Central haemodynamic variability during sleep in subjects with and without atrial fibrillation

The aim of this study was to develop a Matlab programme for quantitative evaluation and graphical presentation of haemodynamic variability parameters calculated from electrocardiography (ECG) and impedance cardiography (ICG) signals. We also sought to identify and compare patterns of variability in healthy subjects (N) and subjects with atrial fibrillation (AF). We analysed 3000 seconds of ECG and ICG signals acquired continuously during overnight, home recordings of six healthy male subjects (20–22, group N) and five male patients with atrial fibrillation (ages 56–78, group AF). Haemodynamic variability was determined using standard deviation and indices describing asymmetry of stroke volume, systolic time intervals (STI), and their derivatives. Stroke volume (SV), amplitude of impedance signal (dz/dtmax), ejection time (ET) and pre-ejection period (PEP) were calculated automatically. The standard deviation (SD) of SV varied in the range of 9–16 ml in normal subjects, versus 12–39 ml in AF subjects. SD for ET was 14–42 ms for N and 31–68 ms for AF. PEP SD varied 8–33 ms for N and 12–38 ms for AF. Preliminary results show that the patterns of haemodynamic variability in AF patients differed from those observed in non-patients for most of the analysed indices.

[1]  G. Cybulski Influence of age on the immediate cardiovascular response to orthostatic manoeuvre , 1996, European Journal of Applied Physiology and Occupational Physiology.

[2]  W. Niewiadomski,et al.  Early hemodynamic response to the tilt test in patients with syncope , 2014, Archives of medical science : AMS.

[3]  Jaroslaw Piskorski,et al.  Heart rate asymmetry by Poincaré plots of RR intervals , 2006, Biomedizinische Technik. Biomedical engineering.

[4]  Janusz Siebert,et al.  Stroke volume variability and heart rate power spectrum in relation to posture changes in healthy subjects. , 2004, Medical science monitor : international medical journal of experimental and clinical research.

[5]  G. Cybulski,et al.  Ambulatory impedance cardiography: new possibilities. , 2000, Journal of applied physiology.

[6]  Gerard Cybulski,et al.  Impedance cardiography: recent advancements. , 2012, Cardiology journal.

[7]  R. Patterson,et al.  Development and evaluation of an impedance cardiac output system. , 1966, Aerospace medicine.

[8]  Gerard Cybulski,et al.  The Quality of Automatic Artifact Identification in Ambulatory Impedance Cardiography Monitoring , 2017 .

[9]  T. Palko,et al.  Ambulatory monitoring device for central hemodynamic and ECG signal recording on PCMCI flash memory cards , 1995, Computers in Cardiology 1995.

[10]  M. Elstad,et al.  Cardiac stroke volume variability measured non-invasively by three methods for detection of central hypovolemia in healthy humans , 2016, European Journal of Applied Physiology.

[11]  G. Cybulski,et al.  Application of impedance cardiographic ambulatory monitoring device for analysis of central hemodynamic variability in atrial fibrillation , 1999, Computers in Cardiology 1999. Vol.26 (Cat. No.99CH37004).

[12]  M. Fischer,et al.  Electrical bioimpedance cardiography: an old technology with new hopes for the future. , 2014, Journal of cardiothoracic and vascular anesthesia.

[13]  K. Nazar,et al.  Hemodynamic and neuroendocrine predictors of lower body negative pressure (LBNP) intolerance in healthy young men. , 2005, Journal of physiology and pharmacology : an official journal of the Polish Physiological Society.

[14]  G. Cybulski,et al.  Stroke volume and systolic time intervals: Beat-to-beat comparison between echocardiography and ambulatory impedance cardiography in supine and tilted positions , 2004, Medical and Biological Engineering and Computing.

[15]  G. Berntson,et al.  Where to B in dZ/dt. , 2007, Psychophysiology.