The response of fluoroscopic image intensifier-TV systems.

Three different types of X-ray fluoroscopic TV chains were investigated: two standard clinical units, one with a vidicon camera tube, the other with a plumbicon camera tube; and the third was a large flat-screen unit. In each an X-ray beam generated at 100 kVp was passed through 10 cm of H2O before aluminum absorbers of varying thickness were introduced. Five video recordings were made at each absorber thickness. The video image was digitized directly from the disk recording and quantized into 128 gray levels. The five recordings were averaged on a point-to-point basis, and the central 900 averaged points were again averaged to yield a value for the gray level assigned to each particular image. This 30 by 30 matrix of points corresponds to input screen areas of 29, 8.2, and 3.6 sq cm for the three units investigated.