Preliminary resolution performance of the prototype system for a 4-Layer DOI-PET scanner: jPET-D4

We are developing a high-performance brain PET scanner, jPET-D4, which provides 4-layer depth-of-interaction (DOI) information. The scanner is designed to achieve not only high spatial resolution but also high scanner sensitivity with the DOI information obtained from multi-layered thin crystals. The scanner has 5 rings of 24 detector blocks each, and each block consists of 1024 GSO crystals of 2.9 mm/spl times/2.9 mm/spl times/7.5 mm, which are arranged in 4 layers of 16/spl times/16 arrays. At this stage, a pair of detector blocks and a coincidence circuit have been assembled into an experimental prototype gantry. In this paper, as a preliminary experiment, we investigated the performance of the jPET-D4's spatial resolution using the prototype system. First, spatial resolution was measured from a filtered backprojection reconstructed image. To avoid systematic error and reduce computational cost in image reconstruction, we applied the DOI compression (DOIC) method followed by maximum likelihood expectation maximization that we had previously proposed. Trade-off characteristics between background noise and resolution were investigated because improved spatial resolution is possible only when enhanced noise is avoided. Experimental results showed that the jPET-D4 achieves better than 3 mm spatial resolution over the field-of-view.

[1]  Claude Comtat,et al.  Preserving Poisson characteristics of PET data with weighted OSEM reconstruction , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[2]  Keishi Kitamura,et al.  DOI-PET image reconstruction with accurate system modeling that reduces redundancy of the imaging system , 2003 .

[3]  G. Herman,et al.  Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and x-ray photography. , 1970, Journal of theoretical biology.

[4]  H. Murayama,et al.  Three-dimensional array of scintillation crystals with proper reflector arrangement for a depth of interaction detector , 2003, IEEE Transactions on Nuclear Science.

[5]  H. Murayama,et al.  Design of a depth of interaction detector with a PS-PMT for PET , 1999, 1999 IEEE Nuclear Science Symposium. Conference Record. 1999 Nuclear Science Symposium and Medical Imaging Conference (Cat. No.99CH37019).

[6]  L. Shepp,et al.  Maximum Likelihood Reconstruction for Emission Tomography , 1983, IEEE Transactions on Medical Imaging.

[7]  H. Murayama,et al.  A performance study of the next generation PET using a new histogramming method for the DOI detector , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[8]  H. Murayama,et al.  Count rate analysis of PET scanner designs based on a GSO depth of interaction detector with a large area PS-PMT , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).