Imaging the living retina at the cellular level with AO parallel spectral-domain optical coherence tomography

We investigate a novel camera that incorporates adaptive optics (AO) and optical coherence tomography (OCT) to determine if it can achieve the necessary 3-D resolution, sensitivity, and speed for imaging individual cells in the living human retina. An AO spectral-domain OCT system was constructed that is based on a free-space Michelson interferometer design. The OCT sub-system consists of a broadband superluminescent diode whose beam passes through an astigmatic lens to form a line illumination pattern on the retina, which is then imaged onto the slit of an imaging spectrometer. The detector of the spectrometer is a scientific-grade areal CCD. Conventional flood illumination, also with AO, was integrated into the camera and provided confirmation of the focus position in the retina. Short bursts of narrow B-scans (100x560 microns) of the living retina were subsequently acquired at 500 Hz during dynamic compensation that corrected the most significant ocular aberrations across a dilated 6 mm pupil. Camera sensitivity (up to 94 dB) was sufficient for observing reflections from essentially all neural layers of the retina. The 3-D resolution of the B-scans (3.0x3.0x5.7 microns) is the highest reported to date in the living human eye. It was sufficient to observe the interface between the inner and outer segments of individual photoreceptor cells, resolved in both lateral and axial dimensions. The waveguiding nature of the photoreceptors is suggestive at multiple reflective sites. Micro-movements of the retina during short burst imaging allow averaging to reduce speckle contrast, but they appear insufficient for significant speckle reduction.

[1]  Yudong Zhang,et al.  Small table-top adaptive optical systems for human retinal imaging , 2002, SPIE Optics + Photonics.

[2]  W Drexler,et al.  Ultrahigh resolution Fourier domain optical coherence tomography. , 2004, Optics express.

[3]  A. Hendrickson,et al.  Human photoreceptor topography , 1990, The Journal of comparative neurology.

[4]  G. M. Morris,et al.  Images of cone photoreceptors in the living human eye , 1996, Vision Research.

[5]  Teresa C. Chen,et al.  Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography. , 2004, Optics express.

[6]  J. Schuman,et al.  Optical coherence tomography. , 2000, Science.

[7]  Wenhan Jiang,et al.  Hartmann-Shack wavefront sensing and wavefront control algorithm , 1990, Other Conferences.

[8]  J. Duker,et al.  Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation. , 2004, Optics express.

[9]  Francois Lacombe,et al.  Towards wide-field retinal imaging with adaptive optics , 2004 .

[10]  A. Fercher,et al.  Parallel Fourier domain optical coherence tomography for in vivo measurement of the human eye. , 2005, Optics express.

[11]  M. Larsen,et al.  Enhanced optical coherence tomography imaging by multiple scan averaging , 2005, British Journal of Ophthalmology.

[12]  B. Singer,et al.  Improvement in retinal image quality with dynamic correction of the eye's aberrations. , 2001, Optics express.

[13]  D R Williams,et al.  Supernormal vision and high-resolution retinal imaging through adaptive optics. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[14]  J. Fujimoto,et al.  Ultrahigh-resolution ophthalmic optical coherence tomography , 2001, Nature Medicine.

[15]  A. Bradley,et al.  Statistical variation of aberration structure and image quality in a normal population of healthy eyes. , 2002, Journal of the Optical Society of America. A, Optics, image science, and vision.

[16]  Robert K. Tyson Principles of Adaptive Optics , 1991 .

[17]  T. Hebert,et al.  Adaptive optics scanning laser ophthalmoscopy. , 2002, Optics express.

[18]  Ravi S. Jonnal,et al.  Coherence gating and adaptive optics in the eye , 2003, SPIE BiOS.

[19]  Teresa C. Chen,et al.  In vivo human retinal imaging by ultrahigh-speed spectral domain optical coherence tomography. , 2004, Optics letters.

[20]  Joseph A. Izatt,et al.  Exposure time dependence of image quality in high-speed retinal in vivo Fourier-domain OCT , 2005, SPIE BiOS.

[21]  P. Artal,et al.  Adaptive-optics ultrahigh-resolution optical coherence tomography. , 2004, Optics letters.