Reproducibility of Anterior Segment Angle Metrics Measurements Derived From Cirrus Spectral Domain Optical Coherence Tomography

Purpose:To investigate the reproducibility of anterior segment angle (ACA) metrics measurements in normal subjects on Cirrus spectral domain optical coherence tomography (SD-OCT). Methods:40 eyes from 20 healthy, normal subjects underwent anterior segment imaging using a Cirrus SD-OCT. For each eye, 2 acquisitions of 5-line raster scans were performed perpendicularly on the inferior (270 degrees) angle. The Schwalbe’s line-angle opening distance (SL-AOD) and Schwalbe’s line-trabecular-iris space area (SL-TISA) measurements were performed by masked certified reading center graders using customized grading software. Intra-acquisition, intergrader, and intragrader reproducibility of SL-AOD and SL-TISA measurements were evaluated by intraclass correlation coefficients (ICC), Bland-Altman plots, and computation of mean percent difference (MPD) and coefficient of variability (CV). Results:The mean SL-AOD (average of 2 acquisitions) was 0.75 mm (range, 0.32 mm to 1.39 mm); SL-TISA was 0.28 mm2 (range, 0.082 mm2 to 0.569 mm2). The repeatability of Cirrus SD-OCT was excellent for both SL-AOD (MPD 4.74%, CV=0.92, ICC=0.99) and SL-TISA (MPD 9.4%, CV=0.8, ICC=0.99). The intragrader reproducibility was high for SL-AOD (MPD 4.28%, CV=0.94, ICC=0.995) and SL-TISA (MPD 6.05%, CV=0.89, ICC=0.993). The inter-grader reproducibility was not as good but still excellent for both SL-AOD (MPD 15.47%, CV=0.95, ICC=0.94) and SL-TISA (MPD 19.43%, CV=0.99, ICC=0.93). Bland-Altman plots of all comparisons did not demonstrate any apparent bias, with similar repeatability at various SL-AOD and SL-TISA values. Conclusions:In our population of young healthy adults with normal eyes, there was excellent intra-acquisition, intragrader, and intergrader reproducibility for Schwalbe’s line-based ACA metrics obtained from Cirrus SD-OCT images. These SD-OCT-derived measures may serve as reliable descriptors of angle morphometry for use in clinical trials and clinical practice.

[1]  Yan Li,et al.  Anterior segment optical coherence tomography for imaging the anterior chamber after laser peripheral iridotomy. , 2007, American journal of ophthalmology.

[2]  Ki Ho Park,et al.  Structure-function relationships in normal and glaucomatous eyes determined by time- and spectral-domain optical coherence tomography. , 2010, Investigative ophthalmology & visual science.

[3]  J. Mehta,et al.  Biometry of the cornea and anterior chamber in chinese eyes: an anterior segment optical coherence tomography study. , 2010, Investigative ophthalmology & visual science.

[4]  Chi Pui Pang,et al.  Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography. , 2008, Ophthalmology.

[5]  T. Aung,et al.  Imaging of the iridocorneal angle with the RTVue spectral domain optical coherence tomography. , 2012, Investigative ophthalmology & visual science.

[6]  J. Mehta,et al.  Reproducibility of cornea measurements in anterior segment OCT images of normal eyes and eyes with bullous keratopathy analyzed with the Zhongshan Assessment Program. , 2011, Investigative ophthalmology & visual science.

[7]  Dariusz Dobrowolski,et al.  Anterior segment imaging: Fourier‐domain optical coherence tomography versus time‐domain optical coherence tomography , 2009, Journal of cataract and refractive surgery.

[8]  Pina Marziliano,et al.  Automatic detection of schwalbe's line in the anterior chamber angle of the eye using HD-OCT images , 2010, 2010 Annual International Conference of the IEEE Engineering in Medicine and Biology.

[9]  D. Friedman,et al.  Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes. , 2007, Ophthalmology.

[10]  Tien Yin Wong,et al.  Novel anterior-chamber angle measurements by high-definition optical coherence tomography using the Schwalbe line as the landmark , 2010, British Journal of Ophthalmology.

[11]  U. Dinç,et al.  Assessment of Anterior Chamber Angle Using Visante OCT, Slit-Lamp OCT, and Pentacam , 2010, European journal of ophthalmology.

[12]  Gabor Nemeth,et al.  Assessment and reproducibility of anterior chamber depth measurement with anterior segment optical coherence tomography compared with immersion ultrasonography , 2007, Journal of cataract and refractive surgery.

[13]  D. Lam,et al.  Repeatability and reproducibility of pachymetric mapping with Visante anterior segment-optical coherence tomography. , 2007, Investigative ophthalmology & visual science.

[14]  G. Dahmen,et al.  Anterior chamber angle measurement with optical coherence tomography: Intraobserver and interobserver variability , 2006, Journal of cataract and refractive surgery.

[15]  L. Módis,et al.  Reliability and repeatability of swept‐source Fourier‐domain optical coherence tomography and Scheimpflug imaging in keratoconus , 2012, Journal of cataract and refractive surgery.

[16]  Tin Aung,et al.  Comparison of two spectral domain optical coherence tomography devices for angle-closure assessment. , 2012, Investigative ophthalmology & visual science.

[17]  David Huang,et al.  Optical coherence tomography assessment of angle anatomy changes after cataract surgery. , 2007, American journal of ophthalmology.

[18]  Tin Aung,et al.  Reproducibility of anterior chamber angle measurements obtained with anterior segment optical coherence tomography. , 2007, Investigative ophthalmology & visual science.

[19]  C. Cheung,et al.  Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography , 2007, British Journal of Ophthalmology.

[20]  H. Quigley,et al.  The number of people with glaucoma worldwide in 2010 and 2020 , 2006, British Journal of Ophthalmology.

[21]  Shu Liu,et al.  Anterior chamber angle imaging with swept-source optical coherence tomography: an investigation on variability of angle measurement. , 2011, Investigative ophthalmology & visual science.

[22]  Tin Aung,et al.  Assessment of the scleral spur in anterior segment optical coherence tomography images. , 2008, Archives of ophthalmology.

[23]  Jing Tian,et al.  Automatic Anterior Chamber Angle Assessment for HD-OCT Images , 2011, IEEE Transactions on Biomedical Engineering.

[24]  B. Lorenz,et al.  Ocular Phenotype in 3 Young Siblings With a Homozygous KCNV2-Mutation Followed for 14 Years , 2007 .

[25]  Kyung R Sung,et al.  Characteristics and reproducibility of anterior chamber angle assessment by anterior‐segment optical coherence tomography , 2011, Acta ophthalmologica.

[26]  H. Quigley Number of people with glaucoma worldwide. , 1996, The British journal of ophthalmology.

[27]  Jing Liu,et al.  Anterior chamber angle measurement with anterior segment optical coherence tomography: a comparison between slit lamp OCT and Visante OCT. , 2008, Investigative ophthalmology & visual science.

[28]  Tin Aung,et al.  Angle imaging: Advances and challenges , 2011, Indian journal of ophthalmology.

[29]  Valéria Lima Passos,et al.  Reproducibility of anterior chamber angle measurements with anterior segment optical coherence tomography. , 2011, Investigative ophthalmology & visual science.