Acoustic assessment of the physical integrity of Björk-Shiley convexo-concave heart valves.

BACKGROUND Several lines of evidence indicate a two-stage failure mode for the Björk-Shiley convexo-concave (C/C) heart valve, in which one of the two outlet strut legs separates from the flange before the other, potentially providing an opportunity to identify and prophylactically replace failure-prone valves. Radiographic single leg separation (SLS) detection, although successful, is subjective and skill intensive, implying a need for both an objective preliminary screen and subsequent corroboration of the radiographic findings. METHODS AND RESULTS We developed a time-windowed, power density analysis of C/C valve closing sounds to detect the vibrational resonance that characterizes the presence of an intact outlet strut in clinically functioning, 29-mm-flange size C/C valves. Recordings from more than 800 patients enrolled in radiographic SLS detection studies were analyzed, and the assessment algorithm was evaluated through a blinded test of 32 study valves for which the true status became known consequent to an autopsy or surgical explantation. Valves were objectively scored on a 0-to- 1 scale, with 1 being assuredly intact and scores of < 0.50 indicating a probable SLS. All except five valves (incorrectly designated probable SLS) were classified correctly, for a sensitivity of 1.00 (95% confidence interval, 0.79 to 1.00) and a specificity of 0.69 (0.41 to 0.89). CONCLUSIONS This level of accuracy is sufficient to serve as an effective preliminary screen, potentially allowing a threefold concentration of SLS prevalence among the C/C valves of patients undergoing radiographic assessment. The value of acoustic classification in avoiding unnecessary operations prompted by false-positive radiographs is less certain.

[1]  James G. Chandler,et al.  Radiographic detection of strut separations in Björk-Shiley convexo-concave mitral valves. , 1995, The New England journal of medicine.

[2]  T D Plemons,et al.  Acoustic classification of the state of artificial heart valves. , 1995, The Journal of the Acoustical Society of America.

[3]  H. Griswold,et al.  Sound Spectrographic Diagnosis of Aortic Ball Variance , 1969, Circulation.

[4]  J V Candy,et al.  Classification of prosthetic heart valve sounds: a parametric approach. , 1995, The Journal of the Acoustical Society of America.

[5]  Terry D. Plemons,et al.  Multiple feature acoustic classification of BSCC artificial heart valves , 1994, Proceedings of IEEE Symposium on Computer-Based Medical Systems (CBMS).

[6]  J. Hanley,et al.  The meaning and use of the area under a receiver operating characteristic (ROC) curve. , 1982, Radiology.

[7]  A. C. Eberhardt,et al.  Acoustic characterization of mechanical valve condition and loading. , 1995, The Journal of heart valve disease.

[8]  H. Griswold,et al.  Phonocardiographic Diagnosis of Aortic Ball Variance , 1968, Circulation.

[9]  J A Swets,et al.  Measuring the accuracy of diagnostic systems. , 1988, Science.

[10]  M. Kendall,et al.  The advanced theory of statistics , 1945 .

[11]  N. Dreyer,et al.  Manufacturing characteristics associated with strut fracture in Björk-Shiley 60 degrees Convexo-Concave heart valves. , 1995, The Journal of heart valve disease.

[12]  James G. Chandler,et al.  Radiographic Detection of Single Strut Leg Separations as a Putative Basis for Prophylactic Explantation of Björk-Shiley Convexo-Concave Heart Valves , 1996, World Journal of Surgery.

[13]  Terry D. Plemons,et al.  Stability of the BSCC heart valve outlet strut resonant frequency under changing physiological conditions and observation times , 1994, Proceedings of IEEE Symposium on Computer-Based Medical Systems (CBMS).