Effect of clearance on cartilage tribology in hip hemi-arthroplasty

Hemi-arthroplasty of the hip (an artificial femoral head articulating against the natural acetabulum) is used to treat fractured necks of femur; however, there is evidence that articulation causes erosion of the cartilage, resulting in pain for the patient. Parameters that may influence this cartilage erosion include head material and roughness, clearance between the head and acetabulum and activity levels of the patient. This study has assessed the effect of clearance of hemi-arthroplasty articulations on the contact stress, friction and cartilage deformation in an in vitro tribological simulation of the hemi-arthroplasty joint that applied dynamic loads and motion. It has been demonstrated that peak contact stress increased from 5.6 to 10.6 MPa as radial clearance increased from small (<0.6 mm) to extra-large (>1.8 mm). In all samples, friction factor increased with time and was significantly less with extra-large clearances compared to small (<0.6 mm), medium (0.6–1.2 mm) and large (1.2–1.8 mm) clearances. The cartilage deformation observed was significantly greater in acetabulum samples paired to give small or extra-large clearances compared to those with medium or large clearances.

[1]  Z. Jin,et al.  The effect of contact stress on cartilage friction, deformation and wear , 2010, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[2]  R. N. Stauffer,et al.  Fixed-head and bipolar hip endoprostheses. A retrospective clinical and roentgenographic study. , 1987, The Journal of arthroplasty.

[3]  C. McCutchen The frictional properties of animal joints , 1962 .

[4]  Zhongmin Jin,et al.  Friction of total hip replacements with different bearings and loading conditions. , 2007, Journal of biomedical materials research. Part B, Applied biomaterials.

[5]  M. Devas,et al.  Prevention of acetabular erosion after hemiarthroplasty for fractured neck of femur. , 1983, The Journal of bone and joint surgery. British volume.

[6]  D. Mehler,et al.  Friction of ceramic and metal hip hemi-endoprostheses against cadaveric acetabula , 2004, Archives of Orthopaedic and Trauma Surgery.

[7]  J. Fisher,et al.  Comparison of human and animal femoral head chondral properties and geometries , 2012, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[8]  J Fisher,et al.  The influence of continuous sliding and subsequent surface wear on the friction of articular cartilage , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[9]  Van C. Mow,et al.  Recent Developments in Synovial Joint Biomechanics , 1980 .

[10]  J Fisher,et al.  The Influence of Loading Time and Lubricant on the Friction of Articular Cartilage , 1996, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[11]  J Fisher,et al.  An investigation of the effect of conformity of knee hemiarthroplasty designs on contact stress, friction and degeneration of articular cartilage: a tribological study. , 2009, Journal of biomechanics.

[12]  J. Kofod,et al.  Moore prosthesis in the treatment of fresh femoral neck fractures. A critical review with special attention to secondary acetabular degeneration. , 1983, Injury.

[13]  N. Sharkey,et al.  Statically equivalent load and support conditions produce different hip joint contact pressures and periacetabular strains. , 1997, Journal of biomechanics.

[14]  W C Hutton,et al.  Contact pressures in the human hip joint. , 1987, The Journal of bone and joint surgery. British volume.

[15]  Z. Jin,et al.  Tribological testing of articular cartilage of the medial compartment of the knee using a friction simulator , 2008 .

[16]  Z. Jin,et al.  Effect of load variation on the friction properties of articular cartilage , 2007 .

[17]  A Rohlmann,et al.  A comparison of hip joint forces in sheep, dog and man. , 1984, Journal of biomechanics.

[18]  G. Beaupré,et al.  Factors influencing changes in articular cartilage following hemiarthroplasty in sheep , 2002, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[19]  Gerard A Ateshian,et al.  The role of interstitial fluid pressurization in articular cartilage lubrication. , 2009, Journal of biomechanics.

[20]  T. Phillips Thompson hemiarthroplasty and acetabular erosion. , 1989, The Journal of bone and joint surgery. American volume.

[21]  D Dowson,et al.  Lubrication and cartilage. , 1976, Journal of anatomy.

[22]  F Eckstein,et al.  Quantitative determination of joint incongruity and pressure distribution during simulated gait and cartilage thickness in the human hip joint , 1999, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.