Optimal focusing by spatio-temporal inverse filter. II. Experiments. Application to focusing through absorbing and reverberating media.

To focus ultrasonic waves in an unknown heterogeneous medium using a phased array, one has to calculate the optimal set of signals to be applied on the transducers of the array. (In most applications of ultrasound, medical imaging, medical therapy, nondestructive testing, the first step consists of focusing a broadband ultrasound beam deeply inside the medium to be investigated.) Focusing in a homogeneous medium simply requires to compensate for the varying focus-array elements geometrical distances. Nevertheless, heterogeneities in the medium, in terms of speed of sound, density, or absorption, may strongly degrade the focusing. Different techniques have been developed in order to correct such aberrations induced by heterogeneous media (time reversal, speckle brightness, for example). In the companion to this paper, a new broadband focusing technique was investigated: the spatio-temporal inverse filter. Experimental results obtained in various media, such as reverberating and absorbing media, are presented here. In particular, intraplate echoes suppression and high-quality focusing through a human skull, as well as hyper-resolution in a reverberating medium, will be shown. It is important to notice that all these experiments were performed with fully programmable multichannel electronics whose use is required to fully exploit the spatio-temporal technique.

[1]  J. Barger,et al.  Acoustical properties of the human skull. , 1978, The Journal of the Acoustical Society of America.

[2]  Mathias Fink,et al.  Time reversal in multiply scattering media , 1998 .

[3]  J.-L. Thomas,et al.  Ultrasonic beam focusing through tissue inhomogeneities with a time reversal mirror: application to transskull therapy , 1996, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[4]  J L Thomas,et al.  Time reversal and the inverse filter. , 2000, The Journal of the Acoustical Society of America.

[5]  W. R. Scott,et al.  Ultrasonic spectrum analysis for nondestructive testing of layered composite materials , 1977 .

[6]  S. W. Smith,et al.  Phase aberration correction in medical ultrasound using speckle brightness as a quality factor. , 1989, The Journal of the Acoustical Society of America.

[7]  M. O'Donnell,et al.  Phase Aberration Measurements in Medical Ultrasound: Human Studies , 1988 .

[8]  R C Waag,et al.  In vivo and in vitro ultrasound beam distortion measurements of a large aperture and a conventional aperture focussed transducer. , 1988, Ultrasound in medicine & biology.

[9]  G E Trahey,et al.  The impact of sound speed errors on medical ultrasound imaging. , 2000, The Journal of the Acoustical Society of America.

[10]  J L Thomas,et al.  Optimal focusing by spatio-temporal inverse filter. I. Basic principles. , 2001, The Journal of the Acoustical Society of America.

[11]  J. L. Thomas,et al.  Focusing and steering through absorbing and aberrating layers: application to ultrasonic propagation through the skull. , 1998, The Journal of the Acoustical Society of America.

[12]  J. Campbell,et al.  Effect of the skull in degrading the display of echoencephalographic B and C scans. , 1968, The Journal of the Acoustical Society of America.

[13]  H. Kuttruff A mathematical model for noise propagation between buildings , 1982 .

[14]  M. Fink,et al.  Time reversal of ultrasonic fields. I. Basic principles , 1992, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.