CT simulation in stereotactic brain radiotherapy--analysis of isocenter reproducibility with mask fixation.

BACKGROUND AND PURPOSE CT verification and measurement of isocenter deviation using repeated mask fixation in linac-based stereotactic high dose radiotherapy of brain metastases were performed in this study. MATERIALS AND METHODS For stereotactic radiotherapy of brain metastases a commercial head mask fixation device based on thermoplastic materials (BrainLAB) was used. A two-step planning-treatment procedure was performed. Immediately before treatment the patient was relocated in the mask and a verification CT scan of the radiopaque marked isocenter was performed and if necessary its position was corrected. The verification procedure is described in detail. Twenty-two CT verifications in 16 patients were analyzed. Deviations were measured separately for each direction. A 3D-deviation vector was calculated. Additionally the average amount of deviation in each of the three dimensions was calculated. RESULTS The mean deviation and standard deviation (SD) of the isocenter was 0.4 mm (SD 1.5 mm) in the longitudinal direction, -0.1 mm (SD 1.8 mm) in the lateral direction and 0.1 mm (SD 1.2 mm) in the anterior-posterior direction. The mean three-dimensional distance (3D-vector) between the verified and the corrected isocenter was 2.4 mm (SD 1.3 mm). The average deviation (without consideration of direction) was 1.1 mm (SD 1.1 mm), 1.3 mm (SD 1.3 mm) and 0.8 mm (SD 0.9 mm) in the longitudinal, lateral and sagittal directions, respectively. No correlation was found between 3D-deviation and the distance of the isocenter from the reference plane nor between deviation and the position of metastases in the brain (central versus peripheral or between different lobes), or the date of treatment. CONCLUSION Reproducibility of the isocenter using the presented mask fixation is in the range of positioning reproducibility reported for other non-invasive fixation devices for stereotactic brain treatment. Our results underline the importance of CT verification as a quality assurance method in stereotactic radiotherapy. Under the condition of a preceding CT verification the mask can be used for single dose stereotactic radiotherapy. For fractionated stereotactic irradiation of small target volumes we recommend repeated CT verifications to assure reproducibility.

[1]  S. Rosenthal,et al.  A precision cranial immobilization system for conformal stereotactic fractionated radiation therapy. , 1995, International journal of radiation oncology, biology, physics.

[2]  R. Laing,et al.  Feasibility of patient immobilization for conventional cranial irradiation with a relocatable stereotactic frame. , 1993, The British journal of radiology.

[3]  L. Laitinen,et al.  An adapter for computed tomography-guided stereotaxis. , 1985, Surgical neurology.

[4]  J. Goble,et al.  A noninvasive thermoplastic head immobilization system. Technical note. , 1995, Journal of neurosurgery.

[5]  J. Sabatier,et al.  Fractionated radiotherapy of small inoperable lesions of the brain using a non-invasive stereotactic frame. , 1991, International journal of radiation oncology, biology, physics.

[6]  M. Phillips,et al.  Stereotactic frame for neuroradiology and charged particle Bragg peak radiosurgery of intracranial disorders. , 1989, International journal of radiation oncology, biology, physics.

[7]  M I Hariz,et al.  Reproducibility of repeated mountings of a noninvasive CT/MRI stereoadapter. , 1986, Applied neurophysiology.

[8]  A P Warrington,et al.  Relocatable frame for stereotactic external beam radiotherapy. , 1991, International journal of radiation oncology, biology, physics.

[9]  L. Laitinen,et al.  A non-invasive method for fractionated stereotactic irradiation of brain tumors with linear accelerator. , 1990, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[10]  Michael Brada,et al.  A non-invasive, relocatable stereotactic frame for fractionated radiotherapy and multiple imaging. , 1991 .

[11]  J Palta,et al.  Systematic analysis of errors in target localization and treatment delivery in stereotactic radiosurgery (SRS). , 1994, International journal of radiation oncology, biology, physics.

[12]  M Bellerive,et al.  Adaptation and verification of the relocatable Gill-Thomas-Cosman frame in stereotactic radiotherapy. , 1994, International journal of radiation oncology, biology, physics.

[13]  H. Frommhold,et al.  Quality assurance for non-invasive patient fixation during stereotactic convergent beam irradiation. , 1994, Acta neurochirurgica. Supplement.