Feasibility of 2D antiscatter grid concept for flat panel detectors: preliminary investigation of primary transmission properties.

Suppressing the effects of scattered radiation in flat panel detector, FPD, based CBCT still remains to be a challenge. To address the scatter problem, we have been investigating the feasibility of a two dimensional antiscatter grid (2D ASG) concept for FPDs. Although a 2D ASG can potentially provide high scatter rejection capability, primary transmission characteristics of a 2D ASG and its implications in image quality plays a more critical role in implementation of the 2D ASG concept. Thus, in this work, a computational model was developed to investigate the primary transmission properties of the 2D ASG for various grid and FPD pixel geometries, and the improvement in signal to noise ratio,SNR, was calculated analytically to demonstrate the impact of 2D ASG's transmission characteristics on image quality. Computational model showed that average primary transmission fraction (Tp) strongly depends on the septal thickness of 2D ASG, and 2D ASG can provide higher Tp than existing radiographic ASGs at a septal thickness of 0.1 mm. Due to the higher Tp, 2D ASG was also predicted to provide SNR improvements in projections in low to moderate scatter environments typically observed in CBCT imaging. On the other hand, the model also indicated that the shadow or footprint of the 2D ASG leads to spatially nonuniform variations in primary signal in FPD pixels. Reduction of septal thickness and optimization of 2D ASG's pitch may play an essential role in reducing such variations in primary image signal, and avoiding potential image artifacts associated with 2D ASG's footprint.

[1]  S. Sanada,et al.  Comparison of anti-scatter grids for digital imaging with use of a direct-conversion flat-panel detector , 2011, Radiological Physics and Technology.

[2]  D. L. Bailey Nuclear medicine physics : a handbook for teachers and students , 2014 .

[3]  Yuncheng Zhong,et al.  Reduction of ring artifacts in CBCT: detection and correction of pixel gain variations in flat panel detectors. , 2014, Medical physics.

[4]  J. Boone,et al.  Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner. , 2005, Medical physics.

[5]  Kenneth A Fetterly,et al.  Experimental evaluation of fiber-interspaced antiscatter grids for large patient imaging with digital x-ray systems , 2007, Physics in medicine and biology.

[6]  Jan-Jakob Sonke,et al.  Improved image quality of cone beam CT scans for radiotherapy image guidance using fiber-interspaced antiscatter grid. , 2014, Medical physics.

[7]  Y Higashida,et al.  Performance of antiscatter grids in diagnostic radiology: experimental measurements and Monte Carlo simulation studies. , 1985, Medical physics.

[8]  G Kleinszig,et al.  Antiscatter grids in mobile C-arm cone-beam CT: effect on image quality and dose. , 2011, Medical physics.

[9]  Masahiro Endo,et al.  Effect of scatter radiation on image noise in cone-beam CT , 2000, Medical Imaging.

[10]  Lingyun Chen,et al.  Visibility of microcalcification in cone beam breast CT: effects of X-ray tube voltage and radiation dose. , 2007, Medical physics.

[11]  Selin Carkaci,et al.  Dedicated cone-beam breast CT: feasibility study with surgical mastectomy specimens. , 2007, AJR. American journal of roentgenology.

[12]  R. Jaszczak,et al.  Half-cone beam collimation for triple-camera SPECT systems. , 1996, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[13]  Nicolai A. Moldovan,et al.  Focused two-dimensional antiscatter grid for mammography , 2002, SPIE Optics + Photonics.

[14]  Bill Randolph,et al.  Two-dimensional anti-scatter grids for computed tomography detectors , 2008, SPIE Medical Imaging.

[15]  Jeffrey F Williamson,et al.  Monte Carlo evaluation of scatter mitigation strategies in cone-beam CT. , 2010, Medical physics.

[16]  U Neitzel,et al.  Grids or air gaps for scatter reduction in digital radiography: a model calculation. , 1992, Medical physics.

[17]  Matthias Bertram,et al.  Performance of standard fluoroscopy antiscatter grids in flat-detector-based cone-beam CT , 2004, SPIE Medical Imaging.

[18]  Boyd McCurdy,et al.  Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapy , 2009, Physics in medicine and biology.

[19]  D. Jaffray,et al.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. , 2001, Medical physics.

[20]  A. Sisniega,et al.  Monte Carlo study of the effects of system geometry and antiscatter grids on cone-beam CT scatter distributions. , 2013, Medical physics.

[21]  S Webb,et al.  Removal and effects of scatter-glare in cone-beam CT with an amorphous-silicon flat-panel detector , 2011, Physics in medicine and biology.

[22]  Klaus Klingenbeck,et al.  A general framework and review of scatter correction methods in x-ray cone-beam computerized tomography. Part 1: Scatter compensation approaches. , 2011, Medical physics.