Physical Aspects of Quality Assurance in Radiation Therapy

All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means (electronic, mechanical, photocopying, recording, or otherwise) without the prior written permission of the publisher. 48 50 51 52 53 54-4-CHAPTER 1 INTRODUCTION Quality assurance in radiation therapy includes those procedures that ensure a consistent and safe fulfillment of the dose prescription to the target volume, with minimal dose to norma1 tissues and minimal exposure to personnel. A comprehensive quality assurance program is necessary because of the importance of accuracy in dose delivery in radiation therapy. The dare-response curve in radiation therapy is quite steep in certain cases, and there is evidence that a 7-10% change in the dose to the target volume may result in a significant change in tuner control probability [53]. Similarly, such a dose change may also result in a sharp change in the incidence and severity of radiation induced morbidity. Surveying the evidence on effective and excessive dose levels, Herring and Compton [38] concluded that the therapeutic system should be capable of delivering a dose to the tumor volume within 5% of the dose prescribed. Report 24 from the International Commission on Radiation Units and Measurements [53] lists several studier in support of this conclusion. Surveys have indicated that errors occur with some finite frequency even in institutions which are regularly reviewed by physicists from the Radiological Physics Center [31] and the Centers for Radiological Physics [117]. A nationwide survey of Co-60 teletherapy units was conducted by the National Center for Devices and Radiological Health (NCDRH), which was fomerly the bureau of Radiological Earth and the National Bureau of Standards [128}. The survey included 75% of the Co-60 units (751) in use and showed that doses delivered by 17% of these units differed from the requested doses by at least 5% and that 4% of the delivered doses differed by 10% or more. Sources of error in radiation therapy include tumor localization, lack of patient immobilization, field placement, human errors in calibration, calculation, daily patient setup, and equipment-related problems. Many of these equipment and calculational errors can be minimized through a program of periodic checks. The discussion and accompanying flow chart of a systems approach to radiation therapy given in ICRU Report No.24 [53] and in the CR0S Blue Book [106] points out that the planning and delivery of a course …

[1]  J. S. Laughlin,et al.  The Physics of Radiology. 4th ed , 1984 .

[2]  M S Rosenthal,et al.  An automatic seed identification technique for interstitial implants using three isocentric radiographs. , 1983, Medical physics.

[3]  F. Waterman,et al.  Absorbed dose determination for interstitial 125I boost therapy. , 1983, Medical physics.

[4]  R L Morin,et al.  Dose calibrator response to brachytherapy sources: A Monte Carlo and analytic evaluation. , 1983, Medical physics.

[5]  A. Brodsky Principles and practices for keeping occupational radiation exposures at medical institutions as low as reasonably achievable , 1982 .

[6]  R. Siddon,et al.  Utilization of parasternal lymphoscintigraphy in radiation therapy of breast carcinoma. , 1982, International journal of radiation oncology, biology, physics.

[7]  S. C. Sharma,et al.  Methods for routine calibration of brachytherapy sources. , 1982, Radiology.

[8]  B. Bjärngard,et al.  Beam alignment tests for therapy accelerators. , 1981, International journal of radiation oncology, biology, physics.

[9]  R. F. Moyer Quality assurance of electron-beam energy using an aluminum wedge. , 1981, Radiology.

[10]  M. E. Masterson,et al.  Radiation therapy dosimetry reviews by the Centers for Radiological Physics. , 1981, International journal of radiation oncology, biology, physics.

[11]  H I Amols,et al.  A three-film technique for reconstruction of radioactive seed implants. , 1981, Medical physics.

[12]  E C McCullough,et al.  Performance evaluation of computerized treatment planning systems for radiotherapy: external photon beams. , 1980, International journal of radiation oncology, biology, physics.

[13]  C. J. Karzmark Concerning radiotherapy standard dose meter calibration. , 1980, Medical physics.

[14]  H. Wyckoff,et al.  Nationwide survey of cobalt-60 teletherapy. Final report , 1980 .

[15]  T D Kampp,et al.  Fluoroscopic visualization of megavoltage therapeutic x ray beams. , 1980, International journal of radiation oncology, biology, physics.

[16]  Z Russell,et al.  A comprehensive radiology quality assurance program. , 1980, QRB. Quality review bulletin.

[17]  H V Fineberg,et al.  Evaluation of medical practices. The case for technology assessment. , 1979, The New England journal of medicine.

[18]  L E Reinstein,et al.  Contrast enhancement of high-energy radiotherapy films. , 1979, The British journal of radiology.

[19]  O. Linton,et al.  Quality assurance: an idea whose time has come. , 1979, AJR. American journal of roentgenology.

[20]  R T Droege,et al.  Influence of metal screens on contrast in megavoltage x-ray imaging. , 1979, Medical physics.

[21]  P. Wootton,et al.  Dose Specification for Reporting External Beam Therapy with Photons and Electrons , 1979 .

[22]  Jörg W. Müller Some second thoughts on error statements , 1979 .

[23]  R. Derzon How will the new technologies affect medical care? , 1979, Investigative radiology.

[24]  L. Anderson,et al.  Dose inhomogeneity in interstitial implants using 125I seeds. , 1979, International journal of radiation oncology, biology, physics.

[25]  R Lindstrom,et al.  Part two: Implementing a shared quality assurance program. , 1979, Dimensions in health service.

[26]  C. J. Karzmark,et al.  Scheduling of radiological equipment maintenance and generation of archival records using a mini-computer. , 1978, The British journal of radiology.

[27]  W. Hanson,et al.  An analysis of discrepancies encountered by the AAPM radiological physics center. , 1978, Medical physics.

[28]  W. Grant,et al.  A review of the reliability of chamber factors used clinically in the United States (1968--1976). , 1978, Medical physics.

[29]  W. Hendee Computed tomography in radiation therapy treatment planning. , 1978, International journal of radiation oncology, biology, physics.

[30]  R. Stanton,et al.  Ionization collection efficiencies of some ionization chambers in pulsed and continuous radiation beams. , 1978, Medical physics.

[31]  H. Mellins ARRS presidential address: accounting to society: radiologic concern for patient care. , 1978, AJR. American journal of roentgenology.

[32]  D. Buchler,et al.  Errors in reproducing pelvic radiation portals. , 1977, International journal of radiation oncology, biology, physics.

[33]  Robert J. Shalek,et al.  Determination of Absorbed Dose in a Patient Irradiated by Beams of X or Gamma Rays in Radiotherapy Procedures , 1977 .

[34]  D. R. White,et al.  The formulation of tissue substitute materials using basic interaction data. , 1977, Physics in medicine and biology.

[35]  E C McCullough,et al.  The use of CT scanners in megavoltage photon-beam therapy planning. , 1977, Radiology.

[36]  F. Hendrickson,et al.  Accuracy in radiotherapy treatment. , 1977, International journal of radiation oncology, biology, physics.

[37]  L M Rosenstein,et al.  A simple computer program for optimization of source loading in cervical intracavitary applicators. , 1977, The British journal of radiology.

[38]  G. Svensson,et al.  Electron scattering and collimation system for a 12-MeV linear accelerator. , 1976, Medical physics.

[39]  A. Wambersie,et al.  Specification of gamma-ray brachytherapy sources , 1975 .

[40]  P. Winter Computers in Radiotherapy—Physical Aspects. , 1975 .

[41]  C. J. Karzmark,et al.  Code of practice for x-ray therapy linear accelerators. , 1975, Medical physics.

[42]  Reynold F. Brown Protection Against Radiation From Brachytherapy Sources , 1972 .

[43]  James Bernard Kelley,et al.  Precision Measurement and Calibration , 1970 .

[44]  P. Lamperti,et al.  Uniformity of high-energy electron-beam calibrations. , 1969, Physics in medicine and biology.

[45]  M. Stovall,et al.  The M. D. Anderson method for the computation of isodose curves around interstitial and intracavitary radiation sources. I. Dose from linear sources. , 1968, The American journal of roentgenology, radium therapy, and nuclear medicine.

[46]  B. M. Wheatley,et al.  Radiation dosimetry, 2nd edition, Vol. II, Instrumentation , 1968 .

[47]  C. J. Karzmark,et al.  A PROGRAMMED MAINTENANCE SYSTEM FOR COMPLEX RADIOLOGICAL EQUIPMENT. , 1965, The British journal of radiology.

[48]  Williams Kl A code of practice for the dosimetry of 2 to 35 MV X-ray and caesium-137 and cobalt-60 gamma-ray beams , 1964, Physics in medicine and biology.

[49]  A. Stacey,et al.  CAESIUM 137 SOURCES FOR USE IN INTRACAVITARY AND INTERSTITIAL RADIOTHERAPY. , 1964, The British journal of radiology.

[50]  V. Sampiere,et al.  The radiation distribution and dose specification in volume implants of radioactive seeds. , 1957, The American journal of roentgenology, radium therapy, and nuclear medicine.

[51]  O. Glasser Physical Foundations of Radiology , 1953 .

[52]  W. J. Meredith,et al.  Radium Dosage: The Manchester System , 1947, The Indian Medical Gazette.

[53]  P. C. Williams Dosimetry of X-ray and Gamma-Ray Beams for Radiation Therapy in the Energy Range 10 keV to 50 MeV , 1983 .

[54]  E. McCullough,et al.  Airborne concentrations of toxic metals resulting from the use of low melting point lead alloys to construct radiotherapy shielding. , 1981, Medical physics.

[55]  I. Lax,et al.  Collimation of high energy electron beams. , 1980, Acta radiologica. Oncology.

[56]  S. M. Glaser,et al.  Reject film study. Cost and quality considerations in a radiology department. , 1980, QRB. Quality review bulletin.

[57]  R. Finston,et al.  Instrumentation and Monitoring Methods for Radiation Protection , 1980 .

[58]  P. Nuttall A Handbook of Radioactivity Measurement Procedures , 1980 .

[59]  J. Cooper,et al.  Measuring the quality of medical care: a beginning. , 1979, Cancer clinical trials.

[60]  Moseley Rd International conference: the impact of new radiological technology on health care, research and teaching. Keynote address. , 1979, Investigative radiology.

[61]  J. Holt,et al.  Teletherapy alignment: a simplified method of quality control. , 1977, International journal of radiation oncology, biology, physics.

[62]  P. J. Campion,et al.  Code of practice for the detailed statement of accuracy , 1973 .

[63]  G. Fletcher Textbook of radiotherapy , 1973 .

[64]  P. C. Williams,et al.  Dose distribution and stability of radiotherapy electron beams from a linear accelerator. , 1972, The British journal of radiology.

[65]  U. Henschke,et al.  Dimension averaging a simple method for dosimetry of interstitial implants. , 1968, Radiobiologia, radiotherapia.

[66]  C. J. Karzmark,et al.  Some aspects of radiation safety for electron accelerators used for both x-ray and electron therapy. , 1967, The British journal of radiology.