Comparison of MacKay-Marg, Goldmann, and Perkins tonometers in abnormal corneas.

Intraocular pressure was determined manometrically in owl monkey eyes with normal and edematous corneas. The pressure was measured with Perkins hand-held, Goldmann, and MacKay-Marg tonometers to compare relative accuracy. At monometric levels of 20, 30, and 40 mm Hg, a statistically significant difference in readings was found between normal and edematous corneas with use of the Perkins and Goldmann applanators, whereas no such difference was found with the MacKay-Marg tonometer. The MacKay-Marg tonometer most closely approximated the true intraocular pressure in edematous corneas.

[1]  H. Kaufman,et al.  Pressure measurement: which tonometer? , 1972, Investigative ophthalmology.

[2]  H. Kaufman,et al.  Intraocular Pressure Following Penetrating Keratoplasty , 1969 .

[3]  E. Marg,et al.  Evaluation of the basic validity and clinical usefulness of the Mackay-Marg tonometer. , 1962, Investigative ophthalmology.

[4]  S. Waltman,et al.  Validity of Mackay-Marg electronic applanation tonometer in patients with scarred irregular corneas. , 1970, American journal of ophthalmology.

[5]  M. Rubin,et al.  A clinical evaluation of the electronic applanation tonometer. , 1966, American journal of ophthalmology.

[6]  E. Perkins Hand-held applanation tonometer. , 1965, The British journal of ophthalmology.

[7]  H. Kaufman,et al.  Validity of MacKay-Marg applanation tonometry following penetrating keratoplasty in man. , 1971, American journal of ophthalmology.

[8]  R. Moses The Goldmann applanation tonometer. , 1958, American journal of ophthalmology.

[9]  A. Irvine,et al.  Electronic applanation tonometry in corneal edema and keratoplasty. , 1969, Investigative ophthalmology.

[10]  G. Hilton,et al.  Electronic applanation tonometry. , 1966, American journal of ophthalmology.

[11]  S. Mishima,et al.  Measurement of corneal thickness with the Haag-Streit pachometer. , 1968, Archives of ophthalmology.