Critical examination of heat capacity measurements made on a Quantum Design physical property measurement system

Abstract We examine the operation and performance of an automated heat-capacity measurement system manufactured by Quantum Design (QD). QD’s physical properties measurement system (PPMS) employs a thermal-relaxation calorimeter that operates in the temperature range of 1.8–395 K. The accuracy of the PPMS specific-heat data is determined here by comparing data measured on copper and synthetic sapphire samples with standard literature values. The system exhibits an overall accuracy of better than 1% for temperatures between 100 and 300 K, while the accuracy diminishes at lower temperatures. These data confirm that the system operates within the ±5% accuracy specified by QD. Measurements on gold samples with masses of 4.5 and 88 mg indicate that accuracy of ±3% or better can be achieved below 4 K by using samples with heat capacities that are half or greater than the calorimeter addenda heat capacity. The ability of a PPMS calorimeter to accurately measure sharp features in Cp(T) near phase transitions is determined by measuring the specific heat in the vicinity of the first-order antiferromagnetic transition in Sm2IrIn8 (T0=14 K) and the second-order hidden order (HO) transition in URu2Si2 (TN=17 K). While the PPMS measures Cp(T) near the second-order transition accurately, it is unable to do so in the vicinity of the first-order transition. We show that the specific heat near a first-order transition can be determined from the PPMS-measured decay curves by using an alternate analytical approach. This correction is required because the latent heat liberated/absorbed at the transition results in temperature–decay curves that cannot be described by a single relaxation time constant. Lastly, we test the ability of the PPMS to measure the specific heat of Mg11B2, a superconductor of current interest to many research groups, that has an unusually strong field-dependent specific heat in the mixed state. At the critical temperature the discontinuity in the specific heat is nearly 15% lower than measurements made on the same sample using a semi-adiabatic calorimeter at Lawrence Berkeley National Laboratory.

[1]  Douglas L. Martin Tray type calorimeter for the 15-300 K temperature range: copper as a specific heat standard in this range , 1987 .

[2]  D. L. Sisson,et al.  Simple microcalorimeter for measuring microgram samples at low temperatures , 2001 .

[3]  G. Stewart Measurement of low-temperature specific heat , 1983 .

[4]  K. Ema,et al.  Complex Calorimeter with AC- and Relaxation-Mode Operation , 1993 .

[5]  Douglas L. Martin Specific Heat of Copper, Silver, and Gold below 30°K , 1973 .

[6]  J. D. Jorgensen,et al.  The complex nature of superconductivity in MgB2 as revealed by the reduced total isotope effect , 2001, Nature.

[7]  Jenn-Shyong Hwang,et al.  Measurement of heat capacity by fitting the whole temperature response of a heat-pulse calorimeter , 1997 .

[8]  D. G. Archer Thermodynamic Properties of Synthetic Sapphire (α‐Al2O3), Standard Reference Material 720 and the Effect of Temperature‐Scale Differences on Thermodynamic Properties , 1993 .

[9]  G. Seidel,et al.  Steady-State, ac-Temperature Calorimetry , 1968 .

[10]  B. M. Boerstoel,et al.  The specific heat of palladium, platinum, gold and copper below 30 K , 1971 .

[11]  K. H. Kim,et al.  High magnetic field studies of the hidden order transition in URu2Si2. , 2002, Physical review letters.

[12]  D. Mapother,et al.  AC Measurement of the Heat Capacity of Nickel Near its Critical Point , 1967 .

[13]  R. Howard,et al.  Automated small sample calorimeter , 1975 .

[14]  C. A. Swenson,et al.  Effects of Temperature Scale Differences on the Analysis of Heat Capacity Data: The Specific Heat of Copper from 1 to 30 K , 1972 .

[15]  Maple,et al.  Partially gapped Fermi surface in the heavy-electron superconductor URu2Si2. , 1986, Physical review letters.

[16]  F. Hellman,et al.  Thin film microcalorimeter for heat capacity measurements in high magnetic fields , 2002 .

[17]  R. Howard,et al.  Heat Capacity Measurements on Small Samples at Low Temperatures , 1972 .