An automated imaging system for radiation biodosimetry

We describe here an automated imaging system developed at the Center for High Throughput Minimally Invasive Radiation Biodosimetry. The imaging system is built around a fast, sensitive sCMOS camera and rapid switchable LED light source. It features complete automation of all the steps of the imaging process and contains built‐in feedback loops to ensure proper operation. The imaging system is intended as a back end to the RABiT—a robotic platform for radiation biodosimetry. It is intended to automate image acquisition and analysis for four biodosimetry assays for which we have developed automated protocols: The Cytokinesis Blocked Micronucleus assay, the γ‐H2AX assay, the Dicentric assay (using PNA or FISH probes) and the RABiT‐BAND assay. Microsc. Res. Tech. 78:587–598, 2015. © 2015 Wiley Periodicals, Inc.

[1]  C. Redon,et al.  γ-H2AX as a biomarker of DNA damage induced by ionizing radiation in human peripheral blood lymphocytes and artificial skin. , 2009, Advances in space research : the official journal of the Committee on Space Research.

[2]  Y. Lawrence Yao,et al.  Development of a robotically-based automated biodosimetry tool for high-throughput radiological triage , 2010 .

[3]  George Iliakis,et al.  Computational Methods for Analysis of Foci: Validation for Radiation-Induced γ-H2AX Foci in Human Cells , 2006, Radiation research.

[4]  M Zeder,et al.  Multispot live‐image autofocusing for high‐throughput microscopy of fluorescently stained bacteria , 2009, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[5]  Barry S Rosenstein,et al.  Development of an Automated γ-H2AX Immunocytochemistry Assay , 2009, Radiation research.

[6]  Hongliang Wang,et al.  The RABiT: A Rapid Automated Biodosimetry Tool for radiological triage. II. Technological developments , 2011, International journal of radiation biology.

[7]  Jing Zhang,et al.  High-content screening moves to the front of the line. , 2006, Drug discovery today.

[8]  D. Brenner,et al.  High Throughput Measurement of γH2AX DSB Repair Kinetics in a Healthy Human Population , 2015, PloS one.

[9]  Leonora Marro,et al.  Biological Dosimetry by the Triage Dicentric Chromosome Assay - Further validation of International Networking. , 2011, Radiation measurements.

[10]  W. Vogel,et al.  An automated scoring procedure for the micronucleus test by image analysis. , 2004, Mutagenesis.

[11]  B W Glickman,et al.  Human micronucleus counts are correlated with age, smoking, and cesium-137 dose in the Goiânia (Brazil) radiological accident. , 1994, Mutation research.

[12]  D. Brenner,et al.  The RABiT: high-throughput technology for assessing global DSB repair , 2014, Radiation and environmental biophysics.

[13]  D. Lloyd,et al.  The role of cytogenetics in early triage of radiation casualties. , 2000, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[14]  Yanxin Li,et al.  Automating dicentric chromosome detection from cytogenetic biodosimetry data. , 2014, Radiation protection dosimetry.

[15]  Giuseppe Schettino,et al.  Use of the γ-H2AX Assay to Investigate DNA Repair Dynamics Following Multiple Radiation Exposures , 2013, PloS one.

[16]  Keiichi Abe,et al.  Topological structural analysis of digitized binary images by border following , 1985, Comput. Vis. Graph. Image Process..

[17]  Michael Model,et al.  Intensity Calibration and Flat‐Field Correction for Fluorescence Microscopes , 2014, Current protocols in cytometry.

[18]  Wen‐jian Li,et al.  Linear energy transfer dependence of the effects of carbon ion beams on adventitious shoot regeneration from in vitro leaf explants of Saintpaulia ionahta , 2006, International journal of radiation biology.

[19]  Kai Rothkamm,et al.  Evidence for a lack of DNA double-strand break repair in human cells exposed to very low x-ray doses , 2003, Proceedings of the National Academy of Sciences of the United States of America.

[20]  T. Erdogan Optical Filters for Wavelength Selection in Fluorescence Instrumentation , 2011, Current protocols in cytometry.

[21]  A W Smeulders,et al.  Robust autofocusing in microscopy. , 2000, Cytometry.

[22]  Jian Zhang,et al.  THE RABIT: A RAPID AUTOMATED BIODOSIMETRY TOOL FOR RADIOLOGICAL TRIAGE , 2010, Health physics.

[23]  Marianne Hanson,et al.  The International Atomic Energy Agency , 2021, Nature.

[24]  G. Sluder,et al.  Microscope basics. , 2013, Methods in cell biology.

[25]  Oscar Déniz,et al.  An automated system for whole microscopic image acquisition and analysis , 2014, Microscopy research and technique.

[26]  Wi-Ho Ha,et al.  Cytogenetic biodosimetry for Fukushima travelers after the nuclear power plant accident: no evidence of enhanced yield of dicentrics , 2012, Journal of radiation research.

[27]  Wang Li,et al.  A 5.5Mpixel 100 frames/sec wide dynamic range low noise CMOS image sensor for scientific applications , 2010, Electronic Imaging.

[28]  M. J. Silva,et al.  Intra- and inter-laboratory variation in the scoring of micronuclei and nucleoplasmic bridges in binucleated human lymphocytes. Results of an international slide-scoring exercise by the HUMN project. , 2003, Mutation research.

[29]  Mikhail Repin,et al.  Next generation platforms for high-throughput biodosimetry. , 2014, Radiation protection dosimetry.

[30]  Bruno Colicchio,et al.  New tool for biological dosimetry: reevaluation and automation of the gold standard method following telomere and centromere staining. , 2014, Mutation research.

[31]  X Y Liu,et al.  Dynamic evaluation of autofocusing for automated microscopic analysis of blood smear and pap smear , 2007, Journal of microscopy.

[32]  A. Jauch,et al.  mBAND: a high resolution multicolor banding technique for the detection of complex intrachromosomal aberrations , 2004, Cytogenetic and Genome Research.

[33]  J F Barquinero,et al.  Automatic scoring of dicentric chromosomes as a tool in large scale radiation accidents. , 2013, Mutation research.

[34]  Laurence Roy,et al.  Automated gamma-H2AX focus scoring method for human lymphocytes after ionizing radiation exposure , 2011 .

[35]  M. Fenech Cytokinesis-block micronucleus cytome assay , 2007, Nature Protocols.

[36]  D. Brenner,et al.  Fast Image Analysis for the Micronucleus Assay in a Fully Automated High-Throughput Biodosimetry System , 2014, Radiation research.

[37]  Jian Zhang,et al.  Design Considerations for a Minimally Invasive High-Throughput Automation System for Radiation Biodosimetry , 2007, 2007 IEEE International Conference on Automation Science and Engineering.

[38]  Y. Yao,et al.  Adapting the γ-H2AX Assay for Automated Processing in Human Lymphocytes. 1. Technological Aspects , 2011, Radiation research.