Design and Characterization of a Biocompatible Packaging Concept for Implantable Electronic Devices

A biocompatible packaging process for implantable electronic systems is described, combining biocompatibility and hermeticity with extreme miniaturization. In a first phase of the total packaging sequence, all chips are encapsulated in order to realize a bi-directional diffusion barrier preventing body fluids to leach into the package causing corrosion, and preventing IC materials such as Cu to diffuse into the body, causing various adverse effects. For cost effectiveness, this hermetic chip sealing is performed as post-processing at wafer level, using modifications of standard clean room (CR) fabrication techniques. Well known conductive and insulating CR materials are investigated with respect to their biocompatibility, diffusion barrier properties and sensitivity to corrosion. In a second phase of the packaging process, all chips of the final device should be electrically connected, applying a biocompatible metallization scheme using eg. gold or platinum. For electrodes being in direct contact with the...