Rapid acceptance testing of modern linac using on‐board MV and kV imaging systems

Purpose The purpose of this study was to develop a novel process for using on‐board MV and kV Electronic Portal Imaging Devices (EPIDs) to perform linac acceptance testing (AT) for two reasons: (a) to standardize the assessment of new equipment performance, and (b) to reduce the time to clinical use while reducing physicist workload. Methods and materials In this study, Varian TrueBeam linacs equipped with amorphous silicon‐based EPID (aS1000) were used. The conventional set of AT tests and tolerances were used as a baseline guide. A novel methodology was developed or adopted from published literature to perform as many tests as possible using the MV and kV EPIDs. The developer mode on Varian TrueBeam linacs was used to automate the process. In the EPID‐based approach, most of mechanical tests were conducted by acquiring images through a custom phantom and software tools were developed for quantitative analysis to extract different performance parameters. The embedded steel‐spheres in a custom phantom provided both visual and radiographic guidance for beam geometry testing. For photon beams, open field EPID images were used to extract inline/crossline profiles to verify the beam energy, flatness and symmetry. EPID images through a double wedge phantom were used for evaluating electron beam properties via diagonal profile. Testing was augmented with a commercial automated application (Machine Performance Check) which was used to perform several geometric accuracy tests such as gantry, collimator rotations, and couch rotations/translations. Results The developed process demonstrated that the tests, which required customer demonstration, were efficiently performed using EPIDs. The AT tests that were performed using EPIDs were fully automated using the developer mode on the Varian TrueBeam system, while some tests, such as the light field versus radiation field congruence, and collision interlock checks required user interaction. Conclusions On‐board imagers are quite suitable for both geometric and dosimetric testing of linac system involved in AT. Electronic format of the acquired data lends itself to benchmarking, transparency, as well as longitudinal use of AT data. While the tests were performed on a specific model of a linear accelerator, the proposed approach can be extended to other linacs.

[1]  P. Lambin,et al.  A literature review of electronic portal imaging for radiotherapy dosimetry. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[2]  Amarjit Sen,et al.  Commissioning experience and quality assurance of helical tomotherapy machines , 2009, Journal of medical physics.

[3]  Song Gao,et al.  Measurement of changes in linear accelerator photon energy through flatness variation using an ion chamber array. , 2013, Medical physics.

[4]  Jun Tan,et al.  SU‐C‐304‐04: A Compact Modular Computational Platform for Automated On‐Board Imager Quality Assurance , 2015 .

[5]  Mark A. Anastasio,et al.  TH-CD-304-03: A Novel and Intelligent Strategy for Automated On-Board Imager Quality Assurance , 2015 .

[6]  Sridhar Yaddanapudi,et al.  SU-E-T-775: Use of Electronic Portal Imaging Device (EPID) for Quality Assurance (QA) of Electron Beams On Varian Truebeam System , 2015 .

[7]  Peter B Greer,et al.  Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry. , 2005, Medical physics.

[8]  Christof Baltes,et al.  Optimized Varian aSi portal dosimetry: development of datasets for collective use , 2013, Journal of applied clinical medical physics.

[9]  Daniel A Low,et al.  MLC quality assurance using EPID: a fitting technique with subpixel precision. , 2008, Medical physics.

[10]  Mike Kirby,et al.  Measurement possibilities using an electronic portal imaging device , 1994 .

[11]  R Tielenburg,et al.  The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes. , 2004, Medical physics.

[12]  David A. Jaffray,et al.  Quality assurance for the geometric accuracy of cone-beam CT guidance in radiation therapy. , 2008, International journal of radiation oncology, biology, physics.

[13]  Tom Depuydt,et al.  The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[14]  Gwe-Ya Kim,et al.  Automating linear accelerator quality assurance. , 2015, Medical physics.

[15]  Luke Clews,et al.  An EPID based method for efficient and precise asymmetric jaw alignment quality assurance. , 2009, Medical physics.

[16]  A L Boyer,et al.  Application of a video-optical beam imaging system for quality assurance of medical accelerators. , 1998, Physics in medicine and biology.

[17]  T Solberg,et al.  Commissioning of the Varian TrueBeam linear accelerator: a multi-institutional study. , 2013, Medical physics.

[18]  Sridhar Yaddanapudi,et al.  Daily QA of linear accelerators using only EPID and OBI. , 2015, Medical physics.

[19]  J. J. Kruse,et al.  Utilizing an electronic portal imaging device to monitor light and radiation field congruence , 2003, Journal of applied clinical medical physics.

[20]  Fang-Fang Yin,et al.  Task Group 142 report: quality assurance of medical accelerators. , 2009, Medical physics.

[21]  Alberto Boriano,et al.  A new approach for the pixel map sensitivity (PMS) evaluation of an electronic portal imaging device (EPID) , 2013, Journal of applied clinical medical physics.

[22]  Lei Xing,et al.  Quantitative measurement of MLC leaf displacements using an electronic portal image device. , 2004, Physics in medicine and biology.

[23]  G J Budgell,et al.  Daily monitoring of linear accelerator beam parameters using an amorphous silicon EPID , 2007, Physics in medicine and biology.

[24]  E. Bezak,et al.  The linear accelerator mechanical and radiation isocentre assessment with an electronic portal imaging device (EPID) , 2009, Australasian Physics & Engineering Sciences in Medicine.

[25]  J R Sykes,et al.  Measuring linac photon beam energy through EPID image analysis of physically wedged fields. , 2014, Medical physics.

[26]  P. B. Greer,et al.  Long-term two-dimensional pixel stability of EPIDs used for regular linear accelerator quality assurance , 2011, Australasian Physical & Engineering Sciences in Medicine.

[27]  E. Bezak,et al.  Assessment of flatness and symmetry of megavoltage x-ray beam with an electronic portal imaging device (EPID) , 2009, Australasian Physics & Engineering Sciences in Medicine.

[28]  Peter B Greer,et al.  Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy. , 2003, Medical physics.

[29]  Gloria P. Beyer,et al.  Commissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100 linear accelerators , 2013, Journal of applied clinical medical physics.

[30]  Sridhar Yaddanapudi,et al.  A self-sufficient method for calibration of Varian electronic portal imaging device , 2015 .

[31]  Andrea Molineu,et al.  Credentialing results from IMRT irradiations of an anthropomorphic head and neck phantom. , 2013, Medical physics.

[32]  S Yaddanapudi,et al.  TU-C-BRE-02: A Novel, Highly Efficient and Automated Quality Assurance Tool for Modern Linear Accelerators. , 2014, Medical physics.

[33]  Wei Zheng,et al.  Verification of multileaf collimator leaf positions using an electronic portal imaging device. , 2002, Medical physics.

[34]  P M Evans,et al.  Electron beam quality control using an amorphous silicon EPID. , 2009, Medical physics.

[35]  Christof Baltes,et al.  Evaluation of the Machine Performance Check application for TrueBeam Linac , 2015, Radiation oncology.

[36]  Benedick A Fraass,et al.  A method for evaluating quality assurance needs in radiation therapy. , 2008, International journal of radiation oncology, biology, physics.

[37]  Sridhar Yaddanapudi,et al.  SU‐E‐T‐269: Differential Hazard Analysis For Conventional And New Linac Acceptance Testing Procedures , 2015 .

[38]  Sridhar Yaddanapudi,et al.  SU-E-T-781: Using An Electronic Portal Imaging Device (EPID) for Correlating Linac Photon Beam Energies , 2015 .