Immediate and late explantation of endovascular aortic grafts: the endovascular technologies experience.

PURPOSE The morbidity and clinical outcome of the failure to successfully repair an abdominal aortic aneurysm with Endovascular Technologies (EVT) grafts, resulting in explantation of the device, was assessed. METHODS The records of all patients worldwide undergoing attempted endovascular repair with EVT devices from February 1993 to October 1997 were retrospectively reviewed. Of 669 patients, 19 (3%) were converted to open procedure with immediate explantation during the initial attempt at endovascular repair, and 27 patients (4%) required explantation at a later date, ranging from 1 day to 40 months. The incidence, morbidity, mortality, and effect on clinical outcome were evaluated. RESULTS Causes of immediate conversion with explantation were: inaccurate deployment of the proximal or distal attachment systems (11 of 19; 58%); twists in the system (3 of 19; 16%); mechanism malfunction during deployment (4 of 19; 21%); and an aortic tear (1 of 19; 5%). Among the 27 patients undergoing late explantation, 20 (74%) did so because of persistent endoleaks. Three cases (11%) were performed because of aneurysm rupture, three (11%) because of graft occlusion, one because of aortic dissection (4%), and one (4%) because of graft migration into the aneurysm sac. The overall perioperative mortality rate was 11% (2 of 19) for immediate explantation and 7% (2 of 27) for late explantation. The average length-of-stay was 11 days for immediate explantation and 14 days for late explantation (NS). Complications included myocardial infarction (4%), pulmonary insufficiency (13%), wound infection (4%), and permanent renal failure (2%). There were no significant differences in the incidence rates of these complications between immediate and late explants. No cases of limb loss occurred. Median American Society of Anesthetists (ASA) classification was 3, and there was no correlation between ASA classification and mortality rate. Average operating time was 374 minutes for immediate explantation (including the time for the failed endovascular procedure) and 185 minutes for late explantation. CONCLUSION Immediate and late explantation are infrequent events, occurring in 3% and 4%, respectively, of attempted EVT endovascular aortic stent placements. The mortality rate was higher for both immediate (11%; P <.05) and late (7%; NS) explantation when compared with the mortality rate of all patients undergoing EVT aortic endograft placement (1.5%). There does not appear to be increased long-term morbidity among patients undergoing successful explantation. Early recognition of the need to convert to open procedure, device improvement, and increased operator experience should continue to minimize the incidence of immediate and late explantation and their associated complications.

[1]  G. White,et al.  Endovascular grafting for abdominal aortic aneurysms: changing incidence and indication for conversion to open operation. , 1998, Cardiovascular surgery.

[2]  P. Walker,et al.  Clinical experience with a bifurcated endovascular graft for abdominal aortic aneurysm repair. , 1996, Journal of vascular surgery.

[3]  R. Moratto,et al.  The Italian trial of endovascular AAA exclusion using the Parodi endograft. , 1997, Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery.

[4]  D. Baker,et al.  Early results of endovascular aortic aneurysm surgery with aortouniiliac graft, contralateral iliac occlusion, and femorofemoral bypass. , 1997, Journal of vascular surgery.

[5]  P. Ziegler,et al.  Three-Year Experience with Modular Stent-Graft Devices for Endovascular AAA Treatment: , 1997 .

[6]  J. Struyven,et al.  Endovascular aortic stent-grafts: transcatheter embolization of persistent perigraft leaks. , 1997, Radiology.

[7]  G. White,et al.  Conversion from endoluminal to open repair of abdominal aortic aneurysms: a hazardous procedure. , 1997, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[8]  M. Lindh,et al.  Changing aneurysmal morphology after endovascular grafting: relation to leakage or persistent perfusion. , 1997, Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery.

[9]  F. Veith,et al.  Endoleaks after endovascular graft treatment of aortic aneurysms: classification, risk factors, and outcome. , 1998, Journal of vascular surgery.

[10]  M. Lindh,et al.  Options for Treatment of Persistent Aneurysm Perfusion after Endovascular Repair , 1996, World Journal of Surgery.

[11]  G. White,et al.  Endoleak--a proposed new terminology to describe incomplete aneurysm exclusion by an endoluminal graft. , 1996, Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery.

[12]  J. Parodi,et al.  Transfemoral intraluminal graft implantation for abdominal aortic aneurysms , 1991, Annals of vascular surgery.

[13]  J. Matsumura,et al.  Clinical consequences of periprosthetic leak after endovascular repair of abdominal aortic aneurysm. Endovascular Technologies Investigators. , 1998, Journal of vascular surgery.

[14]  W. Moore The EVT Tube and Bifurcated Endograft Systems: Technical Considerations and Clinical Summary , 1997, Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery.

[15]  G. White,et al.  Concurrent comparison of endoluminal versus open repair in the treatment of abdominal aortic aneurysms: analysis of 303 patients by life table method. , 1998, Journal of vascular surgery.

[16]  C. Zarins,et al.  Operative repair for aortic aneurysms: the gold standard. , 1997 .

[17]  P. Walker,et al.  Transfemoral insertion of a bifurcated endovascular graft for aortic aneurysm repair: the first 22 patients. , 1995, Cardiovascular surgery.

[18]  Moore Ws The EVT tube and bifurcated endograft systems: technical considerations and clinical summary. EVI Investigators. , 1997 .

[19]  G. White,et al.  Three-Year Experience with the White-Yu Endovascular GAD Graft for Transluminal Repair of Aortic and Iliac Aneurysms , 1997, Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery.

[20]  W S Moore,et al.  Transfemoral endovascular repair of abdominal aortic aneurysm: results of the North American EVT phase 1 trial. EVT Investigators. , 1996, Journal of vascular surgery.

[21]  G. White,et al.  Surgical management of complications following endoluminal grafting of abdominal aortic aneurysms. , 1995, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.