Rigid versus flexible plate fixation for periprosthetic femoral fracture-computer modelling of a clinical case.

A variety of plate designs have been implemented for treatment of periprosthetic femoral fracture (PFF) fixation. Controversy, however, exists with regard to optimum fixation methods using these plates. A clinical case of a PFF fixation (Vancouver type C) was studied where a rigid locking plate fixation was compared with a more flexible non-locking approach. A parametric computational model was developed in order to understand the underlying biomechanics between these two fixations. The model was used to estimate the overall stiffness and fracture movement of the two implemented methods. Further, the differing aspects of plate design and application were incrementally changed in four different models. The clinical case showed that a rigid fixation using a 4.5 mm titanium locking plate with a short bridging length did not promote healing and ultimately failed. In contrast, a flexible fixation using 5.6 mm stainless steel non-locking plate with a larger bridging length promoted healing. The computational results highlighted that changing the bridging length made a more substantial difference to the stiffness and fracture movement than varying other parameters. Further the computational model predicted the failure zone on the locking plate. In summary, rigid fracture fixation in the case of PFF can suppress the fracture movement to a degree that prevents healing and may ultimately fail. The computational approach demonstrated the potential of this technique to compare the stiffness and fracture movement of different fixation constructs in order to determine the optimum fixation method for PFF.

[1]  R Zdero,et al.  The biomechanics of human femurs in axial and torsional loading: comparison of finite element analysis, human cadaveric femurs, and synthetic femurs. , 2007, Journal of biomechanical engineering.

[2]  B. Masri,et al.  Fractures of the femur after hip replacement. , 1995, Instructional course lectures.

[3]  D. Hak,et al.  The influence of fracture fixation biomechanics on fracture healing. , 2010, Orthopedics.

[4]  D. Berry,et al.  Periprosthetic Femoral Fractures Around Well-Fixed Implants: Use of Cortical Onlay Allografts with or without a Plate , 2002, The Journal of bone and joint surgery. American volume.

[5]  Alison C Jones,et al.  Periprosthetic fracture fixation of the femur following total hip arthroplasty: a review of biomechanical testing. , 2011, Clinical biomechanics.

[6]  Kavanagh Bf,et al.  Femoral fractures associated with total hip arthroplasty. , 1992 .

[7]  P Augat,et al.  Effect of dynamization on gap healing of diaphyseal fractures under external fixation. , 1995, Clinical biomechanics.

[8]  M J Pearcy,et al.  Computational investigations of mechanical failures of internal plate fixation , 2010, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[9]  C. Bourgeault,et al.  Screw Position Affects Dynamic Compression Plate Strain in an In Vitro Fracture Model , 2001, Journal of orthopaedic trauma.

[10]  Michael Bottlang,et al.  Far cortical locking can improve healing of fractures stabilized with locking plates. , 2010, The Journal of bone and joint surgery. American volume.

[11]  Y. Yoon,et al.  Shape optimal design of the stem of a cemented hip prosthesis to minimize stress concentration in the cement layer. , 1989, Journal of biomechanics.

[12]  Georg N Duda,et al.  Mechanical conditions in the internal stabilization of proximal tibial defects. , 2002, Clinical biomechanics.

[13]  G. Farfalli,et al.  Locking compression plate fixation of Vancouver type-B1 periprosthetic femoral fractures. , 2007, The Journal of bone and joint surgery. American volume.

[14]  Benjamin J. Ellis,et al.  Verification, validation and sensitivity studies in computational biomechanics , 2007, Computer methods in biomechanics and biomedical engineering.

[15]  S M Perren,et al.  EVOLUTION OF THE INTERNAL FIXATION OF LONG BONE FRACTURES , 2002 .

[16]  W R Krause,et al.  Finite-element modelling of femoral shaft fracture fixation techniques post total hip arthroplasty. , 1992, Journal of biomechanics.

[17]  M. Viceconti,et al.  Extracting clinically relevant data from finite element simulations. , 2005, Clinical biomechanics.

[18]  S M Perren,et al.  Physical and biological aspects of fracture healing with special reference to internal fixation. , 1979, Clinical orthopaedics and related research.

[19]  K. Doi,et al.  Early breakage of a titanium volar locking plate for fixation of a distal radius fracture: case report. , 2009, The Journal of hand surgery.

[20]  M. Bottlang,et al.  Locked Plating of Distal Femur Fractures Leads to Inconsistent and Asymmetric Callus Formation , 2010, Journal of orthopaedic trauma.

[21]  R Huiskes,et al.  Stress analyses of implanted orthopaedic joint prostheses for optimal design and fixation. , 1980, Acta orthopaedica Belgica.

[22]  Stephan M Perren,et al.  Evolution of the internal fixation of long bone fractures. The scientific basis of biological internal fixation: choosing a new balance between stability and biology. , 2002, The Journal of bone and joint surgery. British volume.

[23]  E. Abel,et al.  A finite element analysis of hollow stemmed hip prostheses as a means of reducing stress shielding of the femur. , 2001, Journal of biomechanics.

[24]  D. Pichora,et al.  Finite element stress analysis of a hybrid fracture fixation plate. , 1996, Medical engineering & physics.

[25]  Perren Sm,et al.  Physical and biological aspects of fracture healing with special reference to internal fixation. , 1979, Clinical orthopaedics and related research.

[26]  B. Hanson,et al.  Locking compression plate loosening and plate breakage: a report of four cases. , 2004, Journal of orthopaedic trauma.

[27]  R. Zdero,et al.  Biomechanical evaluation of periprosthetic femoral fracture fixation. , 2008, The Journal of bone and joint surgery. American volume.

[28]  Mark A. Lee,et al.  Polyaxial Locking Plate Fixation in Distal Femur Fractures: A Biomechanical Comparison , 2008, Journal of orthopaedic trauma.

[29]  D. Weber,et al.  Distal femoral fractures after knee arthroplasty , 1999, International Orthopaedics.

[30]  B F Kavanagh,et al.  Femoral fractures associated with total hip arthroplasty. , 1992, The Orthopedic clinics of North America.

[31]  Kent N Bachus,et al.  Biomechanical Analysis of Distal Femur Fracture Fixation: Fixed-Angle Screw-Plate Construct Versus Condylar Blade Plate , 2007, Journal of orthopaedic trauma.

[32]  D T Davy,et al.  Torsional loads in the early postoperative period following total hip replacement , 1995, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[33]  F. Kummer,et al.  Distal femoral fixation: a biomechanical comparison of the standard condylar buttress plate, a locked buttress plate, and the 95-degree blade plate. , 1997, Journal of orthopaedic trauma.

[34]  Ulrich Dieter,et al.  Biomechanical testing of the LCP--how can stability in locked internal fixators be controlled? , 2003, Injury.

[35]  Erik N. Kubiak,et al.  Biomechanics of Locked Plates and Screws , 2004, Journal of orthopaedic trauma.

[36]  J. Cooper,et al.  Locking plate osteosynthesis for Vancouver Type B1 and Type C periprosthetic fractures of femur: a report on 12 patients. , 2007, Injury.

[37]  H. Schmotzer,et al.  Surgical management of intra- and postoperative fractures of the femur about the tip of the stem in total hip arthroplasty. , 1996, The Journal of arthroplasty.

[38]  S. Perren,et al.  Plate‐Bone Friction in the Compression Fixation of Fractures , 1972, Clinical orthopaedics and related research.

[39]  M. Bottlang,et al.  Does locked plating of periprosthetic supracondylar femur fractures promote bone healing by callus formation? Two cases with opposite outcomes. , 2008, The Iowa orthopaedic journal.

[40]  G S Beaupré,et al.  Stresses in plated long‐bones: The role of screw tightness and interface slipping , 1988, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.