An International Comparison of 50-60-Hz Current-Ratio Standards at Currents up to 60000 A
暂无分享,去创建一个
An intercomparison of the ratio errors of the national power frequency current-ratio standards of the National Research Council of Canada and the Physikalisch-Technische Bundesanstalt of Germany has been carried out by means of transfer standard current transformers at ratios of 60 000 A : 5 A, 50 000 A : 5 A, 40 000 A : 5 A. Typical agreement between calibrations at both laboratories is of the order of 0.5 to 1.5 ppm. Details of the transfer standards are given.
[1] N. L. Kusters,et al. A Phantom Burden for Current Transformer Calibration , 1974 .
[2] William J. M. Moore,et al. The Application of Current Comparators to the Calibration of Current Transformers at Ratios up to 36000/5 Amperes , 1968 .
[3] Andreas Braun. Determination of Current Transformer Errors at Primary Currents up to 100 000 A , 1977, IEEE Transactions on Instrumentation and Measurement.