Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assurance.

PURPOSE The purpose of this work was to assess the monitor unit (MU) values and position accuracy of spot scanning proton beams as recorded by the daily treatment logs of the treatment control system, and furthermore establish the feasibility of using the delivered spot positions and MU values to calculate and evaluate delivered doses to patients. METHODS To validate the accuracy of the recorded spot positions, the authors generated and executed a test treatment plan containing nine spot positions, to which the authors delivered ten MU each. The spot positions were measured with radiographic films and Matrixx 2D ion-chambers array placed at the isocenter plane and compared for displacements from the planned and recorded positions. Treatment logs for 14 patients were then used to determine the spot MU values and position accuracy of the scanning proton beam delivery system. Univariate analysis was used to detect any systematic error or large variation between patients, treatment dates, proton energies, gantry angles, and planned spot positions. The recorded patient spot positions and MU values were then used to replace the spot positions and MU values in the plan, and the treatment planning system was used to calculate the delivered doses to patients. The results were compared with the treatment plan. RESULTS Within a treatment session, spot positions were reproducible within ±0.2 mm. The spot positions measured by film agreed with the planned positions within ±1 mm and with the recorded positions within ±0.5 mm. The maximum day-to-day variation for any given spot position was within ±1 mm. For all 14 patients, with ∼1 500 000 spots recorded, the total MU accuracy was within 0.1% of the planned MU values, the mean (x, y) spot displacement from the planned value was (-0.03 mm, -0.01 mm), the maximum (x, y) displacement was (1.68 mm, 2.27 mm), and the (x, y) standard deviation was (0.26 mm, 0.42 mm). The maximum dose difference between calculated dose to the patient based on the plan and recorded data was within 2%. CONCLUSIONS The authors have shown that the treatment log file in a spot scanning proton beam delivery system is precise enough to serve as a quality assurance tool to monitor variation in spot position and MU value, as well as the delivered dose uncertainty from the treatment delivery system. The analysis tool developed here could be useful for assessing spot position uncertainty and thus dose uncertainty for any patient receiving spot scanning proton beam therapy.

[1]  Radhe Mohan,et al.  The M. D. Anderson proton therapy system. , 2009, Medical physics.

[2]  James F Dempsey,et al.  Validation of dynamic MLC-controller log files using a two-dimensional diode array. , 2003, Medical physics.

[3]  D Robertson,et al.  Intensity modulated proton therapy treatment planning using single-field optimization: the impact of monitor unit constraints on plan quality. , 2010, Medical physics.

[4]  Robert J. Schneider,et al.  Range modulators for protons and heavy ions , 1975 .

[5]  Alessandra Bolsi,et al.  Treatment planning and verification of proton therapy using spot scanning: initial experiences. , 2004, Medical physics.

[6]  E. Mok,et al.  Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log files , 2010, Physics in medicine and biology.

[7]  John R. Zullo,et al.  An overview of the comprehensive proton therapy machine quality assurance procedures implemented at The University of Texas M. D. Anderson Cancer Center Proton Therapy Center-Houston. , 2009, Medical physics.

[8]  L. Verhey,et al.  Clinical experience and expectation with protons and heavy ions. , 1977, International journal of radiation oncology, biology, physics.

[9]  E. Pedroni,et al.  Intensity modulated proton therapy: a clinical example. , 2001, Medical physics.

[10]  Michael Gillin,et al.  Spot scanning proton beam therapy for prostate cancer: treatment planning technique and analysis of consequences of rotational and translational alignment errors. , 2010, International journal of radiation oncology, biology, physics.

[11]  A. Lomax,et al.  Intensity modulation methods for proton radiotherapy. , 1999, Physics in medicine and biology.

[12]  Bijan Arjomandy,et al.  Patient-specific quality assurance for prostate cancer patients receiving spot scanning proton therapy using single-field uniform dose. , 2011, International journal of radiation oncology, biology, physics.

[13]  Uwe Titt,et al.  Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas M.D. Anderson Cancer Center, Proton Therapy Center, Houston. , 2009, Medical physics.

[14]  Toshikazu Imae,et al.  Quality assurance of volumetric modulated arc therapy using Elekta Synergy , 2009, Acta oncologica.

[15]  Martin Grossmann,et al.  More than 10 years experience of beam monitoring with the Gantry 1 spot scanning proton therapy facility at PSI. , 2009, Medical physics.