Influence of measured objects parameters on CMM touch trigger probe accuracy of probing

Influence of the measured objects parameters on coordinate measuring machine (CMM) touch trigger probe inaccuracy of probing is investigated. Factors such as material stiffness of measured workpiece, its surface shape and roughness are taken into consideration. The measuring method out off CMM, based on a rotary table and a reference axis is applied for tests. The concept is based on measurements of the distance between reference and triggering points in various directions. The reference points are established by the rotation of a precise rotary table. The statistical analysis of variance and regression is applied for data analysis.