Comparison of malleable stylet and reusable and disposable bougies by paramedics in a simulated difficult intubation

In a randomised crossover study, 60 ambulance paramedics attempted tracheal intubation of a manikin model of a Cormack and Lehane grade 3/4 view using a Portex stylet, Portex and Frova single‐use bougies, and a Portex reusable bougie. Tracheal intubation within 30 s was achieved by 34/60 (57%) using the stylet, 18/60 (30%) using a Portex single‐use bougie, 16/60 (27%) using a Frova single‐use bougie and 5/60 (8%) using a Portex reusable bougie. The proportion intubating within 30 s was significantly higher with the stylet compared with any bougie (p < 0.001), but significantly lower with a Portex reusable bougie than any other device (p < 0.004). Participants rated the Portex reusable bougie as significantly more difficult to use than the other devices (p < 0.001). There was no evidence of a relationship between previous experience and success rate for any device.

[1]  D. Kupas,et al.  Comparison of Bougie-Assisted Intubation with Traditional Endotracheal Intubation in a Simulated Difficult Airway , 2011, Prehospital emergency care : official journal of the National Association of EMS Physicians and the National Association of State EMS Directors.

[2]  Charles A Czeisler,et al.  Does Simulator-Based Clinical Performance Correlate With Actual Hospital Behavior? The Effect of Extended Work Hours on Patient Care Provided by Medical Interns , 2010, Academic medicine : journal of the Association of American Medical Colleges.

[3]  J. Nolan,et al.  A critical reassessment of ambulance service airway management in prehospital care: Joint Royal Colleges Ambulance Liaison Committee Airway Working Group, June 2008 , 2010, Emergency Medicine Journal.

[4]  M. Woollard,et al.  The College of Paramedics (British Paramedic Association) position paper regarding the Joint Royal Colleges Ambulance Liaison Committee recommendations on paramedic intubation , 2010, Emergency Medicine Journal.

[5]  S. Charman,et al.  The effect of temperature on bougies: a photographic and manikin study * , 2008, Anaesthesia.

[6]  C. Deakin,et al.  Securing the prehospital airway: a comparison of laryngeal mask insertion and endotracheal intubation by UK paramedics , 2004, Emergency Medicine Journal.

[7]  M. Woollard,et al.  Prehospital airway management in Ambulance Services in the United Kingdom * , 2004, Anaesthesia.

[8]  W. Mapleson,et al.  Evaluation of Frova, single‐use intubation introducer, in a manikin. Comparison with Eschmann multiple‐use introducer and Portex single‐use introducer * , 2004, Anaesthesia.

[9]  E. Brizendine,et al.  Emergency physician-verified out-of-hospital intubation: miss rates by paramedics. , 2004, Academic emergency medicine : official journal of the Society for Academic Emergency Medicine.

[10]  A R Wilkes,et al.  Bougie‐assisted difficult airway management in a manikin – the effect of position held on placement and force exerted by the tip * , 2004, Anaesthesia.

[11]  J. Burton,et al.  Unrecognized misplacement of endotracheal tubes in a mixed urban to rural emergency medical services setting. , 2003, Academic emergency medicine : official journal of the Society for Academic Emergency Medicine.

[12]  A R Wilkes,et al.  A comparison of simulated difficult intubation with multiple‐use and single‐use bougies in a manikin * , 2003, Anaesthesia.

[13]  J. Falk,et al.  Misplaced endotracheal tubes by paramedics in an urban emergency medical services system. , 2001, Annals of emergency medicine.

[14]  J. M. Cupitt Microbial contamination of gum elastic bougies , 2000, Anaesthesia.

[15]  J. Benumof Comparison of the gum elastic bougie and the stylet. , 1997, Anaesthesia.

[16]  R S Cormack,et al.  Difficult tracheal intubation in obstetrics , 1983, Anaesthesia.