Assessment of microdamage caused by orthodontic miniscrew pilot holes

Abstract Introduction The clinical success of orthodontic miniscrew implants may be improved after pre-drilling a pilot hole. However, the amount of microdamage to the bone surface produced by the pilot hole is largely unknown. The present study aimed to quantify the damage to cortical bone after the creation of a pilot hole. Materials and methods Porcine tibia bone was prepared into 30 rectangular bone block specimens with widths of 1.5, 2.0 or 2.5 mm. A pilot hole (0.9 mm diameter) was drilled into each bone specimen. Sequential staining allowed the microdamage on the entry and exit surfaces to be imaged by a confocal laser scanning microscope. Image analysis software was used to measure histomorphometric parameters. Results The specimens had a mean total damage area of 0.95 mm2, a maximum damage radius of 0.66 mm and a maximum crack length of 0.18 mm. There were no significant differences between the three bone thicknesses for any of the histomorphometric parameters on the entry and exit surfaces (p > 0.05). The total damage area was significantly greater on the exit surface compared to the entry surface (p < 0.0001). Conclusions Microdamage caused by the creation of a pilot hole in the cortical bone was minimal and did not appear to be influenced by bone thickness. Therefore, pilot hole pre-drill protocols may be implemented without introducing significant cortical bone microdamage.

[1]  J. Codrington,et al.  The influence of miniscrew insertion torque , 2018, European journal of orthodontics.

[2]  J. Codrington,et al.  Influence of cortical bone thickness on miniscrew microcrack formation , 2017, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[3]  E. M. Boeck,et al.  EFFECT OF ANGLED INSTALLATION OF ORTHODONTIC MINI-IMPLANTS ON PRIMARY STABILITY , 2014 .

[4]  M. Allen,et al.  Orthodontic mini-implant diameter does not affect in-situ linear microcrack generation in the mandible or the maxilla. , 2012, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[5]  William P. Neace,et al.  Microdamage of the cortical bone during mini-implant insertion with self-drilling and self-tapping techniques: a randomized controlled trial. , 2012, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[6]  S. Huja,et al.  Bone damage associated with orthodontic placement of miniscrew implants in an animal model. , 2012, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[7]  Seçil Çehreli,et al.  Primary stability and histomorphometric bone-implant contact of self-drilling and self-tapping orthodontic microimplants. , 2012, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[8]  William Robert Walsh,et al.  Drilling of Bone: Practicality, Limitations and Complications Associated with Surgical Drill-Bits , 2011 .

[9]  D. Drescher,et al.  Impact of bone quality, implant type, and implantation site preparation on insertion torques of mini-implants used for orthodontic anchorage. , 2011, International journal of oral and maxillofacial surgery.

[10]  J. Nyman,et al.  Differences in the mechanical behavior of cortical bone between compression and tension when subjected to progressive loading. , 2009, Journal of the mechanical behavior of biomedical materials.

[11]  M. Hans,et al.  Buccal cortical bone thickness for mini-implant placement. , 2009, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[12]  Benedict Wilmes,et al.  Insertion angle impact on primary stability of orthodontic mini-implants. , 2008, The Angle orthodontist.

[13]  P. Buschang,et al.  2008 survey of AAO members on miniscrew usage. , 2008, Journal of clinical orthodontics : JCO.

[14]  H. Fischer-Brandies,et al.  Microdamage in Cortical Bone due to the Overtightening of Orthodontic Microscrews , 2008, Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie.

[15]  J. Granjeiro,et al.  Dental and skeletal characterization of the BR-1 minipig. , 2007, Veterinary journal.

[16]  Benedict Wilmes,et al.  Parameters Affecting Primary Stability of Orthodontic Mini-implants , 2006, Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie.

[17]  B. Melsen Mini-implants: Where are we? , 2005, Journal of clinical orthodontics : JCO.

[18]  L. Kaban,et al.  Histology of the porcine mandibular distraction wound. , 2005, International journal of oral and maxillofacial surgery.

[19]  B. Melsen,et al.  Miniscrew Implants: The Aarhus Anchorage System , 2005 .

[20]  T. Gunnlaugsson,et al.  Detecting microdamage in bone , 2003, Journal of anatomy.

[21]  R. Martin,et al.  Fatigue Microdamage as an Essential Element of Bone Mechanics and Biology , 2003, Calcified Tissue International.

[22]  Kaddour Bouazza-Marouf,et al.  The detection of drill bit break-through for the enhancement of safety in mechatronic assisted orthopaedic drilling , 1999 .

[23]  D. Laskin,et al.  Analysis of seating and fracturing torque of bicortical screws. , 1994, Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons.

[24]  J. Boutrand Biocompatibility and performance of medical devices , 2012 .

[25]  Hee-Moon Kyung,et al.  Biomechanical and histological comparison of self-drilling and self-tapping orthodontic microimplants in dogs. , 2008, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[26]  R Kanomi,et al.  Mini-implant for orthodontic anchorage. , 1997, Journal of clinical orthodontics : JCO.