Accuracy of bedside capillary blood glucose measurements in critically ill patients

ObjectiveTo compare the accuracy of fingerstick with laboratory venous plasma glucose measurements (laboratory glucose) in medical ICU patients and to determine the factors which interfere with the accuracy of fingerstick measurements.ParticipantsThe study included 80 consecutive patients aged 58 ± 7 years, BMI 29.5 ± 9.0, and APACHE II score 15 ± 6 (277 simultaneous paired measurements).MeasurementsThis prospective observational study compared fingerstick measurements to simultaneously sampled laboratory glucose once a day in patients in our medical ICU (twice daily if on an insulin infusion). Data recorded included patient demographics, admission diagnoses, APACHE II score, BMI, daily hematocrit, arterial blood gasses, chemistry results, concomitant medications (including vasopressors and corticosteroids), and upper extremity edema. Accuracy was defined as the percentage of paired values not in accord (> 15 mg dl–1 / 0.83 mmol–1 l–1 difference for laboratory values < 75 mg dl–1 / 4.12 mmol–1 l–1 and > 20% difference for laboratory values ≥ 75 mg/dl). Outliers (blood glucose difference > 100 mg dl–1 / 5.56 mmol–1 l–1) were excluded from the correlation and distribution analyses.ResultsMean fingerstick glucose was 129 ± 45 mg/dl (7.2 ± 2.5 mmol/l) and mean laboratory glucose 123 ± 44 mg/dl (6.8 ± 2.4 mmol/l). The correlation coefficient between the two values was 0.9110 (Clinical and Laboratory Standards Institute threshold 0.9751). The mean difference (bias) between the two methods was 8.6 ± 18.6 mg/dl (0.48 ± 1.0 mmol/l) and limits of agreement +45.8 and –28.6 mg/dl (+2.5 and –1.6 mmol/l). Fifty-three (19%) paired measurements in 22 patients were not in accord (CLSI threshold ≤ 5%). In 44 (83%) of these paired measurements fingerstick glucose was greater than laboratory glucose.ConclusionsThe findings suggest that capillary blood glucose as measured by fingerstick is inaccurate in critically ill ICU patients and does not meet the CLSI standard. It is unclear whether the sampling method, device used, or both contributed to this inaccuracy. The wide limits of agreement suggest that fingerstick measurements should be used with great caution in protocols of tight glycemic control.

[1]  K. Wiener Whole Blood Glucose: What are We Actually Measuring? , 1995, Annals of clinical biochemistry.

[2]  T. Whitley,et al.  Accuracy of fingerstick glucose values in shock patients. , 1995, American journal of critical care : an official publication, American Association of Critical-Care Nurses.

[3]  Brian Hutton,et al.  Reliability of point-of-care testing for glucose measurement in critically ill adults* , 2005, Critical care medicine.

[4]  W. Olszewski,et al.  The pathomechanism of posttraumatic edema of the lower limbs: II--Changes in the lymphatic system. , 2003, The Journal of trauma.

[5]  G. Van den Berghe,et al.  Intensive insulin therapy in the medical ICU. , 2006, The New England journal of medicine.

[6]  P. D'Orazio,et al.  Proposal for standardizing direct-reading biosensors for blood glucose. , 1998, Clinical chemistry.

[7]  B. Afessa,et al.  Agreement between bedside blood and plasma glucose measurement in the ICU setting. , 2005, Chest.

[8]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[9]  G J Kost,et al.  Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing , 2001, Critical care medicine.

[10]  Marcus J Schultz,et al.  Predisposing factors for hypoglycemia in the intensive care unit* , 2006, Critical care medicine.

[11]  G. Price,et al.  Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients , 2004, Intensive Care Medicine.

[12]  P. Marik,et al.  Intensive insulin therapy in the ICU: is it now time to jump off the bandwagon? , 2007, Resuscitation.

[13]  G J Kost,et al.  Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer. , 2000, American journal of clinical pathology.

[14]  F. Brunkhorst,et al.  Intensive insulin therapy in critically ill patients. , 2002, The New England journal of medicine.

[15]  E. A. Robertson,et al.  Is antiglycolysis required for routine glucose analysis? , 1979, Clinical chemistry.

[16]  B. Bistrian,et al.  Intensive insulin therapy in critically ill patients. , 2002, The New England journal of medicine.

[17]  E. Draper,et al.  APACHE II: A severity of disease classification system , 1985, Critical care medicine.

[18]  P. Marik,et al.  ACCURACY OF BEDSIDE CAPILLARY BLOOD GLUCOSE MEASUREMENTS IN CRITICALLY ILL PATIENTS.: 259 , 2006 .

[19]  M. Mahoney Fingerstick glucose determination in shock , 1991 .

[20]  K. Kanarek,et al.  Glucose determinations in plasma and serum: potential error related to increased hematocrit. , 1982, Clinical chemistry.