Investigation of Taper Failure in a Contemporary Metal-on-Metal Hip Arthroplasty System Through Examination of Unused and Explanted Prostheses

Background: Large-diameter (≥36-mm) metal-on-metal (MoM) total hip replacements have been shown to fail at an unacceptably high rate. Globally, the DePuy Pinnacle prosthesis was the most widely used device of this type. There is evidence to suggest that one of the main reasons for the poor clinical performance of large-diameter MoM prostheses is the metal debris released from the head-stem taper junction—i.e., taper junction failure. The aim of this study was to investigate variation in the as-manufactured finish of the female taper surface and to determine its influence on material loss. We hypothesized that rougher surfaces with higher relative material peaks would be significantly associated with greater taper wear rates. Methods: We analyzed 93 Articul/eze femoral head tapers with a 36-mm bearing diameter that had been used in combination with a Corail titanium uncemented stem. The influence of the surface topography of the as-manufactured female taper surface on taper wear was examined by means of a multiple regression model, taking into account other known variables. Results: We identified great variation in the as-manufactured surface finish of the female taper surface, with a range of measured Ra values from 0.14 to 4.20 &mgr;m. The roughness of the female taper surface appeared to be the most important variable associated with taper wear (p < 0.001). The best-fitting regression model, including duration in vivo, head offset, reduced peak height (Rpk) value, stem shaft angle, and bearing surface wear rate, explained approximately 44% of the variation in taper wear rates. Conclusions: We concluded that the roughness of the female taper surface appears to be a significant factor in metal debris release from head-stem taper junctions. Clinical Relevance: This study shows evidence that previously unappreciated variations in manufacturing processes may have a major impact on the clinical outcomes of patients.

[1]  T. Joyce,et al.  Reducing metal ion release following hip resurfacing arthroplasty. , 2011, The Orthopedic clinics of North America.

[2]  Paul J. Scott The case of surface texture parameter RSm , 2006 .

[3]  R. Barrack,et al.  The effects of blood and fat on Morse taper disassembly forces. , 2009, American journal of orthopedics.

[4]  D. Loch,et al.  Assessing Morse Taper Function: The Relationship between Impaction Force, Disassembly Force, and Design Variables , 1997 .

[5]  D. J. Langton,et al.  Practical considerations for volumetric wear analysis of explanted hip arthroplasties , 2014, Bone & joint research.

[6]  M. Whitehouse,et al.  Adverse local tissue reactions in metal-on-polyethylene total hip arthroplasty due to trunnion corrosion: the risk of misdiagnosis. , 2015, The bone & joint journal.

[7]  L. Blunt,et al.  Material loss at the taper junction of retrieved large head metal‐on‐metal total hip replacements , 2013, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[8]  T J Joyce,et al.  Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear. , 2010, The Journal of bone and joint surgery. British volume.

[9]  J. Hua,et al.  Enhanced wear and corrosion in modular tapers in total hip replacement is associated with the contact area and surface topography , 2013, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[10]  T. Joyce,et al.  Determining material loss from the femoral stem trunnion in hip arthroplasty using a coordinate measuring machine , 2015, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[11]  T. J. Joyce,et al.  Taper junction failure in large-diameter metal-on-metal bearings , 2012, Bone & joint research.

[12]  T. Joyce,et al.  Aseptic lymphocyte-dominated vasculitis-associated lesions are related to changes in metal ion handling in the joint capsules of metal-on-metal hip arthroplasties , 2018, Bone & joint research.

[13]  S. Bull,et al.  Why does titanium alloy wear cobalt chrome alloy despite lower bulk hardness: A nanoindentation study? , 2013 .

[14]  S. Bhatnagar,et al.  Cup anteversion in hip resurfacing: validation of EBRA and the presentation of a simple clinical grading system. , 2010, The Journal of arthroplasty.

[15]  G. Blunn,et al.  Influence of stem type on material loss at the metal-on-metal pinnacle taper junction , 2015, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[16]  T. Joyce,et al.  The clinical implications of metal debris release from the taper junctions and bearing surfaces of metal-on-metal hip arthroplasty: joint fluid and blood metal ion concentrations. , 2016, The bone & joint journal.

[17]  W. M. Rainforth,et al.  Characterisation of the oxide film on the taper interface from retrieved large diameter metal on polymer modular total hip replacements , 2015 .

[18]  R. Wixson,et al.  What do we know about taper corrosion in total hip arthroplasty? , 2014, The Journal of arthroplasty.

[19]  M. Morlock,et al.  Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses. , 2012, Clinical biomechanics.

[20]  R. Berger,et al.  Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. , 2012, The Journal of bone and joint surgery. American volume.

[21]  Joshua J. Jacobs,et al.  Corrosion of metal orthopaedic implants. , 1998, The Journal of bone and joint surgery. American volume.

[22]  R. P. Sidaginamale,et al.  Blood metal ion testing is an effective screening tool to identify poorly performing metal-on-metal bearing surfaces , 2013, Bone & joint research.

[23]  T. Joyce,et al.  Accelerating failure rate of the ASR total hip replacement. , 2011, The Journal of bone and joint surgery. British volume.

[24]  T. Joyce,et al.  Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal‐on‐metal total hip arthroplasty system , 2015, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.