Microneedle Insertion Force Reduction Using Vibratory Actuation

The effect of vibratory actuation on microneedle insertion force was investigated. Hollow micro hypodermic injection needles were fabricated by a two-wafer polysilicon micromolding process. A vibratory actuator operating in the kHz range was coupled with the hypodermic microneedles. The force to insert microneedles into excized animal tissue was measured with a load cell. Results showed a greater than 70% reduction in microneedle insertion force by using vibratory actuation. The application of vibratory actuation provides a promising method to precisely control the microneedle insertion forces to overcome microneedle structural material limitations, minimize insertion pain, and enhance the efficiency of drug delivery.

[1]  A. B. Frazier,et al.  Micromachined needle arrays for drug delivery or fluid extraction. , 1999, IEEE engineering in medicine and biology magazine : the quarterly magazine of the Engineering in Medicine & Biology Society.

[2]  Howard I. Maibach,et al.  Evaluation of the Barrier Function of Skin Using Transepidermal Water Loss (TEWL): A Critical Overview , 2014 .

[3]  Mark G. Allen,et al.  Lack of pain associated with microfabricated microneedles. , 2001 .

[4]  Joseph Kost,et al.  Ultrasound induced delivery of peptides , 1993 .