Pulse Arrival Time Based Cuff-Less and 24-H Wearable Blood Pressure Monitoring and its Diagnostic Value in Hypertension

Ambulatory blood pressure monitoring (ABPM) has become an essential tool in the diagnosis and management of hypertension. Current standard ABPM devices use an oscillometric cuff-based method which can cause physical discomfort to the patients with repeated inflations and deflations, especially during nighttime leading to sleep disturbance. The ability to measure ambulatory BP accurately and comfortably without a cuff would be attractive. This study validated the accuracy of a cuff-less approach for ABPM using pulse arrival time (PAT) measurements on both healthy and hypertensive subjects for potential use in hypertensive management, which is the first of its kind. The wearable cuff-less device was evaluated against a standard cuff-based device on 24 subjects of which 15 have known hypertension. BP measurements were taken from each subject over a 24-h period by the cuff-less and cuff-based devices every 15 to 30 minutes during daily activities. Mean BP of each subject during daytime, nighttime and over 24-h were calculated. Agreement between mean nighttime systolic BP (SBP) and diastolic (DBP) measured by the two devices evaluated using Bland-Altman plot were −1.4 ± 6.6 and 0.4 ± 6.7 mmHg, respectively. Receiver operator characteristics (ROC) statistics was used to assess the diagnostic accuracy of the cuff-less approach in the detection of BP above the hypertension threshold during nighttime (>120/70 mmHg). The area under ROC curves were 0.975/0.79 for nighttime. The results suggest that PAT-based approach is accurate and promising for ABPM without the issue of sleep disturbances associated with cuff-based devices.

[1]  D. Seals,et al.  The influence of aging on the human sympathetic nervous system and brain norepinephrine turnover. , 2002, American journal of physiology. Regulatory, integrative and comparative physiology.

[2]  Ju Han Kim,et al.  Not nondipping but nocturnal blood pressure predicts left ventricular hypertrophy in the essential hypertensive patients: the Korean Ambulatory Blood Pressure multicenter observational study , 2014, Journal of hypertension.

[3]  P. Wagner Muscle intracellular oxygenation during exercise: optimization for oxygen transport, metabolism, and adaptive change , 2011, European Journal of Applied Physiology.

[4]  Carmen C. Y. Poon,et al.  Attenuation of Systolic Blood Pressure and Pulse Transit Time Hysteresis During Exercise and Recovery in Cardiovascular Patients , 2014, IEEE Transactions on Biomedical Engineering.

[5]  Kazuo Eguchi,et al.  What is the optimal interval between successive home blood pressure readings using an automated oscillometric device? , 2009, Journal of hypertension.

[6]  Wang Yi,et al.  Noninvasive and Continuous Blood Pressure Monitoring Using Wearable Body Sensor Networks , 2015, IEEE Intelligent Systems.

[7]  Survi Kyal,et al.  Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice , 2015, IEEE Transactions on Biomedical Engineering.

[8]  A. Sacchetti,et al.  Interrogation of Patient Smartphone Activity Tracker to Assist Arrhythmia Management. , 2016, Annals of emergency medicine.

[9]  Carmen C. Y. Poon,et al.  An Armband Wearable Device for Overnight and Cuff-Less Blood Pressure Measurement , 2014, IEEE Transactions on Biomedical Engineering.

[10]  G. Grassi,et al.  Sympathetic activation in the pathogenesis of hypertension and progression of organ damage. , 1999, Hypertension.

[11]  J. Ruiz-Rodríguez,et al.  Innovative continuous non-invasive cuffless blood pressure monitoring based on photoplethysmography technology , 2013, Intensive Care Medicine.

[12]  A. Patzak,et al.  Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method , 2011, European Journal of Applied Physiology.

[13]  Satoru Miyano,et al.  Global implementation of genomic medicine: We are not alone , 2015, Science Translational Medicine.

[14]  C. Goodman Association for the Advancement of Medical Instrumentation , 1988 .

[15]  P. Challande,et al.  Intrinsic stiffness of the carotid arterial wall material in essential hypertensives. , 2000, Hypertension.

[16]  Robert P Light,et al.  The effect of measuring ambulatory blood pressure on nighttime sleep and daytime activity--implications for dipping. , 2010, Clinical journal of the American Society of Nephrology : CJASN.

[17]  Mehmet Rasit Yuce,et al.  A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time , 2015, Physiological measurement.

[18]  G. Stergiou,et al.  Prediction of treatment-induced changes in target-organ damage using changes in clinic, home and ambulatory blood pressure , 2014, Hypertension Research.

[19]  Giuseppe Mancia,et al.  Ambulatory Blood Pressure Is Superior to Clinic Blood Pressure in Predicting Treatment-Induced Regression of Left Ventricular Hypertrophy , 1997 .

[20]  Carmen C. Y. Poon,et al.  A clip-free eyeglasses-based wearable monitoring device for measuring photoplethysmograhic signals , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[21]  Jeroen J. Bax,et al.  2007 ESH-ESC Guidelines for the management of arterial hypertension: the task force for the management of arterial hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC). , 2007, Blood pressure.

[22]  Da Tao,et al.  A 3-Month Randomized Controlled Pilot Trial of a Patient-Centered, Computer-Based Self-Monitoring System for the Care of Type 2 Diabetes Mellitus and Hypertension , 2016, Journal of Medical Systems.

[23]  J. Quenot,et al.  Pulse Wave Transit Time Measurements of Cardiac Output in Septic Shock Patients: A Comparison of the Estimated Continuous Cardiac Output System with Transthoracic Echocardiography , 2015, PloS one.

[24]  A. Jula,et al.  Office, Home, and Ambulatory Blood Pressures as Predictors of Cardiovascular Risk , 2014, Hypertension.

[25]  Carmen C. Y. Poon,et al.  Evaluation scale to assess the accuracy of cuff-less blood pressure measuring devices , 2009, Blood pressure monitoring.

[26]  Yong-Seok Park,et al.  Beat-to-Beat Tracking of Systolic Blood Pressure Using Noninvasive Pulse Transit Time During Anesthesia Induction in Hypertensive Patients , 2013, Anesthesia and analgesia.

[27]  Mirza Mansoor Baig,et al.  Smart Health Monitoring Systems: An Overview of Design and Modeling , 2013, Journal of Medical Systems.

[28]  L. Geddes,et al.  Measurements of Young's Modulus of Elasticity of the Canine Aorta with Ultrasound , 1979 .

[29]  A Mathewson,et al.  An examination of calibration intervals required for accurately tracking blood pressure using pulse transit time algorithms , 2013, Journal of Human Hypertension.

[30]  Zhilin Zhang,et al.  Photoplethysmography-Based Heart Rate Monitoring in Physical Activities via Joint Sparse Spectrum Reconstruction , 2015, IEEE Transactions on Biomedical Engineering.

[31]  Guanqun Zhang,et al.  Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure. , 2011, Journal of applied physiology.

[32]  Carmen C. Y. Poon,et al.  Cuff-less and Noninvasive Measurements of Arterial Blood Pressure by Pulse Transit Time , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[33]  Eric J. Topol,et al.  The emerging field of mobile health , 2015, Science Translational Medicine.

[34]  Joseph Finkelstein,et al.  Introducing Contactless Blood Pressure Assessment Using a High Speed Video Camera , 2016, Journal of Medical Systems.

[35]  Jan A. Staessen,et al.  Daytime and Nighttime Blood Pressure as Predictors of Death and Cause-Specific Cardiovascular Events in Hypertension , 2008, Hypertension.

[36]  R. Payne,et al.  Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure. , 2006, Journal of applied physiology.

[37]  Daniel W. Jones,et al.  Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. , 2005, Hypertension.

[38]  Carmen C. Y. Poon,et al.  Time-frequency analysis of variabilities of heart rate, systolic blood pressure and pulse transit time before and after exercise using the recursive autoregressive model , 2011, Biomed. Signal Process. Control..

[39]  R. Hermida,et al.  Ambulatory Blood Pressure Monitoring: Importance of Sampling Rate and Duration—48 Versus 24 Hours—on the Accurate Assessment of Cardiovascular Risk , 2013, Chronobiology international.

[40]  Emma Pickwell-MacPherson,et al.  The effects of pre-ejection period on post-exercise systolic blood pressure estimation using the pulse arrival time technique , 2010, European Journal of Applied Physiology.

[41]  Charles Sodini,et al.  A wearable vital signs monitor at the ear for continuous heart rate and Pulse Transit Time measurements , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.