A new method to detect ventricular fibrillation from CPR artifact-corrupted ECG based on the ECG alone

Abstract During cardiopulmonary resuscitation, chest compressions (CCs) introduce mechanical activity in the ECG and thus preclude a reliable electrocardiographic (ECG) rhythm diagnosis. To achieve a reliable rhythm analysis, chest compression must therefore be interrupted, and therefore, the probability of the restoration of spontaneous circulation (ROSC) is adversely affected. In recent years, a number of algorithms have been developed to distinguish ventricular fibrillation (VF) rhythm from normal sinus rhythm (SR) without chest compression (CC) interruptions. However, the implementation of most of these algorithms relies on the acquisition of reference signals that are strongly correlated with CC artifacts and makes additional hardware alteration inevitable. In the present work, a novel method (the enhanced LMS method) that effectively suppresses CPR artifacts and can easily use the corrupted ECG signal alone is developed for the reliable detection of the VF rhythm during uninterrupted CCs. The enhanced LMS method was tested using mixtures of CC artifacts and real out-of-hospital ECG recordings for different corruption levels, and it was compared with other established algorithms that use the corrupted ECG signal alone, including the morphology consistency evaluation algorithm and the adaptive stop-band filtering algorithm. The validation results indicate that the enhanced LMS method has superior performance in VF/SR rhythm classification under different artifact interference levels. It is shown that the VF rhythm can be reliably detected using only the corrupted ECG alone. The novel method proposed in this study is promising for identification VF from SR with no hardware alterations for clinical cardiopulmonary resuscitation practice.

[1]  T. Rea,et al.  Incidence of EMS-treated out-of-hospital cardiac arrest in Europe. , 2005, Resuscitation.

[2]  Wanchun Tang,et al.  Adverse Outcomes of Interrupted Precordial Compression During Automated Defibrillation , 2002, Circulation.

[3]  Trygve Eftestøl,et al.  Removal of cardiopulmonary resuscitation artifacts from human ECG using an efficient matching pursuit-like algorithm , 2002, IEEE Transactions on Biomedical Engineering.

[4]  Yongqin Li,et al.  Techniques for artefact filtering from chest compression corrupted ECG signals: good, but not enough. , 2009, Resuscitation.

[5]  Hans G. Feichtinger,et al.  CPR Artifact Removal in Ventricular Fibrillation ECG Signals Using Gabor Multipliers , 2009, IEEE Transactions on Biomedical Engineering.

[6]  George A. Mensah,et al.  Sudden Cardiac Death in the United States, 1989 to 1998 , 2001, Circulation.

[7]  Tobias Werther,et al.  Reduction of CPR artifacts in the ventricular fibrillation ECG by coherent line removal , 2010, Biomedical engineering online.

[8]  Roger D. White,et al.  Technologic advances and program initiatives in public access defibrillation using automated external defibrillators , 2001, Current opinion in critical care.

[9]  Jesus Ruiz,et al.  Detection of ventricular fibrillation in the presence of cardiopulmonary resuscitation artefacts. , 2007, Resuscitation.

[10]  Cantab Ma,et al.  Emanuel Miller. M.A. Cantab., F.R.C.P., D.P.M. , 1970 .

[11]  Jeffrey M. Hausdorff,et al.  Physionet: Components of a New Research Resource for Complex Physiologic Signals". Circu-lation Vol , 2000 .

[12]  M. Eisenberg,et al.  SENSITIVITY, ACCURACY, AND SAFETY OF AN AUTOMATIC EXTERNAL DEFIBRILLATOR Report of a Field Evaluation , 1984, The Lancet.

[13]  Karl Unterkofler,et al.  Detecting Ventricular Fibrillation by Time-Delay Methods , 2007, IEEE Transactions on Biomedical Engineering.

[14]  Jesus Ruiz,et al.  A method to remove CPR artefacts from human ECG using only the recorded ECG. , 2008, Resuscitation.

[15]  D J Roe,et al.  Outcomes of rapid defibrillation by security officers after cardiac arrest in casinos. , 2000, The New England journal of medicine.

[16]  Maaret Castrén,et al.  European Resuscitation Council Guidelines for Resuscitation 2010 Section 2. Adult basic life support and use of automated external defibrillators , 2010, Resuscitation.

[17]  David E. Snyder,et al.  Wide variation in cardiopulmonary resuscitation interruption intervals among commercially available automated external defibrillators may affect survival despite high defibrillation efficacy , 2004, Critical care medicine.

[18]  Trygve Eftestøl,et al.  CPR artifact removal from human ECG using optimal multichannel filtering , 2000, IEEE Transactions on Biomedical Engineering.

[19]  Wanchun Tang,et al.  Adverse effects of interrupting precordial compression during cardiopulmonary resuscitation. , 1997, Critical care medicine.

[20]  Trygve Eftestøl,et al.  A Least Mean-Square Filter for the Estimation of the Cardiopulmonary Resuscitation Artifact Based on the Frequency of the Compressions , 2009, IEEE Transactions on Biomedical Engineering.

[21]  P. Steen,et al.  Reducing CPR artefacts in ventricular fibrillation in vitro. , 2001, Resuscitation.

[22]  Yongqin Li,et al.  An Algorithm Used for Ventricular Fibrillation Detection Without Interrupting Chest Compression , 2012, IEEE Transactions on Biomedical Engineering.

[23]  S. O. Aase,et al.  Feasibility of shock advice analysis during CPR through removal of CPR artefacts from the human ECG. , 2004, Resuscitation.

[24]  Greg Sinibaldi,et al.  Incidence of EMS-treated out-of-hospital cardiac arrest in the United States. , 2004, Resuscitation.

[25]  Tobias Werther,et al.  Strong corruption of electrocardiograms caused by cardiopulmonary resuscitation reduces efficiency of two-channel methods for removing motion artefacts in non-shockable rhythms. , 2009, Resuscitation.

[26]  Joel Xue,et al.  Statistical evaluation of reproducibility of automated ECG measurements: An example from arrhythmogenic right ventricular dysplasia/cardiomyopathy clinic , 2014, Biomed. Signal Process. Control..

[27]  Robert A. Berg,et al.  Importance of Continuous Chest Compressions During Cardiopulmonary Resuscitation: Improved Outcome During a Simulated Single Lay-Rescuer Scenario , 2002, Circulation.

[28]  Klaus Rheinberger,et al.  Removal of CPR Artifacts From the Ventricular Fibrillation ECG by Adaptive Regression on Lagged Reference Signals , 2008, IEEE Transactions on Biomedical Engineering.

[29]  R W Koster,et al.  Out-of-hospital cardiac arrests in Amsterdam and its surrounding areas: results from the Amsterdam resuscitation study (ARREST) in 'Utstein' style. , 1998, Resuscitation.

[30]  P. Steen,et al.  Effects of Interrupting Precordial Compressions on the Calculated Probability of Defibrillation Success During Out-of-Hospital Cardiac Arrest , 2002 .

[31]  R. Koster,et al.  Interruption of cardiopulmonary resuscitation with the use of the automated external defibrillator in out-of-hospital cardiac arrest. , 2003, Annals of emergency medicine.