Radiation eye dose to medical staff during respiratory endoscopy under X-ray fluoroscopy

Abstract Although the clinical value of fluoroscopically guided respiratory endoscopy (bronchoscopy) is clear, there have been very few studies on the radiation dose received by staff during fluoroscopically guided bronchoscopy. The International Commission on Radiological Protection (ICRP) is suggesting reducing the occupational lens dose limit markedly from 150 to 20 mSv/year, averaged over defined periods of five years. The purpose of this study was to clarify the current occupational eye dose of bronchoscopy staff conducting fluoroscopically guided procedures. We measured the occupational eye doses (3-mm-dose equivalent, Hp(3)) of bronchoscopy staff (physicians and nurses) over a 6-month period. The eye doses of eight physicians and three nurses were recorded using a direct eye dosimeter, the DOSIRIS. We also estimated eye doses using personal dosimeters worn at the neck. The mean ± SD radiation eye doses (DOSIRIS) to physicians and nurses were 7.68 ± 5.27 and 2.41 ± 1.94 mSv/6 months, respectively. The new lens dose limit, 20 mSv/year, may be exceeded among bronchoscopy staff, especially physicians. The eye dose of bronchoscopy staff (both physicians and nurses) was underestimated when measured using a neck dosimeter. Hence, the occupational eye dose of bronchoscopy staff should be monitored. To reduce the occupational eye dose, we recommend that staff performing fluoroscopically guided bronchoscopy wear Pb glasses. correct evaluation of the lens dose [Hp(3)] using an eye dosimeter such as the DOSIRIS is necessary for bronchoscopy staff.

[1]  K. Chida,et al.  A PHANTOM STUDY TO DETERMINE THE OPTIMAL PLACEMENT OF EYE DOSEMETERS ON INTERVENTIONAL CARDIOLOGY STAFF. , 2019, Radiation protection dosimetry.

[2]  H. Toyoshima,et al.  OCCUPATIONAL RADIATION EXPOSURE DOSE OF THE EYE IN DEPARTMENT OF CARDIAC ARRHYTHMIA PHYSICIAN. , 2019, Radiation protection dosimetry.

[3]  H. Toyoshima,et al.  OCCUPATIONAL RADIATION EXPOSURE OF THE EYE IN NEUROVASCULAR INTERVENTIONAL PHYSICIAN. , 2019, Radiation protection dosimetry.

[4]  K. Chida,et al.  Performance of the DOSIRIS™ eye lens dosimeter , 2019, Journal of radiological protection : official journal of the Society for Radiological Protection.

[5]  Koichi Chida,et al.  New real-time patient radiation dosimeter for use in radiofrequency catheter ablation , 2019, Journal of radiation research.

[6]  Colin J. Martin Protecting interventional radiology and cardiology staff: Are current designs of lead glasses and eye dosemeters fit for purpose? , 2018, Journal of radiological protection : official journal of the Society for Radiological Protection.

[7]  R. Guleria,et al.  A survey of flexible bronchoscopy practices in India: The Indian bronchoscopy survey (2017) , 2018, Lung India : official organ of Indian Chest Society.

[8]  K. Chida,et al.  Effectiveness of additional lead shielding to protect staff from scattering radiation during endoscopic retrograde cholangiopancreatography procedures , 2018, Journal of radiation research.

[9]  K. Chida,et al.  Occupational eye dose in interventional cardiology procedures , 2017, Scientific Reports.

[10]  M Ginjaume,et al.  FIRST EURADOS INTERCOMPARISON EXERCISE OF EYE LENS DOSEMETERS FOR MEDICAL APPLICATIONS. , 2016, Radiation protection dosimetry.

[11]  Koichi Chida,et al.  Estimation of the Dose of Radiation Received by Patient and Physician During a Videofluoroscopic Swallowing Study , 2016, Dysphagia.

[12]  R. Eberhardt,et al.  Cross-Sectional Survey on Bronchoscopy in Germany – The Current Status of Clinical Practice , 2016, Pneumologie.

[13]  Masayuki Zuguchi,et al.  Fundamental study of a real-time occupational dosimetry system for interventional radiology staff , 2014, Journal of radiological protection : official journal of the Society for Radiological Protection.

[14]  M. Kitagawa,et al.  The dose and risk factors for radiation exposure to medical staff during endobronchial ultrasonography with a guide sheath for peripheral pulmonary lesions under X-ray fluoroscopy. , 2014, Japanese journal of clinical oncology.

[15]  R. Booton,et al.  British Thoracic Society guideline for diagnostic flexible bronchoscopy in adults: accredited by NICE , 2013, Thorax.

[16]  M. Aoe,et al.  Bronchoscopic practice in Japan: A survey by the Japan Society for Respiratory Endoscopy in 2010 , 2013, Respirology.

[17]  G. O'reilly,et al.  Occupational radiation dose to eyes from endoscopic retrograde cholangiopancreatography procedures in light of the revised eye lens dose limit from the International Commission on Radiological Protection. , 2013, The British journal of radiology.

[18]  B Geiger,et al.  ICRP Publication 120: Radiological Protection in Cardiology , 2013, Annals of the ICRP.

[19]  Masayuki Zuguchi,et al.  Occupational dose in interventional radiology procedures. , 2013, AJR. American journal of roentgenology.

[20]  K. Chida,et al.  Clarifying and visualizing sources of staff-received scattered radiation in interventional procedures. , 2011, AJR. American journal of roentgenology.

[21]  F. J. F. Herth,et al.  Bronchoscopic techniques in diagnosis and staging of lung cancer , 2011, Breathe.

[22]  Masayuki Zuguchi,et al.  Radiation dose to the pediatric cardiac catheterization and intervention patient. , 2010, AJR. American journal of roentgenology.

[23]  Daniel P Steinfort,et al.  Radiation dose to patients and clinicians during fluoroscopically-guided biopsy of peripheral pulmonary lesions. , 2010, Respiratory care.

[24]  T. Ishibashi,et al.  Radiation dose and radiation protection for patients and physicians during interventional procedure. , 2010, Journal of radiation research.

[25]  Masayuki Zuguchi,et al.  Radiation dose of interventional radiology system using a flat-panel detector. , 2009, AJR. American journal of roentgenology.

[26]  S. Takahashi,et al.  Effect of radiation monitoring method and formula differences on estimated physician dose during percutaneous coronary intervention , 2009, Acta radiologica.

[27]  Masayuki Zuguchi,et al.  Usefulness of non-lead aprons in radiation protection for physicians performing interventional procedures. , 2008, Radiation protection dosimetry.

[28]  Masayuki Zuguchi,et al.  Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. , 2006, AJR. American journal of roentgenology.

[29]  J. Valentin,et al.  Abstract: Avoidance of radiation injuries from medical interventional procedures, ICRP Publication 85 , 2000 .

[30]  Masayuki Zuguchi,et al.  Evaluating the performance of a MOSFET dosimeter at diagnostic X-ray energies for interventional radiology , 2009, Radiological physics and technology.

[31]  J. Valentin,et al.  Avoidance of radiation injuries from medical interventional procedures. , 2000, Annals of the ICRP.