Brown-Roberts-Wells stereotactic frame modifications to accomplish magnetic resonance imaging guidance in three planes.
暂无分享,去创建一个
The Brown-Roberts-Wells (BRW) computer tomography (CT) stereotactic guidance system has been modified to accommodate magnetic resonance imaging (MRI). A smaller head ring, which fits in standard MRI head coils, is constructed of a non-ferromagnetic aluminum ring that is split to prevent eddy currents and anodized to prevent MRI image distortion and resolution degradation. A new localizing device has been designed in a box configuration, which allows BRW stereotactic coordinates to be calculated from coronal and sagittal MRI images, in addition to axial images. The system was tested utilizing a phantom and T1- and T2-weighted images. Using 5-mm MRI scan slices, targets were localized accurately to a 5-mm cube in three combined planes. Optimized calibration of both low field strength (0.3 T) and high field strength (1.5 T) MRI systems is necessary to obtain thin slice (5 mm) images with acceptable image resolution. To date, 10 patients have had MRI stereotactic localization of brain lesions that were better defined by MRI than CT.