Direct Current Make and Break Thresholds for Pacemaker Electrodes on the Canine Ventricle

Thresholds to anodal make, anodal break, cathodal make and cathodal break have been studied throughout the cardiac cycle in 13 dogs under pentobarbital anesthesia. Unipolar direct current pulses were applied through epicardial pacemaker electrodes to the left ventricle. Make and break responses were separated by letting the break and make occur in the refractory periods of the following and preceding cycle, respectively. Four distinct threshold interval curves were obtained from each of 17 electrodes. The anodal break and cathodal break curves showed an early diastolic dip. Dips were deepest in the anodal break curves. These also had the shortest effective refractory period. Thresholds for cathodal make and anodal make first dipped steeply, then gradually sloped down towards end-diastolic levels. The magnitude of end-diastolic mean threshold values increased in the following order: cathodal make, anodal make, cathodal break, anodal break. Comparison of threshold interval curves for make and break responses with those for shorter rectangular anodal and cathodal curves throughout the cardiac cycle supports the hypothesis that the threshold behavior to a short rectangular pulse is determined by whichever of two thresholdsis lowest: make or break at that particular cycle interval.

[1]  C. Brooks,et al.  Excitability of the mammalian ventricle throughout the cardiac cycle. , 1950, American Journal of Physiology.

[2]  E. Benson,et al.  THE ULTRASTRUCTURE OF FROG VENTRICULAR CARDIAC MUSCLE AND ITS RELATIONSHIP TO MECHANISMS OF EXCITATION-CONTRACTION COUPLING , 1968, The Journal of cell biology.

[3]  B. Hoffman,et al.  Effect of rate on excitability of dog's ventricle. , 1951, The American journal of physiology.

[4]  E. Sowton,et al.  Electrical threshold of the human heart. , 1966, British heart journal.

[5]  C. Brooks,et al.  Excitability cycle of mammalian auricle. , 1950, The American journal of physiology.

[6]  C. Brooks,et al.  Anodal effects during the refactory period of cardiac muscle. , 1956, Journal of cellular and comparative physiology.

[7]  E H Wood,et al.  Inotropic Effects of Electric Currents , 1969, Circulation research.

[8]  E. S. An Introduction to Physiology , 1903, Nature.

[9]  Charles H. Altshuler Excitability of the Heart , 1956 .

[10]  B. Hoffman,et al.  Anodal excitation of cardiac muscle. , 1957, The American journal of physiology.

[11]  B. Hoffman,et al.  PROPAGATED REPOLARIZATION IN HEART MUSCLE , 1958, The Journal of general physiology.

[12]  L. Barr,et al.  Propagation of Action Potentials and the Structure of the Nexus in Cardiac Muscle , 1965, The Journal of general physiology.

[13]  C. Brooks,et al.  Membrane excitability of the frog ventricle examined by long pulses. , 1969, The American journal of physiology.

[14]  S. Pettersson,et al.  Artificial pacemaker for treatment of Adams-Stokes syndrome and slow heart rate☆ , 1963 .

[15]  C. Y. Kao,et al.  Refractoriness in cardiac muscle. , 1957, The American journal of physiology.

[16]  J. Strackee,et al.  The Excitability Cycle of the Dog's Left Ventricle Determined by Anodal, Cathodal, and Bipolar Stimulation , 1956, Circulation research.

[17]  R. Hoshi,et al.  Excitability cycle of cardiac muscle examined by intracellular stimulation. , 1962, The Japanese journal of physiology.

[18]  C. Brooks,et al.  Vulnerability to fibrillation and the ventricular-excitability curve. , 1951, The American journal of physiology.

[19]  A. Rosenblueth,et al.  THE ACTION OF ELECTRICAL STIMULI ON THE TURTLE'S VENTRICLE , 1942 .