An improved fast-response vacuum-UV resonance fluorescence CO instrument

The fast-response resonance fluorescence instrument for the airborne measurement of carbon monoxide described by Gerbig et al. [1996] was modified by implementing an improved optical filter with more efficient optics and an optimized resonance lamp. Besides reductions in size and weight, the new instrument achieves a sensitivity 10 times higher, a lower background (65 ppb compared with 250 ppb), and a faster time response (<0.1 s) than the original instrument. The precision is ±1.5 ppb at an atmospheric mixing ratio of 100 ppb CO, and the detection limit is 3 ppb (2σ) for an integration time of 1 s. First results from the North Atlantic Regional Aerosol Characterization Experiment (ACE-2) campaign during July 1997, when the new instrument was deployed aboard the U.K. Meteorological Office C-130 aircraft, are used to demonstrate the performance of the new instrument.

[1]  G. Sachse,et al.  Correlations between ozone and carbon monoxide in the lower stratosphere, folded tropopause, and maritime troposphere , 1987 .

[2]  W. Spicer,et al.  Photoemission and photomultipliers , 1963 .

[3]  Wolfgang Seiler,et al.  The cycle of atmospheric CO. , 1974 .

[4]  Four-laser airborne infrared spectrometer for atmospheric trace gas measurements. , 1996, Applied optics.

[5]  Glen W. Sachse,et al.  Fast‐response, high‐precision carbon monoxide sensor using a tunable diode laser absorption technique , 1987 .

[6]  Paul J. Crutzen,et al.  The origin of ozone in the troposphere , 1978, Nature.

[7]  R. Dickerson,et al.  Modification of a Commercial Gas Filter Correlation CO Detector for Enhanced Sensitivity , 1988 .

[8]  D. McKenna,et al.  Fast response resonance fluorescence CO measurements aboard the C‐130: Instrument characterization and measurements made during North Atlantic Regional Experiment 1993 , 1996 .

[9]  W. Seiler,et al.  Detection of carbon monoxide and hydrogen by conversion of mercury oxide to mercury vapor , 1980 .

[10]  Anne M. Thompson,et al.  Free tropospheric ozone production following entrainment of urban plumes into deep convection , 1992 .

[11]  S. Wofsy,et al.  Tropospheric chemistry: A global perspective , 1981 .

[12]  A resonance-fluorescence instrument for the in-situ measurement of atmospheric carbon monoxide , 1985 .

[13]  W. Seiler,et al.  Correlative nature of ozone and carbon monoxide in the troposphere - Implications for the tropospheric ozone budget , 1983 .

[14]  P. Crutzen,et al.  Biomass Burning in the Tropics: Impact on Atmospheric Chemistry and Biogeochemical Cycles , 1990, Science.

[15]  John S. Holloway,et al.  Export of North American Ozone Pollution to the North Atlantic Ocean , 1993, Science.

[16]  A. Marenco,et al.  Meridional and vertical ozone distribution in the background troposphere (70°n-60°s; 0–12 km altitude) from scientific aircraft measurements during the STRATOZ III experiment (june 1984) , 1989 .

[17]  Kenji Ito,et al.  Absorption cross section measurements of water vapor in the wavelength region 120 nm to 188 nm (Chem. Phys. 211 (1996) 387 391) , 1997 .