A fast convolution-based methodology to simulate 2-Dd/3-D cardiac ultrasound images

This paper describes a fast convolution-based methodology for simulating ultrasound images in a 2-D/3-D sector format as typically used in cardiac ultrasound. The conventional convolution model is based on the assumption of a space-invariant point spread function (PSF) and typically results in linear images. These characteristics are not representative for cardiac data sets. The spatial impulse response method (IRM) has excellent accuracy in the linear domain; however, calculation time can become an issue when scatterer numbers become significant and when 3-D volumetric data sets need to be computed. As a solution to these problems, the current manuscript proposes a new convolution-based methodology in which the data sets are produced by reducing the conventional 2-D/3-D convolution model to multiple 1-D convolutions (one for each image line). As an example, simulated 2-D/3-D phantom images are presented along with their gray scale histogram statistics. In addition, the computation time is recorded and contrasted to a commonly used implementation of IRM (Field II). It is shown that COLE can produce anatomically plausible images with local Rayleigh statistics but at improved calculation time (1200 times faster than the reference method).

[1]  G. Tupholme Generation of acoustic pulses by baffled plane pistons , 1969 .

[2]  A. Douglas,et al.  Description of the deformation of the left ventricle by a kinematic model. , 1992, Journal of biomechanics.

[3]  J C Bamber,et al.  Ultrasonic B-scanning: a computer simulation , 1980, Physics in medicine and biology.

[4]  J. Meunier,et al.  Echographic image mean gray level changes with tissue dynamics: a system-based model study , 1995, IEEE Transactions on Biomedical Engineering.

[5]  W. Gellert,et al.  The VNR concise encyclopedia of mathematics , 1977 .

[6]  Jacques Ohayon,et al.  Adapting the Lagrangian speckle model estimator for endovascular elastography: theory and validation with simulated radio-frequency data. , 2004, The Journal of the Acoustical Society of America.

[7]  R. F. Wagner,et al.  Statistical properties of radio-frequency and envelope-detected signals with applications to medical ultrasound. , 1987, Journal of the Optical Society of America. A, Optics and image science.

[8]  S.W. Smith,et al.  High-speed ultrasound volumetric imaging system. I. Transducer design and beam steering , 1991, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[9]  P. Stepanishen Transient Radiation from Pistons in an Infinite Planar Baffle , 1970 .

[10]  J Meunier Tissue motion assessment from 3D echographic speckle tracking. , 1998, Physics in medicine and biology.

[11]  J A Jensen,et al.  A model for the propagation and scattering of ultrasound in tissue. , 1991, The Journal of the Acoustical Society of America.

[12]  W. A. Verhoef,et al.  Texture of B-Mode Echograms: 3-D Simulations and Experiments of the Effects of Diffraction and Scatterer Density , 1985 .