Hydrodynamic focusing of conducting fluids for conductivity-based biosensors.

Hydrodynamic focusing of a conducting fluid by a non-conducting fluid to form a constricted current path between two sensing electrodes is implemented in order to enhance the sensitivity of a 4-electrode conductance-based biosensor. The sensor has a simple two-inlet T-junction design and performs four-point conductivity measurements to detect particles immobilized between the sensing electrode pair. Computational simulations conducted in conjunction with experimental flow studies using confocal microscopy show that a flat profile for the focused layer is dependent on the Reynolds number for the chosen flow parameters. The results also indicate that a flat focused layer is desirable for both increased sensitivity as well as surface-binding efficiency. Proof of concept for conductance measurements in a hydrodynamically focused conducting fluid was demonstrated with entrapped magnetic beads.

[1]  S. Shoji,et al.  Surface hydrophilic treatment of polyurea film realized by vacuum ultraviolet light irradiation and its application for poly(methylmethacrylate) blood analysis chip , 2008 .

[2]  Victor M Ugaz,et al.  Multivortex micromixing. , 2006, Proceedings of the National Academy of Sciences of the United States of America.

[3]  Stephen C Jacobson,et al.  Diffusion coefficient measurements in microfluidic devices. , 2002, Talanta.

[4]  E. Walsh,et al.  On the out-of-plane divergence of streamtubes in planar mini-scale flow focusing devices , 2007 .

[5]  Alex Groisman,et al.  High-throughput and high-resolution flow cytometry in molded microfluidic devices. , 2006, Analytical chemistry.

[6]  E. Alocilja,et al.  A high density microelectrode array biosensor for detection of E. coli O157:H7. , 2005, Biosensors & bioelectronics.

[7]  S. Hua,et al.  A microfluidic chip for real-time studies of the volume of single cells. , 2009, Lab on a chip.

[8]  Guy Voirin,et al.  Three-dimensional microfluidic confinement for efficient sample delivery to biosensor surfaces. application to immunoassays on planar optical waveguides. , 2002, Analytical chemistry.

[9]  Joel P Golden,et al.  Attachment of plastic fluidic components to glass sensing surfaces. , 2002, Biosensors & bioelectronics.

[10]  H P Schwan,et al.  ELECTRODE POLARIZATION IMPEDANCE AND MEASUREMENTS IN BIOLOGICAL MATERIALS * , 1968, Annals of the New York Academy of Sciences.

[11]  Joel P Golden,et al.  Target delivery in a microfluidic immunosensor. , 2007, Biosensors & bioelectronics.

[12]  D. Beebe,et al.  Controlled microfluidic interfaces , 2005, Nature.

[13]  R. Tompkins,et al.  Continuous inertial focusing, ordering, and separation of particles in microchannels , 2007, Proceedings of the National Academy of Sciences.

[14]  Victor M Ugaz,et al.  Fluid mixing in planar spiral microchannels. , 2006, Lab on a chip.

[15]  P. Yager,et al.  Biotechnology at low Reynolds numbers. , 1996, Biophysical journal.

[16]  David W. M. Marr,et al.  Hydrodynamic focusing for vacuum-pumped microfluidics , 2005 .

[17]  Gwo-Bin Lee,et al.  The hydrodynamic focusing effect inside rectangular microchannels , 2006 .

[18]  S. Takayama,et al.  Gravity-driven microfluidic particle sorting device with hydrodynamic separation amplification. , 2007, Analytical chemistry.

[19]  G. Whitesides,et al.  Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels , 2000 .

[20]  P. Yager,et al.  A rapid diffusion immunoassay in a T-sensor , 2001, Nature Biotechnology.

[21]  S. Takayama,et al.  Microfluidics for flow cytometric analysis of cells and particles , 2005, Physiological measurement.

[22]  Kangjin Kim,et al.  Surface treatment and characterization of PMMA, PHEMA, and PHPMA , 2001 .

[23]  Robert Langer,et al.  Design and Testing of an Impedance-Based Sensor for Monitoring Drug Delivery , 2005 .

[24]  Martin Arundell,et al.  High-speed particle detection in a micro-Coulter counter with two-dimensional adjustable aperture. , 2008, Biosensors & bioelectronics.

[25]  Paul Yager,et al.  Suppression of non-specific adsorption using sheath flow. , 2004, Lab on a chip.