A Versatile and Reproducible Multi-Frequency Electrical Impedance Tomography System

A highly versatile Electrical Impedance Tomography (EIT) system, nicknamed the ScouseTom, has been developed. The system allows control over current amplitude, frequency, number of electrodes, injection protocol and data processing. Current is injected using a Keithley 6221 current source, and voltages are recorded with a 24-bit EEG system with minimum bandwidth of 3.2 kHz. Custom PCBs interface with a PC to control the measurement process, electrode addressing and triggering of external stimuli. The performance of the system was characterised using resistor phantoms to represent human scalp recordings, with an SNR of 77.5 dB, stable across a four hour recording and 20 Hz to 20 kHz. In studies of both haeomorrhage using scalp electrodes, and evoked activity using epicortical electrode mats in rats, it was possible to reconstruct images matching established literature at known areas of onset. Data collected using scalp electrode in humans matched known tissue impedance spectra and was stable over frequency. The experimental procedure is software controlled and is readily adaptable to new paradigms. Where possible, commercial or open-source components were used, to minimise the complexity in reproduction. The hardware designs and software for the system have been released under an open source licence, encouraging contributions and allowing for rapid replication.

[1]  I Frerichs,et al.  Electrical impedance tomography (EIT) in applications related to lung and ventilation: a review of experimental and clinical activities. , 2000, Physiological measurement.

[2]  David S. Holder,et al.  Imaging fast electrical activity in the brain with electrical impedance tomography , 2016, NeuroImage.

[3]  M. A. Friedman,et al.  Thalamo‐cortical processing of vibrissal information in the rat. I. Intracortical origins of surround but not centre‐receptive fields of layer IV neurones in the rat S1 barrel field cortex , 1991, The Journal of comparative neurology.

[4]  Richard Bayford,et al.  Bioimpedance imaging: an overview of potential clinical applications. , 2012, The Analyst.

[5]  Rebecca L. Robinson,et al.  A portable instrument for high-speed brain function imaging: FEITER , 2011, 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[6]  R Bayford,et al.  Design and calibration of a compact multi-frequency EIT system for acute stroke imaging. , 2006, Physiological measurement.

[7]  D S Holder,et al.  An electrode addressing protocol for imaging brain function with electrical impedance tomography using a 16-channel semi-parallel system , 2009, Physiological measurement.

[8]  H. Jasper Report of the committee on methods of clinical examination in electroencephalography , 1958 .

[9]  N. Logothetis,et al.  In Vivo Measurement of Cortical Impedance Spectrum in Monkeys: Implications for Signal Propagation , 2007, Neuron.

[10]  Jin Keun Seo,et al.  The Linearized Inverse Problem in Multifrequency Electrical Impedance Tomography , 2016, SIAM J. Imaging Sci..

[11]  J. B. Ranck,et al.  Analysis of specific impedance of rabbit cerebral cortex. , 1963, Experimental neurology.

[12]  Alexander Hartov,et al.  Intracranial Electrical Impedance Tomography: A Method of Continuous Monitoring in an Animal Model of Head Trauma , 2013, Anesthesia and analgesia.

[13]  Simon R. Arridge,et al.  Multifrequency Electrical Impedance Tomography Using Spectral Constraints , 2014, IEEE Transactions on Medical Imaging.

[14]  R H Bayford,et al.  Multi-frequency electrical impedance tomography (EIT) of the adult human head: initial findings in brain tumours, arteriovenous malformations and chronic stroke, development of an analysis method and calibration , 2006, Physiological measurement.

[15]  Andy Adler,et al.  Whither lung EIT: Where are we, where do we want to go and what do we need to get there? , 2012, Physiological measurement.

[16]  C. Petersen The Functional Organization of the Barrel Cortex , 2007, Neuron.

[17]  C Blochet,et al.  In vivo bioimpedance changes during haemorrhagic and ischaemic stroke in rats: towards 3D stroke imaging using electrical impedance tomography , 2016, Physiological measurement.

[18]  Eung Je Woo,et al.  Multi-frequency EIT system with radially symmetric architecture: KHU Mark1 , 2007, Physiological measurement.

[19]  Richard H. Bayford,et al.  Three-Dimensional Electrical Impedance Tomography of Human Brain Activity , 2001, NeuroImage.

[20]  Xuetao Shi,et al.  Performance Evaluation of Five Types of Ag/AgCl Bio-Electrodes for Cerebral Electrical Impedance Tomography , 2011, Annals of Biomedical Engineering.

[21]  C. Gabriel,et al.  Electrical conductivity of tissue at frequencies below 1 MHz , 2009, Physics in medicine and biology.

[22]  Ryan J. Halter,et al.  FPGA-Based Voltage and Current Dual Drive System for High Frame Rate Electrical Impedance Tomography , 2015, IEEE Transactions on Medical Imaging.

[23]  D S Holder,et al.  Analysis of resting noise characteristics of three EIT systems in order to compare suitability for time difference imaging with scalp electrodes during epileptic seizures , 2007, Physiological measurement.

[24]  Robert Oostenveld,et al.  The five percent electrode system for high-resolution EEG and ERP measurements , 2001, Clinical Neurophysiology.

[25]  Anna N. Vongerichten,et al.  Characterisation and imaging of cortical impedance changes during interictal and ictal activity in the anaesthetised rat , 2016, NeuroImage.

[26]  D S Holder,et al.  Comparison of frequency difference reconstruction algorithms for the detection of acute stroke using EIT in a realistic head-shaped tank , 2012, Physiological measurement.

[27]  Simon Arridge,et al.  Stroke type differentiation using spectrally constrained multifrequency EIT: evaluation of feasibility in a realistic head model , 2014, Physiological measurement.

[28]  Eung Je Woo,et al.  Validation of weighted frequency-difference EIT using a three-dimensional hemisphere model and phantom. , 2011, Physiological measurement.

[29]  Simon Arridge,et al.  A Reconstruction-Classification Method for Multifrequency Electrical Impedance Tomography , 2015, IEEE Transactions on Medical Imaging.

[30]  C Blochet,et al.  In vivo bioimpedance measurement of healthy and ischaemic rat brain: implications for stroke imaging using electrical impedance tomography , 2015, Physiological measurement.

[31]  Kirill Y Aristovich,et al.  A method for reconstructing tomographic images of evoked neural activity with electrical impedance tomography using intracranial planar arrays , 2014, Physiological measurement.

[32]  Kirill Y Aristovich,et al.  Investigation of potential artefactual changes in measurements of impedance changes during evoked activity: implications to electrical impedance tomography of brain function , 2015, Physiological measurement.

[33]  H Griffiths,et al.  A Cole phantom for EIT. , 1995, Physiological measurement.

[34]  Bing Li,et al.  Use of Electrical Impedance Tomography to Monitor Regional Cerebral Edema during Clinical Dehydration Treatment , 2014, PloS one.

[35]  E. De Schutter,et al.  Comparing BOLD fMRI signal changes in the awake and anesthetized rat during electrical forepaw stimulation. , 2001, Magnetic resonance imaging.

[36]  Seong-Gi Kim,et al.  Relationship between neural, vascular, and BOLD signals in isoflurane-anesthetized rat somatosensory cortex. , 2006, Cerebral cortex.

[37]  Andrew S. Lowe,et al.  Small animal, whole brain fMRI: Innocuous and nociceptive forepaw stimulation , 2007, NeuroImage.

[38]  Eung Je Woo,et al.  Multi-Frequency Electrical Impedance Tomography System With Automatic Self-Calibration for Long-Term Monitoring , 2014, IEEE Transactions on Biomedical Circuits and Systems.

[39]  O Gilad,et al.  Empirical validation of statistical parametric mapping for group imaging of fast neural activity using electrical impedance tomography , 2016, Physiological measurement.

[40]  J. Seo,et al.  Detection of admittivity anomaly on high-contrast heterogeneous backgrounds using frequency difference EIT , 2015, Physiological measurement.

[41]  Xuetao Shi,et al.  In vivo Monitoring by EIT for the Pig’s Bleeding after Liver Injury , 2009 .

[42]  Andreas Dedner,et al.  A Fast Parallel Solver for the Forward Problem in Electrical Impedance Tomography , 2015, IEEE Transactions on Biomedical Engineering.

[43]  Josselin Garnier,et al.  Spectroscopic imaging of a dilute cell suspension , 2013, 1310.1292.

[44]  C D Binnie,et al.  Factors limiting the application of electrical impedance tomography for identification of regional conductivity changes using scalp electrodes during epileptic seizures in humans , 2006, Physiological measurement.

[45]  H. S. Wolff,et al.  iRun: Horizontal and Vertical Shape of a Region-Based Graph Compression , 2022, Sensors.

[46]  D. C. Barber,et al.  Three-dimensional electrical impedance tomography , 1996, Nature.

[47]  Martin Schuettler,et al.  A novel method for recording neuronal depolarization with recording at 125–825 Hz: implications for imaging fast neural activity in the brain with electrical impedance tomography , 2011, Medical & Biological Engineering & Computing.

[48]  Liu Ruigang,et al.  High precision Multifrequency Electrical Impedance Tomography System and Preliminary imaging results on saline tank , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[49]  Keith D. Paulsen,et al.  A Broadband High-Frequency Electrical Impedance Tomography System for Breast Imaging , 2008, IEEE Transactions on Biomedical Engineering.