Improved isotopic SIMS measurements of uranium particles for nuclear safeguard purposes

The isotopic analysis of particles containing sub-pg to pg levels of uranium, released from nuclear material handling, has been proven as an efficient tool for international safeguard purposes. Precise and accurate measurement of both enrichment and the minor isotopes is, however, a challenging analytical task due to the low levels of material. One of the mainstay techniques for particle measurement is Secondary Ion Mass Spectrometry (SIMS), this study evaluates the analytical benefit of an alternative in the form of large geometry SIMS (LG-SIMS), which combines high transmission with high mass resolution. We report here that LG-SIMS instruments provide a significantly better measurement quality than the small geometry SIMS as almost all isobaric background interferences are removed at a high useful ion yield. Useful yield measurements, performed on uranium oxide particles with calibrated uranium content, showed an overall useful yield of 1.2% for the LG-SIMS at a mass resolution of 3000. These improvements were then demonstrated by comparing results from actual nuclear inspection samples measured on both instruments. Additional benefits include an increased ability to detect particles of interest in a dust matrix while simultaneously reducing the time of sample analysis. An evaluation on the performance of LG-SIMS compared to Thermal Ion Mass Spectrometry (TIMS) is also presented. This evaluation shows that LG-SIMS has an advantage due to its high ion yield but with a limitation in the detection limit of 236U at higher enrichments due to the necessity for a hydrogen correction.

[1]  M. Betti,et al.  Application of secondary ion mass spectrometry to the identification of single particles of uranium and their isotopic measurement , 1998 .

[2]  K. Habfast Fractionation correction and multiple collectors in thermal ionization isotope ratio mass spectrometry , 1998 .

[3]  M. Betti,et al.  Characterisation of Radioactive Particles by SIMS , 2000, Microchimica Acta.

[4]  G. Wasserburg,et al.  Isotopic Determination of Uranium in Picomole andSubpicomole Quantities , 1981 .

[5]  Development in fission track-thermal ionization mass spectrometry for particle analysis of safeguards environmental samples , 2007 .

[6]  N. Erdmann,et al.  Production of monodisperse uranium oxide particles and their characterization by scanning electron microscopy and secondary ion mass spectrometry , 2000 .

[7]  S. Hart,et al.  Isotope fractionation in secondary ion mass spectrometry , 1982 .

[8]  J. A. Carter,et al.  A simple method to enhance thermal emission of metal ions , 1981 .

[9]  L. Nittler,et al.  Automated isotopic measurements of micron-sized dust: application to meteoritic presolar silicon carbide , 2003 .

[10]  S. Hart,et al.  Applications of the Ion Microprobe to Geochemistry and Cosmochemistry , 1982 .

[11]  K. Watanabe,et al.  Particle isolation for analysis of uranium minor isotopes in individual particles by secondary ion mass spectrometry. , 2007, Talanta.

[12]  G. Wasserburg,et al.  Precise isotopic analysis of lead in picomole and subpicomole quantities , 1975 .

[13]  S Richter,et al.  Improved techniques for high accuracy isotope ratio measurements of nuclear materials using thermal ionization mass spectrometry , 2003 .

[14]  Lothar Koch,et al.  Use of Secondary Ion Mass Spectrometry in Nuclear Forensic Analysis for the Characterization of Plutonium and Highly Enriched Uranium Particles , 1999 .

[15]  D. Simons,et al.  A particulate isotopic standard of uranium and plutonium in an aluminosilicate matrix , 1994, Journal of the American Society for Mass Spectrometry.

[16]  A. Leifer,et al.  Secondary ion mass spectrometry measurements of isotopic ratios: correction for time varying count rate , 2005 .

[17]  J. Sarkis,et al.  Evaluation of the n(235U)/n(238U) isotope ratio measurements in a set of uranium samples by thermal ionization mass spectrometry , 2005 .

[18]  F. Esaka,et al.  Efficient Isotope Ratio Analysis of Uranium Particles in Swipe Samples by Total-Reflection X-ray Fluorescence Spectrometry and Secondary Ion Mass Spectrometry , 2004 .