Optimizing the performance of a PET detector using discrete GSO crystals on a continuous lightguide

The authors are designing a new detector for PET using discrete 4/spl times/4/spl times/10 mm/sup 3/ GSO crystals on a continuous lightguide with 39 mm PMTs. The Light Response Function (LRF) of a detector is the amount of light received by a PMT as a function of the source position. It has to be controlled by a careful design of the lightguide in order to identify 4 mm crystals. The ideal LRF should be narrow with a linear variation over the PMT diameter. Simulations show that a 1.81 cm thick lightguide produces a narrow LRF with good crystal discrimination. However, the tails of this LRF are long. A further improvement can be achieved by using a 2.31 cm lightguide with 5 mm slots cut in its front surface. This results in a sharp edged, almost triangular, LRF. The slotted lightguide also minimizes the spatial dependence on varying depths of interaction of the gamma ray. The effect of varying slot depths was also investigated through the simulations. This was done while keeping the thickness of the lightguide continuous area constant. Experiments were performed and shown to be in general agreement with the simulations. The good spatial resolution and narrow LRF of such a detector will result in a high resolution PET scanner with good count rate capability.

[1]  Joel S. Karp,et al.  Design and performance of the HEAD PENN-PET scanner , 1994 .

[2]  G Muehllehner,et al.  Performance of a position-sensitive scintillation detector. , 1985, Physics in medicine and biology.

[3]  R. Lecomte,et al.  Scintillation light emission studies of LSO scintillators , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[4]  G. Muehllehner,et al.  A high-resolution GSO-based brain PET camera , 1999, 1999 IEEE Nuclear Science Symposium. Conference Record. 1999 Nuclear Science Symposium and Medical Imaging Conference (Cat. No.99CH37019).

[5]  E. Hoffman,et al.  Effect of refraction index and light sharing on detector element identification for 2D detector modules in Positron Emission Tomography , 1994 .

[6]  Joel S. Karp,et al.  A model of the high count rate performance of NaI(Tl)-based PET detectors , 1997 .

[7]  G Muehllehner,et al.  Three-dimensional imaging characteristics of the HEAD PENN-PET scanner. , 1997, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[8]  S Grootoonk,et al.  Performance Evaluation of the Positron Scanner ECAT EXACT , 1992, Journal of computer assisted tomography.

[9]  C. Bohm,et al.  Figures of Merit for Different Detector (Configurations Utilized in High Resolution Positron Cameras , 1986, IEEE Transactions on Nuclear Science.

[10]  W. J. Lorenz,et al.  Performance evaluation of the whole-body PET scanner ECAT EXACT HR + , 1997 .

[11]  Takeo Kanade,et al.  Surface Reflection: Physical and Geometrical Perspectives , 1989, IEEE Trans. Pattern Anal. Mach. Intell..

[12]  Slotted surface treatment on position-sensitive NaI(Tl) detectors to improve spatial resolution , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[13]  Wai-Hoi Wong,et al.  An analog decoding BGO block detector using circular photomultipliers , 1995 .

[14]  R. Nutt,et al.  A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography , 1986, IEEE Transactions on Nuclear Science.

[15]  D. Wolski,et al.  Energy resolution of scintillation detectors readout with large area avalanche photodiodes and photomultipliers , 1998 .

[16]  D. Chesler,et al.  A Positron Tomograph Employing a One Dimension BGO Scintillation Camera , 1983, IEEE Transactions on Nuclear Science.

[17]  David A. Chesler,et al.  Cylindrical PET detector design , 1988 .

[18]  Wai-Hoi Wong,et al.  An analog decoding BGO block detector using circular photomultipliers , 1994, Proceedings of 1994 IEEE Nuclear Science Symposium - NSS'94.

[19]  C. Moisan,et al.  A more physical approach to model the surface treatment of scintillation counters and its implementation into DETECT , 1996, 1996 IEEE Nuclear Science Symposium. Conference Record.

[20]  K. Takagi,et al.  Cerium‐activated Gd2SiO5 single crystal scintillator , 1983 .

[21]  Simon R. Cherry,et al.  A comparison of PET detector modules employing rectangular and round photomultiplier tubes , 1995 .