Torsional frequency response of cantilever beams immersed in viscous fluids with applications to the atomic force microscope

The frequency response of a cantilever beam is strongly dependent on the fluid in which it is immersed. In a companion study, Sader [J. Appl. Phys. 84, 64, (1998)] presented a theoretical model for the flexural vibrational response of a cantilever beam, that is immersed in a viscous fluid, and excited by an arbitrary driving force. Due to its relevance to applications of the atomic force microscope (AFM), we extend the analysis of Sader to the related problem of torsional vibrations, and also consider the special case where the cantilever is excited by a thermal driving force. Since longitudinal deformations of AFM cantilevers are not measured normally, combination of the present theoretical model and that of the companion study enables the complete vibrational response of an AFM cantilever beam, that is immersed in a viscous fluid, to be calculated.

[1]  Irene A. Stegun,et al.  Handbook of Mathematical Functions. , 1966 .

[2]  R. Plunkett,et al.  Formulas for Stress and Strain , 1965 .

[3]  Thorsten Hesjedal,et al.  Transverse surface acoustic wave detection by scanning acoustic force microscopy , 1998 .

[4]  Rutland,et al.  Tip friction - torsional spring constant determination. , 2000, Colloids and surfaces. B, Biointerfaces.

[5]  E. Bamberg,et al.  Scan speed limit in atomic force microscopy , 1993 .

[6]  F. Elmer,et al.  Eigenfrequencies of a rectangular atomic force microscope cantilever in a medium , 1997 .

[7]  E. Tuck Calculation of unsteady flows due to small motions of cylinders in a viscous fluid , 1969 .

[8]  A. Love A treatise on the mathematical theory of elasticity , 1892 .

[9]  Hans-Jürgen Butt,et al.  Calculation of thermal noise in atomic force microscopy , 1995 .

[10]  Paul K. Hansma,et al.  Studies of vibrating atomic force microscope cantilevers in liquid , 1996 .

[11]  Gus Gurley,et al.  Short cantilevers for atomic force microscopy , 1996 .

[12]  E. McFarland,et al.  Multi-mode noise analysis of cantilevers for scanning probe microscopy , 1997 .

[13]  G. Batchelor,et al.  An Introduction to Fluid Dynamics , 1968 .

[14]  J. Sader,et al.  Calibration of rectangular atomic force microscope cantilevers , 1999 .

[15]  John E. Sader,et al.  Experimental validation of theoretical models for the frequency response of atomic force microscope cantilever beams immersed in fluids , 2000 .

[16]  J. Sader Frequency response of cantilever beams immersed in viscous fluids with applications to the atomic force microscope , 1998 .

[17]  Stephen P. Timoshenko,et al.  Vibration problems in engineering , 1928 .

[18]  Thomas Thundat,et al.  RESONANCE RESPONSE OF SCANNING FORCE MICROSCOPY CANTILEVERS , 1994 .

[19]  T. Hesjedal,et al.  Influence of surface acoustic waves on lateral forces in scanning force microscopies , 2001 .