A very low frequency ac-dc transfer standard
暂无分享,去创建一个
A low frequency ac-dc transfer standard has been developed to satisfy the NML demand for very low frequency ac-dc transfer calibrations. Tests performed using digitally synthesized signals and synchronised dc measurements showed less than 13 parts per million variation in the ac-dc difference of the standard within the frequency range 0.01 to 100 Hz. >
[1] M. Klonz,et al. Increasing the time constant of a thin film multijunction thermal converter for low frequency application , 1991 .
[2] G. W. Small,et al. Synthesizing signal generator for use with lock-in amplifiers in audiofrequency measurements , 1986, IEEE Transactions on Instrumentation and Measurement.