Statistical process control analysis for patient-specific IMRT and VMAT QA

This work applied statistical process control to establish the control limits of the % gamma pass of patient-specific intensity modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) quality assurance (QA), and to evaluate the efficiency of the QA process by using the process capability index (Cpml). A total of 278 IMRT QA plans in nasopharyngeal carcinoma were measured with MapCHECK, while 159 VMAT QA plans were undertaken with ArcCHECK. Six megavolts with nine fields were used for the IMRT plan and 2.5 arcs were used to generate the VMAT plans. The gamma (3%/3 mm) criteria were used to evaluate the QA plans. The % gamma passes were plotted on a control chart. The first 50 data points were employed to calculate the control limits. The Cpml was calculated to evaluate the capability of the IMRT/VMAT QA process. The results showed higher systematic errors in IMRT QA than VMAT QA due to the more complicated setup used in IMRT QA. The variation of random errors was also larger in IMRT QA than VMAT QA because the VMAT plan has more continuity of dose distribution. The average % gamma pass was 93.7% ± 3.7% for IMRT and 96.7% ± 2.2% for VMAT. The Cpml value of IMRT QA was 1.60 and VMAT QA was 1.99, which implied that the VMAT QA process was more accurate than the IMRT QA process. Our lower control limit for % gamma pass of IMRT is 85.0%, while the limit for VMAT is 90%. Both the IMRT and VMAT QA processes are good quality because Cpml values are higher than 1.0.

[1]  Sandra Nuyts,et al.  Radiation‐induced xerostomia in patients with head and neck cancer , 2006, Cancer.

[2]  C. Perez,et al.  Intensity-modulated radiation therapy reduces late salivary toxicity without compromising tumor control in patients with oropharyngeal carcinoma: a comparison with conventional techniques. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  Cedric X. Yu,et al.  Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. , 2003, Medical physics.

[4]  Vincent Marchesi,et al.  A comprehensive analysis of the IMRT dose delivery process using statistical process control (SPC). , 2009, Medical physics.

[5]  A. Garden,et al.  Radiation‐induced Xerostomia in patients with head and neck cancer: Pathogenesis, impact on quality of life, and management , 2004, Head & neck.

[6]  M. Stock,et al.  Interpretation and evaluation of the γ index and the γ index angle for the verification of IMRT hybrid plans , 2005 .

[7]  Todd Pawlicki,et al.  Statistical process control for radiotherapy quality assurance. , 2005, Medical physics.

[8]  Donald J. Wheeler,et al.  Understanding Statistical Process Control , 1986 .

[9]  Jatinder R. Palta,et al.  Intensity-Modulated Radiation Therapy: The State of the Art , 2003 .

[10]  Mack Roach,et al.  Advances in Radiation Therapy: Conventional to 3D, to IMRT, to 4D, and Beyond , 2005, CA: a cancer journal for clinicians.

[11]  Maurice Pillet,et al.  SPC - Generalization of capability index Cpm - Case of unilateral tolerances , 1997 .

[12]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[13]  Peter Dunscombe,et al.  Experimental validation of the Eclipse AAA algorithm , 2007, Journal of applied clinical medical physics.

[14]  Todd Pawlicki,et al.  Moving from IMRT QA measurements toward independent computer calculations using control charts. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[15]  Benjamin E. Nelms,et al.  A survey on planar IMRT QA analysis , 2007, Journal of applied clinical medical physics.

[16]  C. Fiorino,et al.  Agreement criteria between expected and measured field fluences in IMRT of head and neck cancer: the importance and use of the gamma histograms statistical analysis. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[17]  Stefan Both,et al.  A study to establish reasonable action limits for patient‐specific quality assurance in intensity‐modulated radiation therapy , 2007, Journal of applied clinical medical physics.

[18]  M. Scorsetti,et al.  Early clinical experience with volumetric modulated arc therapy in head and neck cancer patients , 2010, Radiation oncology.

[19]  Michael B Sharpe,et al.  Statistical process control for IMRT dosimetric verification. , 2008, Medical physics.

[20]  B Mijnheer,et al.  Tolerances for the accuracy of photon beam dose calculations of treatment planning systems. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[21]  Parminder S Basran,et al.  An analysis of tolerance levels in IMRT quality assurance procedures. , 2008, Medical physics.

[22]  M. Stock,et al.  Interpretation and evaluation of the gamma index and the gamma index angle for the verification of IMRT hybrid plans. , 2005, Physics in medicine and biology.