Pulse oximeters’ reliability in detecting hypoxemia and bradycardia: Comparison between a conventional and two new generation oximeters

Objective: Pulse oximeters are increasingly used for patient monitoring; however, they are traditionally very prone to motion artifact. Newly developed instruments have lower false alarm rates. We wanted to know whether this is achieved at the expense of an increased proportion of false negative alarms such as missed or delayed identification of hypoxemia and/or bradycardia. Design: Observational study. Setting: Neonatal intensive care unit. Patients: A total of 17 unsedated preterm infants (median gestational age at birth, 25 wks; range, 24‐30 wks). Intervention: Long‐term recordings of transcutaneous partial pressure of oxygen (PTcO2), heart rate, pulse oximeter saturation (SpO2), and pulse rate from a conventional oximeter and two new generation oximeters. Measurements: Recordings were analyzed for episodes with PTcO2 <40 torr or with heart rate <80 beats/min for >5 secs. Hypoxemia was considered identified if SpO2 had fallen to <85% within 2 mins of PTcO2 reaching 40 torr, and bradycardia was considered identified if pulse rate had fallen to <80 beats/min within 2 mins of the heart rate reaching this threshold. Main Results: A total of 202 falls in PTcO2 to <40 torr occurred; 174 (86%) were identified by all three oximeters. Of the remaining episodes, manual analysis of red and infrared absorption signals confirmed that SpO2 had indeed been <85% for ≥10 secs in 11 episodes; therefore, these episodes should have been identified by all three oximeters. None of these had been missed by the conventional oximeter, but 10 (5.4% of the total) were missed by one of the new generation instruments (Nellcor), and one (0.5%) was missed by the other (Masimo). Of 54 bradycardias, only 14 were identified by all three oximeters; 17 (32%) were missed by the conventional, 37 (69%) by the Nellcor, and 4 (7%) by the Masimo instrument. Conclusion: One of the two new generation instruments investigated in this study missed 5.4% of hypoxemic episodes and 69% of bradycardias. It thus appears that this instrument's reduced false alarm rate is achieved at the expense of an unreliable and/or delayed identification of hypoxemia and bradycardia. The other instrument identified both conditions equally as or more reliably than a conventional pulse oximeter.

[1]  D. Shannon,et al.  Apnea and periodic breathing in normal full‐term infants during the first twelve months , 1985, Pediatric pulmonology.

[2]  T. L. Rusch,et al.  Signal processing methods for pulse oximetry , 1996, Comput. Biol. Medicine.

[3]  Christian F Poets,et al.  Gasping and Other Cardiorespiratory Patterns during Sudden Infant Deaths , 1999, Pediatric Research.

[4]  R. B. Cherry,et al.  A FURTHER EXTENSION OF THE IN VIVO OXYGEN-DISSOCIATION CURVE FOR THE BLOOD OF THE NEWBORN INFANT. , 1964, Journal of Clinical Investigation.

[5]  S J Barker,et al.  Effects of Motion on the Performance of Pulse Oximeters in Volunteers , 1996, Anesthesiology.

[6]  Failure of conventional monitoring to detect apnea resulting in hypoxemia. , 1979, Birth defects original article series.

[7]  C. Poets,et al.  Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern. , 1994, Pediatrics.

[8]  Ehud Zmora,et al.  Nurses' response to alarms from monitoring systems in NICU. 1027 , 1997, Pediatric Research.

[9]  S. Lawless Crying wolf: False alarms in a pediatric intensive care unit , 1994, Critical care medicine.

[10]  C. Poets,et al.  Home monitoring of transcutaneous oxygen tension in the early detection of hypoxaemia in infants and young children. , 1991, Archives of disease in childhood.

[11]  K. Tremper,et al.  Clinical Evaluation of a Prototype Motion Artifact Resistant Pulse Oximeter in the Recovery Room , 1996, Anesthesia and analgesia.

[12]  R. Martin,et al.  Relationship between transcutaneous and arterial oxygen tension in sick neonates during mild hyperoxemia , 1982, Critical care medicine.

[13]  Steven J. Barker,et al.  The effects of motion on the performance of pulse oximeters in volunteers (revised publication). , 1997 .