

[31]  R. Bauernschmitt,et al.  Hemodynamic Performance and Incidence of Patient-Prosthesis Mismatch of the Complete Supraannular Perimount Magna Bioprosthesis in the Aortic Position , 2005, The Thoracic and cardiovascular surgeon.

[32]  H. Adachi,et al.  Mid-term outcomes after aortic valve replacement with the 17-mm St. Jude Medical Regent valve. , 2012, Circulation journal : official journal of the Japanese Circulation Society.

[33]  Chad M. House,et al.  How reliable is an effective orifice area indexed chart? , 2009, The Journal of heart valve disease.

[34]  M. Enriquez-Sarano,et al.  Regression in left ventricular mass after aortic valve replacement for chronic aortic regurgitation is unrelated to prosthetic valve size. , 2011, The Journal of thoracic and cardiovascular surgery.

[35]  R. Cohen,et al.  Industry-generated charts for the selection of stented aortic valve prostheses: clinical tool or marketing ploy? , 2011, The Annals of thoracic surgery.

[36]  P. Serruys,et al.  Prosthesis-patient mismatch after transcatheter aortic valve implantation with the medtronic CoreValve system in patients with aortic stenosis. , 2010, The American journal of cardiology.

[37]  S Frank,et al.  Natural history of valvular aortic stenosis. , 1973, British heart journal.

[38]  F J Schoen,et al.  Founder's Award, 25th Annual Meeting of the Society for Biomaterials, perspectives. Providence, RI, April 28-May 2, 1999. Tissue heart valves: current challenges and future research perspectives. , 1999, Journal of biomedical materials research.

[39]  D. Hodge,et al.  Comprehensive echocardiographic assessment of mechanical tricuspid valve prostheses based on early post-implantation echocardiographic studies. , 2011, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[40]  Influence of prosthesis-patient mismatch on diastolic heart failure after aortic valve replacement. , 2008, The Annals of thoracic surgery.

[41]  M. Cheitlin Comparison Between Transcatheter and Surgical Prosthetic Valve Implantation in Patients With Severe Aortic Stenosis and Reduced Left Ventricular Ejection Fraction , 2011 .

[42]  J. Dumesnil,et al.  Validation and applications of indexed aortic prosthetic valve areas calculated by Doppler echocardiography. , 1990, Journal of the American College of Cardiology.

[43]  N. Birkmeyer,et al.  Prosthesis size and long-term survival after aortic valve replacement. , 2003, The Journal of thoracic and cardiovascular surgery.

[44]  E. Braunwald,et al.  Congenital Aortic Stenosis: I. Clinical and Hemodynamic Findings in 100 Patients; II. Surgical Treatment and the Results of Operation , 1963 .

[45]  J. Ross,et al.  Obstruction to left ventricular outflow. Current concepts of management and operative treatment. , 1968, Annals of internal medicine.

[46]  Philippe Pibarot,et al.  Manufacturer-provided effective orifice area index charts and the prevention of prosthesis-patient mismatch. , 2012, The Journal of heart valve disease.

[47]  Gloria Tamborini,et al.  Hemodynamic and clinical impact of prosthesis-patient mismatch after transcatheter aortic valve implantation. , 2011, Journal of the American College of Cardiology.

[48]  K. Hammermeister,et al.  Six month postoperative hemodynamics of the Hancock heterograft and the Björk-Shiley prosthesis: results of a Veterans Administration cooperative prospective randomized trial. , 1988, Journal of the American College of Cardiology.

[49]  H. Baumgartner,et al.  Moderate patient-prosthesis mismatch after valve replacement for severe aortic stenosis has no impact on short-term and long-term mortality , 2008, Heart.

[50]  T. Karamitsos,et al.  The role of cardiovascular magnetic resonance in the evaluation of valve disease. , 2011, Progress in cardiovascular diseases.

[51]  Nag Ac,et al.  Study of non-muscle cells of the adult mammalian heart: a fine structural analysis and distribution. , 1980 .

[52]  Robert Bauernschmitt,et al.  Prediction of valve prosthesis–patient mismatch prior to aortic valve replacement: which is the best method? , 2006, Heart.

[53]  P. Pibarot,et al.  Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. , 2000, Journal of the American College of Cardiology.

[54]  D. Coltart,et al.  Inaccuracies in using aortic valve gradients alone to grade severity of aortic stenosis. , 1989, British heart journal.

[55]  P. Pibarot,et al.  Impact of prosthesis-patient mismatch on hemodynamic and symptomatic status, morbidity and mortality after aortic valve replacement with a bioprosthetic heart valve. , 1998, The Journal of heart valve disease.

[56]  S. Rahimtoola Valvular heart disease: a perspective on the asymptomatic patient with severe valvular aortic stenosis. , 2008, European heart journal.

[57]  R. Bauernschmitt,et al.  Impact of patient-prosthesis mismatch on exercise capacity in patients after bioprosthetic aortic valve replacement , 2007, Heart.

[58]  Bart Meuris,et al.  Prosthesis-Patient Mismatch Predicts Structural Valve Degeneration in Bioprosthetic Heart Valves , 2010, Circulation.

[59]  Yoram Rudy,et al.  Conductive bridges in cardiac tissue: a beneficial role or an arrhythmogenic substrate? , 2004, Circulation research.

[60]  B S Goldman,et al.  Inaccurate and misleading valve sizing: a proposed standard for valve size nomenclature. , 1998, The Annals of thoracic surgery.

[61]  M. Pasque,et al.  POINT: Prosthesis-patient mismatch does not affect survival for patients greater than 70 years of age undergoing bioprosthetic aortic valve replacement. , 2009, The Journal of thoracic and cardiovascular surgery.

[62]  Mario Castaño,et al.  Hemodynamic performance of the Medtronic Mosaic and Perimount Magna aortic bioprostheses: five-year results of a prospectively randomized study. , 2011, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[63]  M. Cheitlin Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient–Prosthesis Mismatch a Relevant Phenomenon? , 2011 .

[64]  M. J. Dalmau,et al.  The Carpentier-Edwards Perimount Magna aortic xenograft: a new design with an improved hemodynamic performance. , 2006, Interactive cardiovascular and thoracic surgery.

[65]  S. Rahimtoola Determining that aortic valve stenosis is severe: back-to-the-future: physical examination and aortic valve area index/energy loss index < or =0.6 cm2/m2. , 2010, JACC. Cardiovascular imaging.

[66]  D. Shavelle Evaluation of Valvular Heart Disease by Cardiac Catheterization and Angiocardiography , 2009 .

[67]  M. De Carlo,et al.  Clinical outcome in patients with 19-mm and 21-mm St. Jude aortic prostheses: comparison at long-term follow-up. , 2002, The Annals of thoracic surgery.

[68]  AlainCarpentier Hemodynamic Factors Affecting the Fate of Valvular Bioprosthesis , 2010 .

[69]  S. Rahimtoola,et al.  The problem of valve prosthesis-patient mismatch. , 1978, Circulation.

[70]  Alexander Albert,et al.  Impact of valve prosthesis-patient mismatch estimated by echocardiographic-determined effective orifice area on long-term outcome after aortic valve replacement. , 2008, American heart journal.

[71]  S. Rahimtoola,et al.  Choice of prosthetic heart valve in adults an update. , 2010, Journal of the American College of Cardiology.

[72]  Jeevanantham Rajeswaran,et al.  Survival after valve replacement for aortic stenosis: implications for decision making. , 2008, The Journal of thoracic and cardiovascular surgery.

[73]  R. De Paulis,et al.  Small "functional" size after mechanical aortic valve replacement: no risk in young to middle-age patients. , 2005, The Annals of thoracic surgery.

[74]  William A. Zoghbi,et al.  Doppler Echocardiographic Assessment of the St. Jude Medical Prosthetic Valve in the Aortic Position Using the Continuity Equation , 1991, Circulation.

[75]  S. Rahimtoola,et al.  Spontaneous changes in left ventricular function between sequential studies. , 1974, The American journal of cardiology.

[76]  C. Joyner,et al.  Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant. , 2002, The Annals of thoracic surgery.

[77]  J. Kovac,et al.  Prosthesis-patient mismatch after transcatheter aortic valve implantation with the Medtronic-Corevalve bioprosthesis. , 2010, European heart journal.