Factors Affecting the Response of the Bubble Detector BD-100 and a Comparison of its Response to CR-39
暂无分享,去创建一个
The BD-100 is a bubble detector available commercially from Chalk River Nuclear Laboratories, Canada for neutron dosimetry. According to the manufac- turer, the BD-100 detects neutrons over an energy range of 100 keV to 14 MeV and the dose equivalent response is independent of energy. The sensitivity of the detector is dependent upon its temperature at the time of irradiation. The sensitized detector self-nucleates upon sharp impact and when heated to temperatures of 48OC or greater. The BD-100 is insensitive to low energy gamma rays but responds to 6 MeV photons. The sensitivity (bubbles/pSv) -of the BD-100 was found to be energy dependent when exposed to standard neutron sources with average energies ranging from 0.5 to 4.5 MeV. The bubbles formed upon irradiation continued to grow in size with time. The response of eledrochemically etched CR-39 to the same neutron sources is also reported for comparison.
[1] D. Busick. BD-100: THE CHALK RIVER NUCLEAR LABORATORIES’ NEUTRON BUBBLE DETECTOR* , 1987 .
[2] H. Ing. The status of the bubble-damage polymer detector , 1986 .
[3] Apfel,et al. Prediction of the minimum neutron energy to nucleate vapor bubbles in superheated liquids. , 1985, Physical review. A, General physics.
[4] S. Roy,et al. Superheated Drop Detector: A Possible Alternative for Neutron Dosimetry , 1985 .
[5] Robert E. Apfel,et al. The superheated drop detector , 1979 .