A systematic error analysis of HP 8510 time-domain gating techniques with experimental verification
暂无分享,去创建一个
The errors caused by using time-domain gating techniques on the HP 8510 automatic network analyzer are investigated systematically. These errors are divided into four categories: out-of-gate attenuation error, truncation error, masking error, and multireflection aliasing error. A method for estimating the order of the magnitude errors resulting from time-domain gating is presented. Experiments for supporting the analysis have been designed and carried out, giving results in good agreement with theory.<<ETX>>
[1] Jin Au Kong,et al. Transient Analysis of Single and Coupled Lines with Capacitively-Loaded Junctions , 1986 .
[2] B. Ulriksson. Conversion of frequency-domain data to the time domain , 1986, Proceedings of the IEEE.