Clinical image quality perception and its relation to NECR measurements in PET

BackgroundThe purpose of this study is to describe a clinical relation of noise equivalent count rate (NECR) - an objective measurement of positron emission tomography (PET) systems - measured in a large number of patients, to clinical image quality of PET and their relation to 18F-fluoro-2-deoxyglucose (FDG) activity and patient's weight.MethodsA total of 71 consecutive patients were evaluated in this retrospective study. All data was automatically analysed using Matlab to estimate the noise equivalent count rate. Then, image quality was evaluated according to two subjective scores: the IQ local score was a 3-point scale assigned to each bed position in all patients and the IQ global score was a 10-point scale assigned after evaluating the coronal whole-body PET. Patient data was also analysed concerning weight, body mass index, FDG dose at the start of acquisition (DAcq), presence of bowel uptake and presence of FDG-positive pathologic lesions. Two additional parameters were defined for each patient: the ratio between DAcq and patient weight (RDW) and the ratio between DAcq and patient BMI (RDBMI).ResultsClinically perceived image quality in PET has a significant positive correlation with NECR measured in patients, RDW, RDBMI and presence of pathologic lesions. Clinical image quality furthermore has significant negative correlation with weight, body mass index (BMI) and presence of bowel uptake. Thresholds of RDW and RDBMI in which clinical IQ is good to excellent in more than 90% of the patients were 2.6 and 8.0, respectively.ConclusionsClinically perceived image quality in PET systems is positively and significantly related to NECR measured in patients. An optimal threshold for the RDW and RDBMI was defined in which clinical IQ is good to excellent in more than 90% of patients. With this data, it is possible to extrapolate technical as well as clinical image quality to other PET system and to predict clinical image perception.

[1]  Z. Keidar,et al.  PET/CT detection of unexpected gastrointestinal foci of 18F-FDG uptake: incidence, localization patterns, and clinical significance. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[2]  A. Elster,et al.  Recommendations on the Use of 18F-FDG PET in Oncology , 2009 .

[3]  L. Kostakoglu,et al.  PET-CT fusion imaging in differentiating physiologic from pathologic FDG uptake. , 2004, Radiographics : a review publication of the Radiological Society of North America, Inc.

[4]  Stefan Eberl,et al.  Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[5]  Tobias J. Hagge,et al.  Physics , 1929, Nature.

[6]  Christian Vanhove,et al.  Optimal dose of 18F-FDG required for whole-body PET using an LSO PET camera , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[7]  Heinrich R Schelbert,et al.  Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006. , 2007, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[8]  Carole Lartizien,et al.  Optimization of injected dose based on noise equivalent count rates for 2- and 3-dimensional whole-body PET. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  F. Fazio,et al.  Generation of the Acquisition-Specific NEC (AS-NEC) Curves to Optimize the Injected Dose in 3D$^18$F-FDG Whole Body PET Studies , 2006, IEEE Transactions on Nuclear Science.

[10]  Chisato Kondo,et al.  Comparison of Imaging Protocols for 18F-FDG PET/CT in Overweight Patients: Optimizing Scan Duration Versus Administered Dose , 2009, Journal of Nuclear Medicine.

[11]  W. Oyen,et al.  FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0 , 2009, European Journal of Nuclear Medicine and Molecular Imaging.

[12]  Keishi Kitamura,et al.  NEC Density and Liver ROI S/N Ratio for Image Quality Control of Whole-Body FDG-PET Scans: Comparison with Visual Assessment , 2009, Molecular Imaging and Biology.

[13]  A. Scarsbrook,et al.  Evaluation of noise equivalent count parameters as indicators of adult whole-body FDG-PET image quality , 2013, Annals of Nuclear Medicine.

[14]  Julian C Matthews,et al.  Optimization of the Injected Activity in Dynamic 3D PET: A Generalized Approach Using Patient-Specific NECs as Demonstrated by a Series of 15O-H2O Scans , 2009, Journal of Nuclear Medicine.

[15]  E. Hoffman,et al.  Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents counts , 1990 .

[16]  John W. Clark,et al.  Effects of injected dose, BMI and scanner type on NECR and image noise in PET imaging , 2011, Physics in medicine and biology.

[17]  V. Bettinardi,et al.  Physical performance of the new hybrid PET∕CT Discovery-690. , 2011, Medical physics.

[18]  John W. Clark,et al.  Reliability of predicting image signal-to-noise ratio using noise equivalent count rate in PET imaging. , 2012, Medical physics.