Camera Augmented Mobile C-arm

The Camera Augmented Mobile C-arm (CamC) system that extends a regular mobile C-arm by a video camera provides an X-ray and video image overlay. Thanks to the mirror construction and one time calibration of the device, the acquired X-ray images are co-registered with the video images without any calibration or registration during the intervention. It is very important to quantify and qualify the system before its introduction into the OR. In this communication, we extended the previously performed overlay accuracy analysis of the CamC system by another clinically important parameter, the applied radiation dose for the patient. Since the mirror of the CamC system will absorb and scatter radiation, we introduce a method for estimating the correct applied dose by using an independent dose measurement device. The results show that the mirror absorbs and scatters 39% of X-ray radiation.