Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose.

BACKGROUND Proposed quality specifications for glucose meters allow results to be in error by 5-10% or more of the "true" concentration. Because meters are used as aids in the adjustment of insulin doses, we aimed to characterize the quantitative effect of meter error on the ability to identify the insulin dose appropriate for the true glucose concentration. METHODS Using Monte Carlo simulation, we generated random "true" glucose values within defined intervals. These values were converted to "measured" glucose values using mathematical models of glucose meters having defined imprecision (CV) and bias. For each combination of bias and imprecision, 10,000-20,000 true and measured glucose concentrations were matched with the corresponding insulin doses specified by selected insulin-dosing regimens. Discrepancies in prescribed doses were counted and their frequencies plotted in relation to bias and imprecision. RESULTS For meters with a total analytical error of 5%, dosage errors occurred in approximately 8-23% of insulin doses. At 10% total error, 16-45% of doses were in error. Large errors of insulin dose (two-step or greater) occurred >5% of the time when the CV and/or bias exceeded 10-15%. Total dosage error rates were affected only slightly by choices of sliding scale among insulin dosage rules or by the range of blood glucose. To provide the intended insulin dosage 95% of the time required that both the bias and the CV of the glucose meter be <1% or <2%, depending on mean glucose concentrations and the rules for insulin dosing. CONCLUSIONS Glucose meters that meet current quality specifications allow a large fraction of administered insulin doses to differ from the intended doses. The effects of such dosage errors on blood glucose and on patient outcomes require study.

[1]  J R Baker,et al.  Analytical Error of Home Glucose Monitors: A Comparison of 18 Systems , 1999, Annals of clinical biochemistry.

[2]  A. Karter,et al.  Self-monitoring of blood glucose: language and financial barriers in a managed care population with diabetes. , 2000, Diabetes care.

[3]  C Ricós,et al.  Desirable routine analytical goals for quantities assayed in serum. Discussion paper from the members of the external quality assessment (EQA) Working Group A on analytical goals in laboratory medicine. , 1995, European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies.

[4]  E. K. Harris,et al.  Biological and analytic components of variation in long-term studies of serum constituents in normal subjects. 3. Physiological and medical implications. , 1970, Clinical chemistry.

[5]  C. Fraser The Necessity of Achieving Good Laboratory Performance , 1990, Diabetic medicine : a journal of the British Diabetic Association.

[6]  C G Fraser,et al.  The application of theoretical goals based on biological variation data in proficiency testing. , 1988, Archives of pathology & laboratory medicine.

[7]  R. Weitgasser,et al.  Newer portable glucose meters--analytical improvement compared with previous generation devices? , 1999, Clinical chemistry.

[8]  D. Cox,et al.  Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood Glucose , 1987, Diabetes Care.

[9]  E. T. Schroeder,et al.  The effects of exercise during hemodialysis on physical performance and nutrition assessment. , 1999, Journal of renal nutrition : the official journal of the Council on Renal Nutrition of the National Kidney Foundation.

[10]  C. Frings,et al.  Improved colorimetric method for assay of amphetamines in urine. , 1971, Clinical chemistry.

[11]  J O Westgard,et al.  Charts of operational process specifications ("OPSpecs charts") for assessing the precision, accuracy, and quality control needed to satisfy proficiency testing performance criteria. , 1992, Clinical chemistry.

[12]  P. H. Petersen,et al.  Analytical performance characteristics should be judged against objective quality specifications. , 1999, Clinical chemistry.

[13]  J. Moss,et al.  Self-Monitoring of Blood Glucose , 1986, Diabetes Care.

[14]  A. Schiffrin,et al.  Comparison Between Continuous Subcutaneous Insulin Infusion and Multiple Injections of Insulin: A One-Year Prospective Study , 1982, Diabetes.