An outbreak of Pseudomonas aeruginosa infections associated with flexible bronchoscopes.

BACKGROUND Endoscopes, including bronchoscopes, are the medical devices most frequently associated with outbreaks of nosocomial infections. We investigated an outbreak of Pseudomonas aeruginosa infections after bronchoscopic procedures. METHODS Microbiologic results were reviewed to determine the rates of recovery of P. aeruginosa from bronchoalveolar-lavage specimens. Environmental samples from endoscopes and the endoscopy suite were cultured. Medical records were reviewed to identify infections in the 14 days after a bronchoscopy. RESULTS The rate of recovery of P. aeruginosa from bronchoalveolar-lavage specimens obtained with use of endoscopy-suite bronchoscopes increased from 10.4 percent at base line to 31.0 percent during the outbreak (relative risk, 2.97; 95 percent confidence interval, 2.28 to 3.90). Cultures of samples from three bronchoscopes grew P. aeruginosa, whereas cultures of samples from the environment, instrument-cleaning machines, and gastrointestinal endoscopes did not. The three bronchoscopes had been part of a nationwide recall. A total of 414 patients underwent bronchoscopy during the outbreak, and there were 48 respiratory tract and bloodstream infections among 39 of these patients (9.4 percent). In 32 infections (66.7 percent), P. aeruginosa was confirmed as a potentially causative organism. Exposure to a potentially contaminated bronchoscope may have had a role in the death of three patients. The rate of recovery of P. aeruginosa returned to base line after the instruments were removed from service. CONCLUSIONS This large outbreak of P. aeruginosa infections related to bronchoscopy was apparently caused by a loose biopsy-port cap in the bronchoscopes. Instrument safety and surveillance methods for bronchoscopy must be improved, and better recall procedures are needed for medical devices.

[1]  Jarvis Wr,et al.  Benchmarking for prevention: the Centers for Disease Control and Prevention's National Nosocomial Infections Surveillance (NNIS) system experience. , 2003 .

[2]  W. Schaffner,et al.  Pseudomonas aeruginosa and Serratia marcescens contamination associated with a manufacturing defect in bronchoscopes. , 2003, The New England journal of medicine.

[3]  L. Kohn,et al.  To Err Is Human : Building a Safer Health System , 2007 .

[4]  H. Colt,et al.  Novel flexible bronchoscope and single-use disposable-sheath endoscope system. A preliminary technology evaluation. , 2000, Chest.

[5]  A. Dimarino,et al.  American Society for Gastrointestinal Endoscopy-Society of Gastroenterology Nurses and Associates Endoscope Reprocessing Guidelines. , 2000, Gastrointestinal endoscopy clinics of North America.

[6]  C. Alvarado,et al.  APIC guideline for infection prevention and control in flexible endoscopy. Association for Professionals in Infection Control. , 1993, American journal of infection control.

[7]  T. Perl,et al.  An outbreak of vancomycin-dependent Enterococcus faecium in a bone marrow transplant unit. , 1999, Clinical infectious diseases : an official publication of the Infectious Diseases Society of America.

[8]  A. Mehta,et al.  Infection control in the bronchoscopy suite. A review. , 1999, Clinics in chest medicine.

[9]  G. Diette,et al.  The higher risk of bleeding in lung transplant recipients from bronchoscopy is independent of traditional bleeding risks: results of a prospective cohort study. , 1999, Chest.

[10]  R. Chaisson,et al.  Transmission of Mycobacterium tuberculosis by a fiberoptic bronchoscope identification by dna fingerprinting , 1997 .

[11]  C. Woodley,et al.  Transmission of a highly drug-resistant strain (strain W1) of Mycobacterium tuberculosis. Community outbreak and nosocomial transmission via a contaminated bronchoscope. , 1997, JAMA.

[12]  M. A. Martin,et al.  APIC guidelines for infection prevention and control in flexible endoscopy. Association for Professionals in Infection Control and Epidemiology, Inc. 1991, 1992, and 1993 APIC Guidelines Committee. , 1994, American journal of infection control.

[13]  R. Goering,et al.  Molecular Epidemiology of Nosocomial Infection: Analysis of Chromosomal Restriction Fragment Patterns by Pulsed-Field Gel Electrophoresis , 1993, Infection Control & Hospital Epidemiology.

[14]  Fred Silverstein,et al.  Transmission of Infection by Gastrointestinal Endoscopy and Bronchoscopy , 1993, Annals of Internal Medicine.

[15]  P. W. Wheeler,et al.  Bronchopulmonary cross-colonization and infection related to mycobacterial contamination of suction valves of bronchoscopes. , 1989, The Journal of infectious diseases.

[16]  J. Sharp,et al.  Contamination of flexible fiberoptic bronchoscopes. , 1983, The American review of respiratory disease.

[17]  K. Nelson,et al.  Transmission of tuberculosis by flexible fiberbronchoscopes. , 1983, The American review of respiratory disease.

[18]  A. Markovitz Inoculation by bronchoscopy. , 1979, The Western journal of medicine.

[19]  S. Webb,et al.  Outbreak of Serratia Marcescens Associated with the Flexible Fiberbronchoscope , 1975 .

[20]  G. Snider,et al.  Fever and pneumonia after flexible fiberoptic bronchoscopy. , 1975, The American review of respiratory disease.